文献速递呼吸衰竭与机械通气

机械通气中持续雾化气流导致触发无效的原因分析

[Reason analysis of ineffective triggering caused by a continuous nebulization airflow in mechanical ventilation].
PubMed 文献 · Zhonghua Wei Zhong Bing Ji Jiu Yi Xue · 2019-01-01
一句话结论雾化气流会增加机械通气吸气触发的难度,存在内源性呼气末正压(PEEPi)时触发更加困难。

该研究用测试肺和氧驱动喷射雾化器连接PB840呼吸机回路,在流量触发模式(VTrig 3 L/min和5 L/min)和压力触发模式(PTrig 2 cmH2O和4 cmH2O)下,分别关闭和打开雾化器,手动外拉测试肺至有效触发,观察回路中流量与压力的对应关系。结果显示,雾化时流量触发模式下肺泡压需从PEEP或PEEPi降至「PEEP-(VTrig+N)R」,压力触发模式下需降至「PEEP-PTrig-(N+1)R」,触发期间肺泡压变化幅度相应增大。结论提示雾化气流增加吸气触发难度,PEEPi使触发更加困难。

要点

  • 流量触发模式下,无雾化时肺泡压需降至「PEEP-VTrigR」方可有效触发,雾化时需降至「PEEP-(VTrig+N)R」,触发难度增加。
  • 压力触发模式下,无雾化时肺泡压需降至「PEEP-PTrig-1R」方可有效触发,雾化时需降至「PEEP-PTrig-(N+1)R」,触发难度增加。
  • 存在PEEPi时,流量触发模式下肺泡压变化幅度为「PEEPi-PEEP+(VTrig+N)R」,压力触发模式下为「PEEPi-PEEP+PTrig+(N+1)R」,触发更加困难。
  • 雾化气流(N)使触发所需的肺泡压下降幅度增大,流量触发模式下增加「N×R」,压力触发模式下增加「(N+1)×R」。

适用人群与场景

  • 接受机械通气并同时进行持续雾化吸入治疗的患者。

研究设计与关键结果

  • 研究类型为呼吸力学实验,使用测试肺和氧驱动喷射雾化器连接PB840呼吸机回路。
  • 流量触发模式(VTrig 3 L/min和5 L/min):无雾化时肺泡压从PEEP或PEEPi降至「PEEP-VTrigR」,变化幅度为「VTrigR」或「PEEPi-PEEP+VTrigR」;雾化时降至「PEEP-(VTrig+N)R」,变化幅度为「(VTrig+N)R」或「PEEPi-PEEP+(VTrig+N)R」。
  • 压力触发模式(PTrig 2 cmH2O和4 cmH2O,1 cmH2O=0.098 kPa):无雾化时肺泡压从PEEP或PEEPi降至「PEEP-PTrig-1R」,变化幅度为「PTrig+1R」或「PEEPi-PEEP+PTrig+1R」;雾化时降至「PEEP-PTrig-(N+1)R」,变化幅度为「PTrig+(N+1)R」或「PEEPi-PEEP+PTrig+(N+1)R」。

证据类型、等级与局限

  • 证据类型:呼吸力学实验研究。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 作者认为,雾化气流会增加机械通气吸气触发的难度,PEEPi使触发更加困难。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

OBJECTIVE: To analyze the ineffective triggering caused by nebulization in the way of respiratory mechanics.

METHODS: A test-lung and an oxygen-driven jet nebulizer were connected to the circuit in a PB840 ventilator. The test-lung was pulled outwards in manual way till an inspiration was effectively triggered separately in different flow-trigger modes [flow-trigger sensitivity (VTrig) 3 L/min and 5 L/min] and pressure-trigger modes [pressure-trigger sensitivity (PTrig) 2 cmH2O and 4 cmH2O, 1 cmH2O = 0.098 kPa] with the nebulizer being closed and opened in turn. The corresponding relationship and characteristics between the flow and pressure in the circuit under different triggering conditions were observed by adjusting the curve amplitude in the screen. The minimum alveolar pressure (Pa) which could cause an effective triggering and the variation span of Pa during the triggering period were analyzed in respiratory mechanics.

RESULTS: (1) In flow-trigger mode: Pa was pulled down from positive end-expiratory pressure (PEEP) or intrinsic positive end-expiratory pressure (PEEPi) to "PEEP-VTrigR" (R meant airway resistance) without nebulization, and the span of Pa was "VTrigR" or "PEEPi-PEEP+VTrigR". Pa was pulled down from PEEP or PEEPi to "PEEP-(VTrig+N)R" (N meant nebulization airflow) with nebulization, and the span of Pa was "(VTrig+N)R" or "PEEPi-PEEP+(VTrig+N)R". (2) In pressure-trigger mode: Pa was pulled down from PEEP or PEEPi to "PEEP-PTrig-1R" without nebulization, and the span of Pa was "PTrig+1R" or "PEEPi-PEEP+PTrig+1R". Pa was pulled down from PEEP or PEEPi to "PEEP-PTrig-(N+1)R" with nebulization, and the span of Pa was "PTrig+(N+1)R" or "PEEPi-PEEP+PTrig+(N+1)R".

CONCLUSIONS: Nebulization airflow increases the difficulty of inspiratory triggering in mechanical ventilation. PEEPi makes it more difficult.