文献速递心脑血管急症重症综合管理与镇静镇痛

双频指数监测仪衍生的简化脑电图在心脏骤停后患者中的验证

The validation of simplified EEG derived from the bispectral index monitor in post-cardiac arrest patients.
PubMed 文献 · Resuscitation · 2018-02-01
一句话结论回顾性验证显示,BIS监测仪衍生的简化脑电图与标准脑电图分类一致性较高,无脑电图经验者也能识别癫痫持续状态。

该研究回顾性验证了心脏骤停后患者中双频指数(BIS)监测仪衍生的简化脑电图的准确性。成功复苏的心脏骤停患者转运至导管室,必要时行经皮冠状动脉介入治疗,抵达冠脉监护病房后启动双侧BIS监测并持续至72小时。在标准脑电图记录的同时点提取BIS简化脑电图,由两名神经电生理医生判读弥漫性慢波、爆发抑制、脑电无活动、周期性癫痫样放电和癫痫持续状态,另由两名无经验研究者仅判读癫痫持续状态。共分析32份简化BIS脑电图样本,与标准脑电图相比,神经电生理医生判读的敏感度为86%、特异度为100%、观察者间变异为0.843;两名无经验医生识别癫痫持续状态的敏感度为80%、特异度为94%。作者结论认为简化BIS脑电图与标准脑电图监测相当,未来需进一步研究确认并明确其预后和治疗价值。

要点

  • 心脏骤停后患者中,神经电生理医生用简化BIS脑电图判读多种脑电模式,与标准脑电图相比敏感度86%、特异度100%、观察者间变异0.843。
  • 两名无脑电图经验医生仅判读癫痫持续状态,敏感度80%、特异度94%。
  • 共分析32份简化BIS脑电图样本,在标准脑电图记录的同时点从BIS监测仪提取。
  • 作者认为简化BIS脑电图与标准脑电图监测相当,但未来研究需确认结果并确定其预后和治疗价值。

适用人群与场景

  • 成功复苏的心脏骤停患者,转运至导管室,必要时行经皮冠状动脉介入治疗,抵达冠脉监护病房后启动双侧BIS监测并持续至72小时。

处置流程与药物剂量

  • 成功复苏的心脏骤停患者转运至导管室,必要时行经皮冠状动脉介入治疗。
  • 抵达冠脉监护病房后启动双侧BIS监测,持续至72小时。
  • 在标准脑电图记录的同时点,从BIS监测仪提取原始简化脑电图描记。

研究设计与关键结果

  • 回顾性验证研究,共分析32份简化BIS脑电图样本。
  • 与标准脑电图相比,神经电生理医生判读简化脑电图(弥漫性慢波、爆发抑制、脑电无活动、周期性癫痫样放电、癫痫持续状态)的敏感度为86%、特异度为100%、观察者间变异为0.843。
  • 两名无经验医生识别癫痫持续状态的敏感度为80%、特异度为94%。

证据类型、等级与局限

  • 证据类型:回顾性验证研究。
  • 作者结论:使用简单分类系统,BIS监测仪衍生的原始简化脑电图与标准脑电图监测相当;无脑电图经验的研究者能够识别心脏骤停后患者的癫痫持续状态。
  • 局限:未来研究需确认这些结果,并确定使用简化BIS脑电图在预后和治疗意义方面的附加价值。

对床旁实践的改变

  • 作者认为,使用简单分类系统,BIS监测仪衍生的原始简化脑电图与标准脑电图监测相当。
  • 作者认为,无脑电图经验的研究者能够识别心脏骤停后患者的癫痫持续状态。
  • 作者提出,未来研究需确认这些结果并确定简化BIS脑电图在预后和治疗意义方面的附加价值。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

AIMS: We aimed to validate retrospectively the accuracy of simplified electroencephalography (EEG) monitoring derived from the bispectral index (BIS) monitor in post-cardiac arrest (CA) patients.

METHODS: Successfully resuscitated CA patients were transferred to the Catherization Lab followed by percutaneous coronary intervention when indicated. On arrival at the coronary care unit, bilateral BIS monitoring was started and continued up to 72 h. Raw simplified EEG tracings were extracted from the BIS monitor at a time point coinciding with the registration of standard EEG monitoring. BIS EEG tracings were reviewed by two neurophysiologists, who were asked to indicate the presence of following patterns: diffuse slowing rhythm, burst suppression pattern, cerebral inactivity, periodic epileptiform discharges and status epilepticus (SE). Additionally, these simplified BIS EEG tracings were analysed by two inexperienced investigators, who were asked to indicate the presence of SE only.

RESULTS: Thirty-two simplified BIS EEG samples were analysed. Compared to standard EEG, neurophysiologists interpreted all simplified EEG samples with a sensitivity of 86%, a specificity of 100% and an interobserver variability of 0.843. Furthermore, SE was identified with a sensitivity of 80% and a specificity of 94% by two unexperienced physicians.

CONCLUSION: Using a simple classification system, raw simplified EEG derived from a BIS monitoring device is comparable to standard EEG monitoring. Moreover, investigators without EEG experience were capable to identify SE in post-CA patients. Future studies will be warranted to confirm our results and to determine the added value of using simplified BIS EEG in terms of prognostic and therapeutic implications.

来源:Resuscitation,查看原文。版权归原作者所有。