模拟研究评估分析误差对新生儿血糖测量分类的影响:基于加拿大儿科学会行动阈值

A Simulation Study to Assess the Effect of Analytic Error on Neonatal Glucose Measurements Using the Canadian Pediatric Society Position Statement Action Thresholds.
指南 / 共识 · J Diabetes Sci Technol · 2019-11-06
一句话结论分析误差可导致新生儿血糖仪结果在加拿大儿科学会行动阈值处误分类,5%变异系数下偏倚+10 mg/dL时假性血糖正常误分类0.8%–5%,偏倚-10 mg/dL时假性低血糖误分类3%–12.5%。

本研究基于新生儿重症监护病房(NICU)的23 749例血糖数据,通过有限混合模型生成100 000个真实血糖值的模拟数据集,并加入偏倚和变异系数(CV)以模拟测量值,采用蒙特卡洛模拟评估在加拿大儿科学会(CPS)行动阈值(32、36、47 mg/dL)处的误分类率。结果显示,在CV 5%且偏倚+10 mg/dL时,32和47 mg/dL阈值处因偏倚导致的假性血糖正常误分类率为0.8%–5%;在CV 5%且偏倚-10 mg/dL时,32和47 mg/dL阈值处因偏倚和变异共同导致的假性低血糖误分类率为3%–12.5%。作者结论认为,使用CPS行动阈值时,该模拟模型可预测在特定偏倚和变异值下NICU血糖仪结果导致不适当临床措施(遗漏治疗或过度治疗)的风险比例。

要点

  • 在CV 5%且偏倚+10 mg/dL时,32和47 mg/dL阈值处因偏倚导致的假性血糖正常误分类率为0.8%–5%。
  • 在CV 5%且偏倚-10 mg/dL时,32和47 mg/dL阈值处因偏倚和变异共同导致的假性低血糖误分类率为3%–12.5%。
  • 评估的偏倚范围为-20至+20 mg/dL,CV范围为0%至20%,在32、36、47 mg/dL阈值处预测误分类率。
  • 模拟数据集来自NICU血糖数据(N=23 749)通过有限混合模型分析生成100 000个真实血糖值。
  • 使用CPS行动阈值时,模拟模型可预测在特定偏倚和变异值下NICU血糖仪结果导致不适当临床措施的风险比例。

适用人群与场景

  • 有新生儿低血糖风险的婴儿,其血糖监测使用CPS指南推荐的筛查和即时管理算法。

诊断标准与关键阈值

  • CPS行动阈值:32、36、47 mg/dL,提供时间依赖性血糖浓度行动阈值。

研究设计与关键结果

  • 研究类型:模拟研究,基于NICU血糖数据(N=23 749)通过有限混合模型分析生成100 000个真实血糖值的模拟数据集。
  • 干预与对照:向真实血糖值加入偏倚(-20至+20 mg/dL)和变异系数(0%至20%)以创建测量值,采用蒙特卡洛模拟评估在CPS行动阈值(32、36、47 mg/dL)处的误分类率。
  • 主要终点结果:在CV 5%且偏倚+10 mg/dL时,32和47 mg/dL阈值处因偏倚导致的假性血糖正常误分类率为0.8%–5%;在CV 5%且偏倚-10 mg/dL时,32和47 mg/dL阈值处因偏倚和变异共同导致的假性低血糖误分类率为3%–12.5%。

证据类型、等级与局限

  • 证据类型:模拟研究,基于NICU血糖数据的有限混合模型分析和蒙特卡洛模拟。
  • 局限:原文未报告局限性

对床旁实践的改变

  • 作者认为,使用CPS行动阈值时,该模拟模型可预测在特定偏倚和变异值下NICU血糖仪结果导致不适当临床措施(遗漏治疗或过度治疗)的风险比例。

指南或共识出处

  • 加拿大儿科学会(CPS)立场声明:新生儿低血糖筛查和即时管理算法,提供时间依赖性血糖浓度行动阈值。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「处置流程与药物剂量」「红旗征与不能漏的情况」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

BACKGROUND: The Canadian Pediatric Society (CPS) has endorsed an algorithm for the screening and immediate management of babies at risk of neonatal hypoglycemia that provides time-dependent glucose concentration action thresholds. The objective of this study was to evaluate the impact of glucose analytic error (bias and imprecision) on the misclassification of glucose meter results from a neonatal intensive care unit (NICU) using the CPS guidelines.

METHODS: A simulation dataset of true glucose values (N = 100 000) was derived by finite mixture model analysis of NICU glucose data (N = 23 749). Bias and imprecision were added to create measured glucose values. The percentages of measured glucose values that were misclassified at CPS action thresholds were determined by Monte Carlo simulation.

RESULTS: Measurement biases ranging from -20 to +20 mg/dL combined with coefficients of variation 0% to 20% were evaluated to predict misclassification rates at 32, 36, and 47 mg/dL. The models demonstrated low risk of false normoglycemia-at 5% CV and +10 mg/dL bias: 0.8% to 5% misclassification at the 32 and 47 mg/dL thresholds due to bias. The models demonstrated risk of false hypoglycemia-at 5% CV and -10 mg/dL bias: 3% to 12.5% misclassification at 32 and 47 mg/dL thresholds due to both bias and imprecision.

CONCLUSION: Using CPS action thresholds, the simulation model predicted the proportion of neonates at risk of inappropriate clinical action-both of omission or "failure to treat" and commission or "overtreatment" in response to NICU glucose meter results at specific bias and imprecision values.