文献速递复苏、休克与脓毒症心脑血管急症

呼气末二氧化碳预测院外心脏骤停除颤成功价值

Value of capnography to predict defibrillation success in out-of-hospital cardiac arrest.
PubMed 文献 · Resuscitation · 2019-03-02
一句话结论首次除颤前1分钟中位EtCO2可预测除颤成功(AUC 0.66),联合VF波形分析可提高预测效能(AUC 0.79–0.83)。

该研究分析214例院外心脏骤停(OHCA)患者的514次除颤(其中75次为首次除颤),自动计算除颤前1分钟中位EtCO2(MEtCO2)及两个VF波形特征(AMSA、FuzzyEn),评估其单独及联合预测除颤成功的能力。首次除颤中,成功组MEtCO2显著高于失败组(31 mmHg vs 25 mmHg,p<0.05),MEtCO2预测首次除颤成功的AUC为0.66;全部除颤中该差异不显著(32 mmHg vs 29 mmHg,p>0.05),MEtCO2不是预测因子(AUC 0.54)。AMSA和FuzzyEn预测首次除颤成功的AUC分别为0.76和0.77,全部除颤均为0.75。首次除颤中联合MEtCO2后,AMSA和FuzzyEn的AUC分别提高至0.79和0.83。

要点

  • 首次除颤前1分钟MEtCO2成功组31 mmHg vs 失败组25 mmHg(p<0.05),可预测首次除颤成功(AUC 0.66)。
  • 全部除颤中MEtCO2成功组32 mmHg vs 失败组29 mmHg(p>0.05),不是预测因子(AUC 0.54)。
  • AMSA预测首次除颤成功AUC 0.76,全部除颤AUC 0.75;FuzzyEn首次AUC 0.77,全部AUC 0.75。
  • 首次除颤中联合MEtCO2使AMSA的AUC提高至0.79、FuzzyEn提高至0.83。
  • MEtCO2与VF波形特征均可从设备文件自动计算,仅需新软件即可引入现有除颤器。

适用人群与场景

  • 院外心脏骤停(OHCA)患者,共214例、514次除颤,其中75次为首次除颤。

研究设计与关键结果

  • 研究类型为对214例OHCA患者514次除颤(75次首次除颤)的设备文件分析,自动计算除颤前1分钟中位EtCO2(MEtCO2)及两个VF波形特征(AMSA、FuzzyEn)。
  • 首次除颤中,成功组MEtCO2为31 mmHg vs 失败组25 mmHg(p<0.05);全部除颤中成功组32 mmHg vs 失败组29 mmHg(p>0.05)。
  • MEtCO2预测首次除颤成功的AUC为0.66,全部除颤AUC为0.54。
  • AMSA预测首次除颤成功AUC 0.76、全部除颤AUC 0.75;FuzzyEn首次AUC 0.77、全部AUC 0.75。
  • 首次除颤中联合MEtCO2后,AMSA的AUC提高至0.79,FuzzyEn提高至0.83。

证据类型、等级与局限

  • 局限:原文未报告局限性。

对床旁实践的改变

  • 作者认为,MEtCO2与VF波形特征可从设备文件自动计算,仅需新软件即可引入现有除颤器。
  • MEtCO2仅在首次除颤中预测除颤成功,全部除颤中不是预测因子,临床解读时需区分首次与后续除颤。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

BACKGROUND AND AIM: Unsuccessful defibrillation shocks adversely affect survival from out-of-hospital cardiac arrest (OHCA). Ventricular fibrillation (VF) waveform analysis is the tool-of-choice for the non-invasive prediction of shock success, but surrogate markers of perfusion like end-tidal CO2 (EtCO2) could improve the prediction. The aim of this study was to evaluate EtCO2 as predictor of shock success, both individually and in combination with VF-waveform analysis.

MATERIALS AND METHODS: In total 514 shocks from 214 OHCA patients (75 first shocks) were analysed. For each shock three predictors of defibrillation success were automatically calculated from the device files: two VF-waveform features, amplitude spectrum area (AMSA) and fuzzy entropy (FuzzyEn), and the median EtCO2 (MEtCO2) in the minute before the shock. Sensitivity, specificity, receiver operating characteristic (ROC) curves and area under the curve (AUC) were calculated, for each predictor individually and for the combination of MEtCO2 and VF-waveform predictors. Separate analyses were done for first shocks and all shocks.

RESULTS: MEtCO2 in first shocks was significantly higher for successful than for unsuccessful shocks (31mmHg/25mmHg, p<0.05), but differences were not significant for all shocks (32mmHg/29mmHg, p>0.05). MEtCO2 predicted shock success with an AUC of 0.66 for first shocks, but was not a predictor for all shocks (AUC 0.54). AMSA and FuzzyEn presented AUCs of 0.76 and 0.77 for first shocks, and 0.75 and 0.75 for all shocks. For first shocks, adding MEtCO2 improved the AUC of AMSA and FuzzyEn to 0.79 and 0.83, respectively.

CONCLUSIONS: MEtCO2 predicted defibrillation success only for first shocks. Adding MEtCO2 to VF-waveform analysis in first shocks improved prediction of shock success. VF-waveform features and MEtCO2 were automatically calculated from the device files, so these methods could be introduced in current defibrillators adding only new software.

来源:Resuscitation,查看原文。版权归原作者所有。

参考来源

  1. Chicote B, Aramendi E, Irusta U, et al. Value of capnography to predict defibrillation success in out-of-hospital cardiac arrest. Resuscitation. 2019. PMID: 30836170. doi:10.1016/j.resuscitation.2019.02.028

供执业医师抢救与诊疗参考,不替代临床判断;剂量请结合体重、肝肾功能和本院药品规格复核,以指南原文为准。