文献速递复苏、休克与脓毒症心脑血管急症

二氧化碳波形图:院外心脏骤停中自主循环恢复检测的支持工具

Capnography: A support tool for the detection of return of spontaneous circulation in out-of-hospital cardiac arrest.
PubMed 文献 · Resuscitation · 2019-04-18
一句话结论在院外心脏骤停中,将呼气末二氧化碳(EtCO2)加入基于心电图和胸阻抗的机器学习模型可提高脉搏检测性能,联合模型检测脉搏的AUC为0.92,回顾性检测ROSC的敏感度96.6%、特异度94.5%。

本研究分析了426例院外心脏骤停病例(117例ROSC,309例无ROSC),比较ROSC与无ROSC病例的EtCO2值,并自动提取5098个无伪差3秒片段(3639个标记为有脉性节律PR,1459个标记为无脉性电活动PEA),以临床医生现场标注的ROSC时刻为金标准。机器学习分类器使用心电图、胸阻抗和EtCO2值特征。结果显示EtCO2从ROSC前3分钟的41 mmHg升至ROSC后1分钟的57 mmHg,PR的EtCO2显著高于PEA(46 mmHg/20 mmHg,p<0.05)。将EtCO2加入模型使曲线下面积(AUC)增加超过2个百分点,心电图、胸阻抗和EtCO2联合检测脉搏的AUC为0.92。回顾性分析显示检测ROSC和无ROSC病例的敏感度和特异度分别为96.6%和94.5%。结论认为加入EtCO2可改善基于心电图和胸阻抗的自动脉搏检测算法性能,这些算法可用于现场识别脉搏和回顾性识别ROSC病例。

要点

  • 在426例院外心脏骤停病例中,EtCO2从ROSC前3分钟的41 mmHg显著升至ROSC后1分钟的57 mmHg。
  • 有脉性节律(PR)的EtCO2显著高于无脉性电活动(PEA),分别为46 mmHg和20 mmHg(p<0.05)。
  • 将EtCO2加入基于心电图和胸阻抗的机器学习模型可使AUC提高超过2个百分点,联合模型检测脉搏的AUC为0.92。
  • 回顾性分析中,联合模型检测ROSC的敏感度为96.6%,检测无ROSC病例的特异度为94.5%。
  • 作者认为加入EtCO2可改善自动脉搏检测算法性能,可用于现场识别脉搏及回顾性识别ROSC病例。

适用人群与场景

  • 院外心脏骤停病例,共426例,其中117例ROSC,309例无ROSC。

研究设计与关键结果

  • 研究类型为对院外心脏骤停病例的回顾性分析,共426例(117例ROSC,309例无ROSC),自动提取5098个无伪差3秒片段,标记为PR(3639个)或PEA(1459个),以临床医生现场标注的ROSC时刻为金标准。
  • EtCO2值从ROSC前3分钟的41 mmHg显著升高至ROSC后1分钟的57 mmHg,PR的EtCO2显著高于PEA(46 mmHg/20 mmHg,p<0.05)。
  • 将EtCO2加入机器学习模型使AUC增加超过2个百分点,心电图、胸阻抗和EtCO2联合检测脉搏的AUC为0.92。
  • 回顾性分析显示联合模型检测ROSC的敏感度为96.6%,检测无ROSC病例的特异度为94.5%。

证据类型、等级与局限

  • 证据类型:回顾性分析。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 作者认为,加入EtCO2可改善基于心电图和胸阻抗的自动脉搏检测算法性能。
  • 作者认为,这些算法可用于现场识别脉搏,并回顾性识别ROSC病例。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

BACKGROUND: Automated detection of return of spontaneous circulation (ROSC) is still an unsolved problem during cardiac arrest. Current guidelines recommend the use of capnography, but most automatic methods are based on the analysis of the ECG and thoracic impedance (TI) signals. This study analysed the added value of EtCO2 for discriminating pulsed (PR) and pulseless (PEA) rhythms and its potential to detect ROSC.

MATERIALS AND METHODS: A total of 426 out-of-hospital cardiac arrest cases, 117 with ROSC and 309 without ROSC, were analysed. First, EtCO2 values were compared for ROSC and no ROSC cases. Second, 5098 artefact free 3-s long segments were automatically extracted and labelled as PR (3639) or PEA (1459) using the instant of ROSC annotated by the clinician on scene as gold standard. Machine learning classifiers were designed using features obtained from the ECG, TI and the EtCO2 value. Third, the cases were retrospectively analysed using the classifier to discriminate cases with and without ROSC.

RESULTS: EtCO2 values increased significantly from 41 mmHg 3-min before ROSC to 57 mmHg 1-min after ROSC, and EtCO2 was significantly larger for PR than for PEA, 46 mmHg/20 mmHg (p < 0.05). Adding EtCO2 to the machine learning models increased their area under the curve (AUC) by over 2 percentage points. The combination of ECG, TI and EtCO2 had an AUC for the detection of pulse of 0.92. Finally, the retrospective analysis showed a sensitivity and specificity of 96.6% and 94.5% for the detection of ROSC and no-ROSC cases, respectively.

CONCLUSION: Adding EtCO2 improves the performance of automatic algorithms for pulse detection based on ECG and TI. These algorithms can be used to identify pulse on site, and to retrospectively identify cases with ROSC.

来源:Resuscitation,查看原文。版权归原作者所有。