文献速递复苏、休克与脓毒症

基于可穿戴无线设备实时反馈的新型心肺复苏测量工具的开发

Development of a novel cardiopulmonary resuscitation measurement tool using real-time feedback from wearable wireless instrumentation.
PubMed 文献 · Resuscitation · 2019-02-21
一句话结论可穿戴设备MYO通过肌电与惯性传感器实时反馈按压频率和深度(准确率97%),但实时识别错误手部位置的分类准确率仅60%,合并相似错误类别后可显著提高准确率。

该研究设计并实现了一种用于改善心肺复苏(CPR)的可穿戴训练设备MYO,集成肌电(EMG)和惯性测量单元(IMU)传感器,可检测有效CPR及四种常见错误手/臂位置(手指松弛、手在胸骨上位置过低、患者太近或太远),通过傅里叶变换和双四元数计算按压频率和深度,并采用六种机器学习分类算法识别位置错误,通过图形用户界面和音频提供反馈。在人体模型上测试并与理论值比较,分类算法在测试中准确率98%,实时对某些类别的准确率仅60%,归因于MYO校准、采样率和相似手部位置的误分类;将相似错误位置合并为更广泛类别后准确率显著提高并产生相同的CPR改善结果;频率和深度测量总体准确率97%。结论认为该系统可实现便携式实时反馈,用于培训和现场,在分类和改进CPR实施方面显示出前景。

要点

  • MYO设备集成EMG和IMU传感器,可检测有效CPR及四种常见错误手/臂位置:手指松弛、手在胸骨上位置过低、患者太近、患者太远。
  • 按压频率和深度分别通过傅里叶变换和双四元数计算,在人体模型上测试的总体准确率为97%。
  • 分类算法在测试中准确率98%,但实时对某些类别的准确率仅60%,原因包括MYO校准、采样率和相似手部位置的误分类。
  • 将相似错误位置合并为更广泛类别后,准确率显著提高,并产生相同的CPR改善结果。
  • 系统通过图形用户界面和音频提供反馈,支持便携式实时反馈,用于培训和现场。

研究设计与关键结果

  • 研究为设备开发与验证研究,在人体模型上进行CPR测试,比较实时结果与理论值。
  • 分类算法在测试中准确率98%,实时对某些类别的准确率仅60%。
  • 将相似错误位置合并为更广泛类别后准确率显著提高,并产生相同的CPR改善结果。
  • 按压频率和深度测量总体准确率97%。
  • 设备包含EMG和IMU传感器,可检测四种常见错误手/臂位置:手指松弛、手在胸骨上位置过低、患者太近、患者太远。

证据类型、等级与局限

  • 证据类型:设备开发与验证研究(体外人体模型测试)。
  • 局限:实时分类准确率低(某些类别60%),归因于MYO校准、采样率和相似手部位置的误分类;原文未报告其他局限性。

对床旁实践的改变

  • 该设备可提供便携式实时反馈,用于CPR培训和现场,作者认为在分类和改进CPR实施方面显示出前景。
  • 实时识别错误手部位置时,将相似错误位置合并为更广泛类别可提高准确率,并产生相同的CPR改善结果。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「适用人群与场景」「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

AIM: The design and implementation of a wearable training device to improve cardiopulmonary resuscitation (CPR) is presented.

METHODS: The MYO contains both Electromyography (EMG) and Inertial Measurement Unit (IMU) sensors which are used to detect effective CPR, and the four common incorrect hand and arm positions viz. relaxed fingers; hands too low on the sternum; patient too close; or patient too far. The device determines the rate and depth of compressions calculated using a Fourier transform and dual-quaternions respectively. In addition, common positional mistakes are determined using classification algorithms (six machine learning algorithms are considered and tested). Feedback via Graphical User Interface (GUI) and audio is integrated.

RESULTS: The system is tested by performing CPR on a mannequin and comparing real-time results to theoretical values. Tests show that although the classification algorithm performed well in testing (98%), in real time, it had low accuracy for certain categories (60%), which are attributable to the MYO calibration, sampling rate and misclassification of similar hand positions. Combining these similar incorrect positions into more general categories significantly improves accuracy, and produces the same improved outcome of improved CPR. The rate and depth measures have a general accuracy of 97%.

CONCLUSION: The system allows for portable, real-time feedback for use in training and in the field, and shows promise toward classifying and improving the administration of CPR.

来源:Resuscitation,查看原文。版权归原作者所有。