智能手表反馈装置在婴儿心脏骤停期间帮助单一施救者进行高质量胸外按压:一项随机对照模拟研究

Smartwatch feedback device for high-quality chest compressions by a single rescuer during infant cardiac arrest: a randomized, controlled simulation study.
PubMed 文献 · Eur J Emerg Med · 2019-01-01
一句话结论模拟婴儿心肺复苏中,智能手表反馈可提高单一施救者胸外按压正确深度比例(99% vs 83%),并加深按压深度(41.2 mm vs 38.6 mm)。

该随机对照模拟研究纳入30名参与者,随机分为使用或不使用智能手表胸外按压反馈应用两组,在婴儿模型上连续进行2分钟仅按压心肺复苏。反馈组正确深度比例中位数99%(IQR 97-100),显著高于无反馈组的83%(IQR 58-97),P=0.002;反馈组按压深度41.2 mm(IQR 39.8-41.7)也显著深于对照组的38.6 mm(IQR 36.1-39.6),P=0.004。两组按压频率和正确回弹深度比例无显著差异。作者认为,接受智能手表胸外按压参数反馈的施救者可在婴儿心肺复苏中实施 adequate 胸外按压。

要点

  • 模拟婴儿心肺复苏中,使用智能手表反馈的施救者正确按压深度比例中位数99%(IQR 97-100),无反馈组为83%(IQR 58-97),P=0.002。
  • 反馈组胸外按压深度为41.2 mm(IQR 39.8-41.7),显著深于无反馈组的38.6 mm(IQR 36.1-39.6),P=0.004。
  • 两组按压频率无显著差异(P=0.482),正确回弹深度比例也无显著差异(P=0.089)。
  • 研究为模拟研究,30名参与者随机分组,在婴儿模型上连续进行2分钟仅按压心肺复苏。

适用人群与场景

  • 参与模拟婴儿心肺复苏的单一施救者,在婴儿模型上实施连续2分钟仅按压心肺复苏。

研究设计与关键结果

  • 随机对照模拟研究,30名参与者随机分为智能手表反馈组和无反馈对照组,在婴儿模型上连续进行2分钟仅按压心肺复苏。
  • 反馈组正确按压深度比例中位数99%(IQR 97-100),对照组83%(IQR 58-97),P=0.002。
  • 反馈组按压深度41.2 mm(IQR 39.8-41.7),对照组38.6 mm(IQR 36.1-39.6),P=0.004。
  • 两组按压频率无显著差异(P=0.482),正确回弹深度比例无显著差异(P=0.089)。

证据类型、等级与局限

  • 证据类型:随机对照模拟研究。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 在婴儿心肺复苏模拟训练中,智能手表反馈可能帮助单一施救者达到更深的胸外按压深度和更高的正确深度比例。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

OBJECTIVE: According to the guidelines, rescuers should provide chest compressions (CC) ~1.5 inches (40 mm) for infants. Feedback devices could help rescuers perform CC with adequate rates (CCR) and depths (CCD). However, there is no CC feedback device for infant cardiopulmonary resuscitation (CPR). We suggest a smartwatch-based CC feedback application for infant CPR.

PARTICIPANTS AND METHODS: We created a smartwatch-based CC feedback application. This application provides feedback on CCD and CCR by colour and text for infant CPR. To evaluate the application, 30 participants were divided randomly into two groups on the basis of whether CC was performed with or without the assistance of the smartwatch application. Both groups performed continuous CC-only CPR for 2 min on an infant mannequin placed on a firm table. We collected CC parameters from the mannequin, including the proportion of correct depth, CCR, CCD and the proportion of correct decompression depth.

RESULTS: Demographics between the two groups were not significantly different. The median (interquartile range) proportion of correct depth was 99 (97-100) with feedback compared with 83 (58-97) without feedback (P = 0.002). The CCR and proportion of correct decompression depth were not significantly different between the two groups (P = 0.482 and 0.089). The CCD of the feedback group was significantly deeper than that of the control group [feedback vs. control: 41.2 (39.8-41.7) mm vs. 38.6 (36.1-39.6) mm; P=0.004].

CONCLUSION: Rescuers who receive feedback of CC parameters from a smartwatch could perform adequate CC during infant CPR.

来源:Eur J Emerg Med,查看原文。版权归原作者所有。