使用带实时反馈的智能手表可提高医护人员高质量心肺复苏的实施质量
该研究将80名急诊科医护人员随机分为干预组(佩戴预装实时视听反馈应用的智能手表)和对照组,两组均在模拟人上以30:2按压通气比进行2分钟心肺复苏。主要结局为按压频率和深度,次要结局为同时满足指南推荐频率(100-120次/分)和深度(50-60mm)的高质量按压百分比。干预组按压深度(50.9±6.6 mm)显著深于对照组(39.0±8.7 mm,P<0.001),高质量按压百分比中位数(39.4%,IQR 27.1-50.1)显著高于对照组(0.0%,IQR 0.0-0.0,P<0.001);对照组按压频率(129.1±14.9次/分)显著快于干预组(112.0±3.5次/分,P<0.001),但对照组频率超出指南推荐范围。作者结论认为,无实时反馈时按压倾向于过快过浅,智能手表实时反馈可提高心肺复苏质量。
要点
- 模拟人2分钟30:2心肺复苏中,干预组按压深度50.9±6.6 mm显著深于对照组39.0±8.7 mm(P<0.001)。
- 干预组高质量按压(频率100-120次/分且深度50-60 mm)百分比中位数39.4%(IQR 27.1-50.1),对照组为0.0%(IQR 0.0-0.0),P<0.001。
- 对照组按压频率129.1±14.9次/分显著快于干预组112.0±3.5次/分(P<0.001),但对照组频率超出指南推荐范围。
- 作者结论:无实时反馈时胸外按压倾向于过快过浅,智能手表实时反馈可提高心肺复苏质量。
适用人群与场景
- 急诊科医护人员,在模拟人上以30:2按压通气比进行2分钟心肺复苏。
研究设计与关键结果
- 研究类型为随机对照模拟人研究,共纳入80名急诊科医护人员,干预组40人、对照组40人。
- 干预组佩戴预装实时视听反馈应用的智能手表,对照组无此反馈;两组均在模拟人上以30:2按压通气比进行2分钟心肺复苏。
- 主要结局:按压频率干预组112.0±3.5次/分 vs 对照组129.1±14.9次/分,对照组显著更快(P<0.001)但超出指南推荐;按压深度干预组50.9±6.6 mm vs 对照组39.0±8.7 mm,干预组显著更深(P<0.001)。
- 次要结局:同时满足指南推荐频率(100-120次/分)和深度(50-60 mm)的高质量按压百分比,干预组中位数39.4%(IQR 27.1-50.1)vs 对照组中位数0.0%(IQR 0.0-0.0),干预组显著更高(P<0.001)。
证据类型、等级与局限
- 证据类型:随机对照模拟人研究。
- 局限:原文未报告局限性。
对床旁实践的改变
- 作者结论认为,无实时反馈时胸外按压倾向于过快过浅,智能手表实时反馈可提高心肺复苏质量。
提炼说明
- 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开
AIM: Cardiopulmonary resuscitation (CPR) quality affects survival after cardiac arrest. We aimed to investigate if a smartwatch with real-time feedback can improve CPR quality by healthcare professionals.
METHODS: An app providing real-time audiovisual feedback was developed for a smartwatch. Emergency Department (ED) professionals were recruited and randomly allocated to either the intervention group wearing a smartwatch with the preinstalled app, or to a control group. All participants were asked to perform a two-minute CPR on a manikin at a 30:2 compression-ventilation ratio. Primary outcomes were the mean CCR and CCD measured on the manikin. A secondary outcome was the percentage of chest compressions meeting both the guideline-recommended rate (100-120 min-1) and depth (50-60 mm) of high-quality CPR during a 2-min period. Differences between groups were evaluated with t-test, Chi-Square test, or Mann-Whitney U test depending on the distribution.
RESULTS: Eighty participants were recruited. 40 people were assigned to the intervention and 40 to the control group. The compression rates (mean ± SD, min-1) were significantly faster (but above the guideline recommendation, P < 0.001) in the control (129.1 ± 14.9) than in the intervention group (112.0 ± 3.5). The compression depths (mean ± SD, mm) were significantly deeper (P < 0.001) in the intervention (50.9 ± 6.6) than in the control group (39.0 ± 8.7). The percentage (%) of high-quality CPR was significantly higher (P < 0.001) in the intervention (median 39.4, IQR 27.1-50.1) than in the control group (median 0.0, IQR 0.0-0.0).
CONCLUSION: Without real-time feedback, chest compressions tend to be too fast and too shallow. CPR quality can be improved with the assistance of a smartwatch providing real-time feedback.