文献速递复苏、休克与脓毒症心脑血管急症

胸外按压心肺复苏期间颈动脉多普勒超声血流测量的可行性研究

[Feasible study of carotid artery Doppler ultrasound blood flow measurement during chest compression cardiopulmonary resuscitation].
PubMed 文献 · Zhonghua Wei Zhong Bing Ji Jiu Yi Xue · 2019-01-01
一句话结论床旁超声可在胸外按压期间实时无创测量颈动脉血流,28/39例获得满意图像,获取率71.8%,可用于评估按压质量。

前瞻性研究纳入2016年5月至2018年11月郑州市人民医院急诊科和ICU接受心肺复苏的成年患者39例,床旁超声记录胸外按压期间右侧颈动脉血流。28例在整个按压过程中获得满意超声图像,获取率71.8%,从开始CPR到获取超声图像时间为1.1-4.9分钟,平均(2.5±1.2)分钟。28例患者CPR开始后1分钟颈动脉收缩期峰值流速(PSV)为(62.9±18.5) cm/s、舒张末期流速(EDV)为(13.9±3.5) cm/s,CPR结束前1分钟PSV为(55.4±18.4) cm/s、EDV为(12.9±3.7) cm/s,两时间点无显著差异(均P>0.05)。按压频率100-149次/min,平均(117±47)次/min,按压达标率85.7%(24/28),按压总中断时间占全部按压时间的4.4%(25.9分钟/587.2分钟)。作者认为超声测量颈总动脉血流具有实时、无创的优点,在临床工作中可行。

要点

  • 39例成人CPR患者中28例在整个胸外按压过程中获得满意颈动脉超声图像,获取率71.8%(28/39)。
  • 从开始CPR到获取超声图像时间为1.1-4.9分钟,平均(2.5±1.2)分钟。
  • 28例患者CPR开始后1分钟PSV为(62.9±18.5) cm/s、EDV为(13.9±3.5) cm/s,CPR结束前1分钟PSV为(55.4±18.4) cm/s、EDV为(12.9±3.7) cm/s,两时间点无显著差异(均P>0.05)。
  • 按压频率<100次/min或颈动脉流速下降(PSV<30 cm/s)时,应及时提醒并纠正胸外按压者。
  • 28例按压频率100-149次/min,平均(117±47)次/min,按压达标率85.7%(24/28),按压总中断时间占全部按压时间的4.4%(25.9分钟/587.2分钟)。

适用人群与场景

  • 2016年5月至2018年11月郑州市人民医院急诊科和ICU收治的接受心肺复苏的成年患者。

诊断标准与关键阈值

  • 按压达标标准:胸外按压频率100-120次/min定义为达标。
  • 按压过程中若按压频率<100次/min或颈动脉PSV<30 cm/s,应及时提醒并纠正胸外按压者。

处置流程与药物剂量

  • 使用床旁超声机记录胸外按压期间右侧颈动脉血流,于CPR开始后1分钟和CPR结束前1分钟记录颈动脉PSV和EDV。
  • 记录整个复苏期间的平均按压频率,通过超声评估按压达标率,并根据CPR期间记录的超声图像资料回顾性计算按压中断时间。

红旗征与不能漏的情况

  • 按压频率<100次/min或颈动脉PSV<30 cm/s时,应及时提醒并纠正胸外按压者。

研究设计与关键结果

  • 前瞻性研究纳入39例成人CPR患者,21例复苏成功,成功率53.8%,恢复自主循环(ROSC)时间为(10.9±5.3)分钟。
  • 从开始CPR到获取超声图像时间为1.1-4.9分钟,平均(2.5±1.2)分钟;28例在整个胸外按压过程中获得满意超声图像,获取率71.8%(28/39)。
  • 28例满意图像患者CPR开始后1分钟PSV为(62.9±18.5) cm/s、EDV为(13.9±3.5) cm/s,CPR结束前1分钟PSV为(55.4±18.4) cm/s、EDV为(12.9±3.7) cm/s,两时间点上述参数无显著差异(均P>0.05),提示整个CPR过程复苏效果满意。
  • 28例按压频率为100-149次/min,平均(117±47)次/min,按压达标率85.7%(24/28),按压总中断时间占全部按压时间的4.4%(25.9分钟/587.2分钟)。

证据类型、等级与局限

  • 证据类型:前瞻性研究。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 作者认为超声测量颈总动脉血流具有实时、无创的优点,在临床工作中可行。
  • 按压过程中若按压频率<100次/min或颈动脉PSV<30 cm/s,应及时提醒并纠正胸外按压者。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

OBJECTIVE: To determine the feasibility of ultrasound to measure blood flow on patients with chest compression cardiopulmonary resuscitation (CPR), and to find out a real-time, noninvasive hemodynamic evaluation method.

METHODS: A prospective study was conducted. All adult patients undergoing CPR admitted to Department of Emergency and intensive care unit (ICU) of Zhengzhou People's Hospital from May 2016 to November 2018 were enrolled. The blood flow over the right carotid arteries during chest compressions was recorded with a bedside ultrasound machine. The peak systolic flow velocity (PSV) and end diastolic flow velocity (EDV) of carotid artery were recorded at 1 minute after the start of CPR and 1 minute before the end of CPR. The mean compression frequency during the whole recovery period was recorded, the rate of compression reaching the standard was evaluated by ultrasound (the rate of compression 100-120 times/min was defined as up to standard), and the interruption time of compression was calculated retrospectively according to the ultrasound image data recorded during CPR.

RESULTS: Thirty-nine patients were enrolled, and 21 patients were successfully rescued, with a successful rate of 53.8%, the time of restoration of spontaneous circulation (ROSC) was (10.9±5.3) minutes. The time from CPR to retrieve an ultrasound image was 1.1-4.9 minutes, with an average of (2.5±1.2) minutes. Satisfactory ultra-sonographic images were obtained in 28 patients during the whole course of chest compression. The acquisition rate was 71.8% (28/39). In the process of compression, if the frequency of compression was less than 100 times/min or the velocity of carotid artery dropped (PSV < 30 cm/s), the chest compressors should be reminded and corrected in time. The PSV at 1 minute after CPR start of 28 patients with satisfactory ultrasound images was (62.9±18.5) cm/s, and the EDV was (13.9±3.5) cm/s, the PSV at 1 minute before the end of CPR was (55.4±18.4) cm/s, and the EDV was (12.9±3.7) cm/s. There was no significant difference in above parameters between the two time points (both P > 0.05), suggesting that satisfactory resuscitation effect was achieved in the whole process of CPR. The compression frequency of 28 patients was 100-149 times/min with an average of (117±47) times/min. The rate of compression with standard was 85.7% (24/28), and the total interruption time of compression accounted for 4.4% of all compression time (25.9 minutes/587.2 minutes).

CONCLUSIONS: Ultrasound measurement of common carotid artery blood flow during CPR has the advantage of real-time and non-invasive, and it is feasible in clinical work.

来源:Zhonghua Wei Zhong Bing Ji Jiu Yi Xue,查看原文。版权归原作者所有。