电池驱动输液加温器性能比较

Comparison of the performance of battery-operated fluid warmers.
PubMed 文献 · Emerg Med J · 2018-06-07
一句话结论在高流速和低温输入条件下,Warrior加温性能最佳,Buddy Lite应限用于中等输入温度和低流速,Thermal Angel受电池容量和低输出温度限制仅适用于小量输液。

该研究在体外条件下比较了四种电池驱动输液加温器(Buddy Lite、enFlow、Thermal Angel、Warrior)的加温性能和抗阻特性。使用体外循环滚压泵和冷却器,在输入温度20°C和10°C、流速50、100和200 mL/min条件下连续记录输出温度。结果发现加温器性能随流速和输入温度变化:在20°C输入、50 mL/min流速下,Buddy Lite、enFlow、Thermal Angel和Warrior分别将3.4、2.4、1和3.6 L液体加温至35°C以上;但在10°C输入、200 mL/min流速下,Buddy Lite无法加温,enFlow将3.3 L加温至25.7°C,Thermal Angel将1.5 L加温至20.9°C,Warrior将3.4 L加温至34.4°C(p<0.0001)。结论认为Warrior在高流速和低输入温度下加温性能最佳,能加温最大体积;Buddy Lite应限用于中等输入温度和低流速;Thermal Angel因电池容量和极端条件下低输出温度仅限用于小量输液。

要点

  • 在输入温度10°C、流速200 mL/min条件下,Warrior将3.4 L液体加温至34.4°C,加温性能最佳。
  • 在相同极端条件下,Buddy Lite无法加温,enFlow将3.3 L加温至25.7°C,Thermal Angel将1.5 L加温至20.9°C(p<0.0001)。
  • 在输入温度20°C、流速50 mL/min条件下,Buddy Lite、enFlow、Thermal Angel和Warrior分别将3.4、2.4、1和3.6 L液体加温至35°C以上。
  • Buddy Lite的使用应限于中等输入温度和低流速。
  • Thermal Angel因电池容量和极端条件下低输出温度,使用限于小量输液。

适用人群与场景

  • 创伤患者,尤其是需要大量输液时,需加温静脉液体以预防低体温。
  • 院前和转运环境,要求加温器小巧、节能且不依赖外部电源。

研究设计与关键结果

  • 体外研究比较四种电池驱动输液加温器(Buddy Lite、enFlow、Thermal Angel、Warrior)的加温性能和抗阻特性。
  • 使用体外循环滚压泵和冷却器,在输入温度20°C和10°C、流速50、100和200 mL/min条件下连续记录输出温度。
  • 在输入温度20°C、流速50 mL/min下,Buddy Lite、enFlow、Thermal Angel和Warrior分别将3.4、2.4、1和3.6 L液体加温至35°C以上。
  • 在输入温度10°C、流速200 mL/min下,Buddy Lite无法加温,enFlow将3.3 L加温至25.7°C,Thermal Angel将1.5 L加温至20.9°C,Warrior将3.4 L加温至34.4°C(p<0.0001)。

证据类型、等级与局限

  • 体外研究。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 在高流速和低输入温度条件下,Warrior提供最佳加温性能,能加温最大体积。
  • Buddy Lite的使用应限于中等输入温度和低流速。
  • Thermal Angel因电池容量和极端条件下低输出温度,使用限于小量输液。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

OBJECTIVES: Warming intravenous fluids is essential to prevent hypothermia in patients with trauma, especially when large volumes are administered. Prehospital and transport settings require fluid warmers to be small, energy efficient and independent of external power supply. We compared the warming properties and resistance to flow of currently available battery-operated fluid warmers.

METHODS: Fluid warming was evaluated at 50, 100 and 200 mL/min at a constant input temperature of 20°C and 10°C using a cardiopulmonary bypass roller pump and cooler. Output temperature was continuously recorded.

RESULTS: Performance of fluid warmers varied with flows and input temperatures. At an input temperature of 20°C and flow of 50 mL/min, the Buddy Lite, enFlow, Thermal Angel and Warrior warmed 3.4, 2.4, 1 and 3.6 L to over 35°C, respectively. However, at an input temperature of 10°C and flow of 200 mL/min, the Buddy Lite failed to warm, the enFlow warmed 3.3 L to 25.7°C, the Thermal Angel warmed 1.5 L to 20.9°C and the Warrior warmed 3.4 L to 34.4°C (p<0.0001).

CONCLUSION: We found significant differences between the fluid warmers: the use of the Buddy Lite should be limited to moderate input temperature and low flow rates. The use of the Thermal Angel is limited to low volumes due to battery capacity and low output temperature at extreme conditions. The Warrior provides the best warming performance at high infusion rates, as well as low input temperatures, and was able to warm the largest volumes in these conditions.

来源:Emerg Med J,查看原文。版权归原作者所有。

参考来源

  1. Lehavi A, Yitzhak A, Jarassy R, et al. Comparison of the performance of battery-operated fluid warmers. Emerg Med J. 2018. PMID: 29880720. doi:10.1136/emermed-2017-207112