呼吸机内部回路消毒在机械通气患者中的应用观察
本研究纳入2017年1月至2018年6月杭州市老年病医院ICU的39例机械通气患者,在常规管路更换、无菌操作、感染预防等处理基础上,机械通气2周后使用呼吸机内部回路消毒器进行消毒。消毒前回路内检出9604株菌,与患者下呼吸道菌株一致率41%,微生物检测合格率仅5.1%,2周内VAP发生13例(33.3%)。消毒后菌落数从150(15,500)cfu/cm2降至0(0,20)cfu/cm2(P<0.01),合格率升至71.8%(P<0.01)。机械通气2周时VAP发生率降至20.5%(8/39),但与消毒前33.3%(13/39)相比无显著差异(P>0.05)。
要点
- 消毒前呼吸机内部回路检出9604株菌,以革兰阴性杆菌为主(8687株,90.4%),与患者下呼吸道菌株一致率41%。
- 消毒前回路微生物检测合格率仅5.1%,消毒后升至71.8%(P<0.01)。
- 消毒后回路菌落数从150(15,500)cfu/cm2降至0(0,20)cfu/cm2(P<0.01)。
- 机械通气2周VAP发生率消毒前33.3%(13/39) vs 消毒后20.5%(8/39),无显著差异(P>0.05)。
- 作者认为呼吸机内部回路可检测病原体且与患者痰培养一致性高,消毒可显著减少呼吸机空气污染并减少VAP发生。
适用人群与场景
- 2017年1月至2018年6月杭州市老年病医院ICU收治的机械通气患者,共39例,男23例、女16例,年龄65-97岁,平均(78.7±7.6)岁。
处置流程与药物剂量
- 常规机械通气治疗包括管路更换、无菌操作、感染预防等。
- 机械通气2周后使用呼吸机内部回路消毒器对呼吸机内部回路进行消毒。
- 消毒前后在呼吸机内部回路呼气端口3 cm处进行微生物采样检测,以菌落数<5 cfu/cm2且未检出致病菌为合格。
- 观察期间若患者并发VAP接受抗感染治疗,感染控制后更换带内部回路消毒的呼吸机,并再次纳入观察组。
研究设计与关键结果
- 单中心研究,纳入39例机械通气患者,全部纳入分析。
- 消毒前呼吸机内部回路检出9604株菌,其中革兰阴性杆菌8687株(90.4%)、革兰阳性球菌902株(9.4%)、真菌15株(0.2%),与患者下呼吸道菌株一致率41%;微生物检测合格率5.1%;2周机械通气期间VAP发生13例(33.3%)。
- 消毒后回路检出785株病原体,其中革兰阴性杆菌688株(87.7%)、革兰阳性球菌92株(11.7%)、真菌5株(0.6%);菌落数较消毒前显著降低[cfu/cm2:0(0,20) vs 150(15,500),P<0.01];微生物检测合格率达71.8%,显著高于消毒前(P<0.01)。
- 机械通气2周时VAP发生率消毒后20.5%(8/39) vs 消毒前33.3%(13/39),差异无统计学意义(P>0.05)。
证据类型、等级与局限
- 研究类型:观察性研究(单中心,39例)。
- 局限:原文未报告局限性。
对床旁实践的改变
- 作者认为呼吸机内部回路可检测病原体且与机械通气患者痰培养一致性高(一致率41%)。
- 作者认为对病原体消毒可显著减少呼吸机空气污染并减少VAP发生,但原文VAP发生率下降无统计学差异(20.5% vs 33.3%,P>0.05)。
提炼说明
- 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开
OBJECTIVE: To understand the internal circuit contamination of ventilator in mechanical ventilation patients, to evaluate the effect of ventilator internal circuit disinfection and the impact on the incidence of ventilator-associated pneumonia (VAP).
METHODS: A total of 39 patients with mechanical ventilation admitted to intensive care unit (ICU) of Hangzhou Geriatric Hospital from January 2017 to June 2018 were enrolled. Routine mechanical ventilation treatments for patients included pipeline replacement, aseptic operation, prevention of infection, etc. After 2 weeks of mechanical ventilation, the internal circuit of the ventilator was disinfected using the internal circuit sterilizer of the ventilator. Microorganism sampling and detection at 3 cm to the exhalation port of the internal circuit of the ventilator was performed before and after disinfection. The number of colonies was < 5 cfu/cm2 and no pathogenic bacteria could be detected. During the observation period, if the patient was complicated by VAP for anti-infective treatment, the ventilator with internal loop disinfection was replaced after infection control, and was incorporated again into the group for observation. The number of microbial colonies in the internal circuit of the ventilator before and after disinfection, the microbiological test pass rate and the incidences of VAP during the 2 weeks were observed.
RESULTS: All 39 patients were included in the analysis, with 23 male and 16 female; with age of 65-97 years old, average (78.7±7.6) years old. Before the disinfection, 9 604 strains were detected in the internal circuit of the ventilator, including 8 687 strains of Gram-negative bacilli (90.4%), 902 strains of Gram-positive cocci (9.4%), and 15 strains of fungi (0.2%), which were detected in the lower respiratory tract of the patients. The strain concordance rate was 41%. The qualified rate of microbial detection in the internal circuit of the ventilator was 5.1%; 13 cases (33.3%) of VAP occurred during 2 weeks of mechanical ventilation. After disinfection, 785 strains of pathogens were detected in the internal circuit of the ventilator, and the number of colonies was significantly reduced compared with that before disinfection [cfu/cm2: 0 (0, 20) vs. 150 (15, 500), P < 0.01], of which 688 strains of Gram-negative bacilli (87.7%), 92 strains of Gram-positive cocci (11.7%) and 5 strains of fungi (0.6%) were found; the qualified rate of microbial detection in the internal circuit of ventilator reached 71.8%, which was significantly higher than that before disinfection (P < 0.01); 2 weeks after mechanical ventilation the incidence of VAP decreased slightly during the period [20.5% (8/39) vs. 33.3% (13/39)], but there was no significant difference (P > 0.05).
CONCLUSIONS: The internal circuit of the ventilator can be used to detect the pathogen and the sputum culture of the patients on mechanical ventilation with a high consistency. The disinfection of the pathogen could significantly reduce the air pollution of the ventilator and reduce the occurrence of VAP in the patients.
参考来源
- Qiu K, Zheng Y, Gu N, et al. [Observation on the application of ventilator internal circuit disinfection in mechanically ventilated patients]. Zhonghua Wei Zhong Bing Ji Jiu Yi Xue. 2019. PMID: 31109419. doi:10.3760/cma.j.issn.2095-4352.2019.04.015
供执业医师抢救与诊疗参考,不替代临床判断;剂量请结合体重、肝肾功能和本院药品规格复核,以指南原文为准。