文献速递心脑血管急症重症综合管理与镇静镇痛

无创脑血管自动调节评估:容积反应性指数的前瞻性研究

Non-invasive Cerebrovascular Autoregulation Assessment Using the Volumetric Reactivity Index: Prospective Study.
PubMed 文献 · Neurocrit Care · 2019-01-01
一句话结论在61例重型颅脑创伤患者中,无创超声容积反应性指数VRx与有创压力反应性指数PRx显著相关(r=0.843,P<0.001),VRx可作为无创脑血管自动调节指数使用。

这项前瞻性研究对61例重型颅脑创伤患者同时进行有创和无创脑血管自动调节监测,有创方法计算压力反应性指数PRx(颅内压与动脉血压慢波的移动相关系数),无创方法计算容积反应性指数VRx(动脉血压与无创测得颅内血容量慢波的移动相关系数)。按患者监测时段平均后,VRx与PRx的线性回归显示显著相关(r=0.843,P<0.001,95%CI 0.751–0.903),两者差值的标准差为0.192,偏倚为-0.065。作者结论认为,基于超声飞行时间测量颅内血容量动态的VRx与有创PRx指数显著一致,VRx可像PRx一样作为无创脑血管自动调节指数使用,并能提供涵盖全部颅内血流动力学的自动调节状态信息。

要点

  • 在61例重型颅脑创伤患者中同时监测VRx与PRx,按患者监测时段平均后两者显著相关(r=0.843,P<0.001,95%CI 0.751–0.903)。
  • VRx为动脉血压与无创测得颅内血容量慢波的移动相关系数,PRx为颅内压与动脉血压慢波的移动相关系数。
  • VRx与PRx差值的标准差为0.192,偏倚为-0.065。
  • 作者认为VRx可像PRx一样作为无创脑血管自动调节指数使用,并可提供涵盖全部颅内血流动力学的自动调节状态信息。
  • 超声飞行时间法反映的是声学路径内(跨越双侧大脑半球)的血容量变化,而非局部有创记录的颅内压慢波。

适用人群与场景

  • 61例重型颅脑创伤的ICU患者,同时接受有创和无创脑血管自动调节监测。

诊断标准与关键阈值

  • PRx为颅内压与动脉血压慢波之间的移动相关系数(有创计算)。
  • VRx为动脉血压与无创测得的颅内血容量慢波之间的移动相关系数。

研究设计与关键结果

  • 前瞻性比较临床研究,61例重型颅脑创伤患者同时进行有创PRx和无创VRx监测。
  • 按患者监测时段平均后,VRx与PRx线性回归显示显著相关:r=0.843,P<0.001,95%CI 0.751–0.903。
  • VRx与PRx差值的标准差为0.192,偏倚为-0.065。

证据类型、等级与局限

  • 证据类型:前瞻性比较临床研究。

对床旁实践的改变

  • 作者认为,基于超声飞行时间测量颅内血容量动态的VRx可像PRx一样作为无创脑血管自动调节指数使用,并能提供涵盖全部颅内血流动力学的自动调节状态信息。
  • 超声飞行时间法反映声学路径内(跨越双侧大脑半球)的血容量变化,不反映局部有创记录的颅内压慢波,而是反映双侧大脑半球的自动调节反应。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

BACKGROUND: This prospective study of an innovative non-invasive ultrasonic cerebrovascular autoregulation (CA) monitoring method is based on real-time measurements of intracranial blood volume (IBV) reactions following changes in arterial blood pressure. In this study, we aimed to determine the clinical applicability of a non-invasive CA monitoring method by performing a prospective comparative clinical study of simultaneous invasive and non-invasive CA monitoring on intensive care patients.

METHODS: CA was monitored in 61 patients with severe traumatic brain injuries invasively by calculating the pressure reactivity index (PRx) and non-invasively by calculating the volumetric reactivity index (VRx) simultaneously. The PRx was calculated as a moving correlation coefficient between intracranial pressure and arterial blood pressure slow waves. The VRx was calculated as a moving correlation coefficient between arterial blood pressure and non-invasively-measured IBV slow waves.

RESULTS: A linear regression between VRx and PRx averaged per patients' monitoring session showed a significant correlation (r = 0.843, p < 0.001; 95% confidence interval 0.751 - 0.903). The standard deviation of the difference between VRx and PRx was 0.192; bias was - 0.065.

CONCLUSIONS: This prospective clinical study of the non-invasive ultrasonic volumetric reactivity index VRx monitoring, based on ultrasonic time-of-flight measurements of IBV dynamics, showed significant coincidence of non-invasive VRx index with invasive PRx index. The ultrasonic time-of-flight method reflects blood volume changes inside the acoustic path, which crosses both hemispheres of the brain. This method does not reflect locally and invasively-recorded intracranial pressure slow waves, but the autoregulatory reactions of both hemispheres of the brain. Therefore, VRx can be used as a non-invasive cerebrovascular autoregulation index in the same way as PRx and can also provide information about the CA status encompassing all intracranial hemodynamics.

来源:Neurocrit Care,查看原文。版权归原作者所有。

参考来源

  1. Petkus V, Preiksaitis A, Krakauskaite S, et al. Non-invasive Cerebrovascular Autoregulation Assessment Using the Volumetric Reactivity Index: Prospective Study. Neurocrit Care. 2019. PMID: 29951960. doi:10.1007/s12028-018-0569-x

供执业医师抢救与诊疗参考,不替代临床判断;剂量请结合体重、肝肾功能和本院药品规格复核,以指南原文为准。