颈动脉多普勒超声与二氧化碳波形图在评估心肺复苏有效性中的比较
本研究在急诊科纳入非创伤性心脏骤停或急诊随访期间发生骤停的患者,在CPR周期第100次胸外按压后测量颈动脉收缩期峰值流速(PSV)、舒张末期流速(EDV)、时间依赖平均血流速度(MNV)及呼气末二氧化碳(ETCO2)。死亡组平均ETCO2低于ROSC组(p<0.05)。ETCO2与PSV呈正相关但程度低,与EDV、MNV呈正相关但程度极低,均无统计学意义(p>0.05)。作者结论认为颈动脉血流测量在展示CPR有效性方面不如ETCO2有价值。
要点
- 在CPR第100次胸外按压后测量颈动脉PSV 67.1±17.3、EDV 16.3±4.5、MNV 25.5±8.1及ETCO2 22.2±8.1。
- 死亡组平均ETCO2值低于ROSC组(p<0.05)。
- ETCO2与PSV呈正相关但程度低,与EDV、MNV呈正相关但程度极低,均无统计学意义(p>0.05)。
- 作者认为颈动脉血流测量结果在展示CPR有效性方面不如ETCO2有价值。
- 院内与院外骤停在患者结局(ROSC与死亡)方面差异有统计学意义(p<0.05)。
适用人群与场景
- 就诊于急诊科的非创伤性心脏骤停患者,或在急诊服务随访期间开始发生骤停的患者。
研究设计与关键结果
- 研究类型为在急诊科开展的研究,患者平均年龄54.5±12.3岁,65.6%为男性。
- 在CPR周期第100次胸外按压后测量并记录颈动脉PSV、EDV、时间依赖MNV及ETCO2值。
- 患者CPR期间颈动脉测量均值:PSV 67.1±17.3,EDV 16.3±4.5,MNV 25.5±8.1,ETCO2 22.2±8.1。
- 院内与院外骤停在患者结局(ROSC与死亡)方面差异有统计学意义(p<0.05)。
- 死亡组平均ETCO2值低于ROSC组(p<0.05)。
- 全部患者中ETCO2与PSV呈正相关但程度低,EDV与MNV呈正相关但程度极低,均无统计学意义(p>0.05)。
证据类型、等级与局限
- 证据类型:研究论文(摘要)。
- 局限:原文未报告局限性。
对床旁实践的改变
- 作者认为颈动脉血流测量结果在展示CPR有效性方面不如ETCO2有价值。
提炼说明
- 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开
INTRODUCTION AND PURPOSE: The end-tidal carbon dioxide (ETCO2) measurement is accepted as the gold standard method for assessing cardiopulmonary resuscitation (CPR) efficacy. In recent studies, the use of Carotid Doppler Ultrasonography has become widespread in showing CPR efficacy. In the present study, the carotid blood flow measurement was compared with ETCO2 measurement and an evaluation was made of whether this method could be used as an alternative method to capnography in the assessment of CPR efficacy.
MATERIAL AND METHOD: This study was conducted on patients who presented at the Emergency Department (ED) with non-traumatic arrest or began to suffer from arrest during emergency service follow-up. The main carotid artery peak systolic velocity (PSV), end diastolic velocity (EDV) and time-dependent mean flow velocity (MNV), and ETCO2 values were measured and recorded after the 100th chest pressure of the CPR cycle and the results were statistically analyzed.
RESULTS: The mean age of the patients was 54.5±12.3years and 65.6% of the patients were male. The mean values of patients measured from the carotid artery during the CPR were PSV 67.1±17.3, EDV 16.3±4.5, MNV 25.5±8.1 and ETCO2 22.2±8.1. A significant difference was found between in-hospital and out-of-hospital arrests in terms of patient outcome (return of spontaneous circulation (ROSC) and death) (p<0.05). The mean ETCO2 values of those who died were found to be lower than those of the ROSC group (p<0.05). Although there was a positive and low-level of correlation between the ETCO2 values and PSV values, and a positive and very low-level of correlation between the EDV and MNV values of all patients, these correlations were not statistically significant. (p>0.05).
CONCLUSION: A low correlation was found between the PSV and ETCO2 values. With effective CPR, the results close to carotid blood flow in normal healthy individuals were obtained. However, the study showed that carotid blood flow measurement results during CPR were not as valuable as ETCO2 in demonstrating CPR efficacy.