急诊危重患者床旁即时检验的准确性
本研究为前瞻性观察性研究,纳入50例急诊危重患者,比较护士使用手持床旁即时检验设备与核心实验室对8项生化及红细胞压积的检测结果。Pearson相关与Lin一致性系数为0.84–1.00,各指标配对差值均值较小,仅3/400次测量超出基于医师调查确定的最大临床可接受偏差。作者结论认为急诊护士床旁即时检验可靠、准确,与培训技师的核心实验室检测无显著偏差。
要点
- 8项分析物(Na+、K+、Cl-、TCO2、葡萄糖、BUN、肌酐、红细胞压积)的Pearson相关与Lin一致性系数均在0.84–1.00,相关性优秀。
- 配对差值均值(95%CI):Na+ 0.30(-0.22至0.82)mmol/L,K+ -0.12(-0.14至-0.09)mmol/L,Cl- 2.10(1.41至2.78)mmol/L,TCO2 -1.68(-2.06至-1.30)mmol/L。
- 葡萄糖差值均值2.46(1.46至3.46)mg/dL,BUN 1.69(0.95至2.42)mg/dL,肌酐0.13(0.08至0.17)mg/dL,红细胞压积-0.39(-0.93至0.15)%。
- 400次测量中仅3次护士床旁即时检验与核心实验室差值超出基于医师调查的最大临床可接受偏差。
- 作者结论:急诊护士床旁即时检验可靠、准确,与培训技师的核心实验室检测无显著偏差。
适用人群与场景
- 急诊危重患者,共50例,采用便利抽样,均开具了床旁生化及红细胞压积检验。
- 平均年龄67岁,50%为男性,82%被收治入院。
研究设计与关键结果
- 研究类型:前瞻性观察性研究,便利抽样,单中心未明确,样本量50例急诊危重患者。
- 干预与对照:同一血样分两份,一份由当班护士用手持床旁即时检验设备检测,另一份送核心实验室由培训技师检测。
- 主要终点:8项分析物的Pearson相关系数与Lin一致性系数均为0.84–1.00,相关性优秀。
- 配对差值均值(95%CI):Na+ 0.30(-0.22至0.82)mmol/L,K+ -0.12(-0.14至-0.09)mmol/L,Cl- 2.10(1.41至2.78)mmol/L,TCO2 -1.68(-2.06至-1.30)mmol/L,葡萄糖2.46(1.46至3.46)mg/dL,BUN 1.69(0.95至2.42)mg/dL,肌酐0.13(0.08至0.17)mg/dL,红细胞压积-0.39(-0.93至0.15)%。
- 400次测量中3次护士床旁即时检验与核心实验室差值超出基于医师调查的最大临床可接受偏差。
证据类型、等级与局限
- 证据类型:前瞻性观察性研究。
- 局限:原文未报告局限性。
对床旁实践的改变
- 作者结论认为急诊护士使用手持设备行床旁即时检验可靠、准确,与培训技师的核心实验室检测无显著偏差。
提炼说明
- 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开
BACKGROUND: Point-of-care (POC) testing reduces laboratory turn-around having the potential to improve timely diagnosis and management. We compared the accuracy of nurse performed POC and core laboratory testing and determined whether deviations between the two were clinically meaningful.
METHODS: We performed a prospective, observational study on a convenience sample of 50 critical care ED patients in whom a POC chemistry and hematocrit was ordered. Blood samples were divided into 2 aliquots; one sample was tested by the treating nurse using a handheld POC device and the other sample was tested in the core laboratory. Paired comparisons of test results were performed using Pearson's correlation coefficients, Lin concordance coefficients, and Bland Altman plots.
RESULTS: Mean patient age was 67, 50% were male, 82% were admitted. Pearson's correlation and Lin concordance coefficients were excellent (0.84-1.00) for all 8 analytes. Mean (95%CI) paired differences between POC and core laboratory measurements were Na+ 0.30 (-0.22 to 0.82) mmol/L, K+-0.12 (-0.14 to - 0.09) mmol/L, Cl- 2.10 (1.41 to 2.78) mmol/L, TCO2-1.68 (-2.06 to -1.30) mmol/L, glucose 2.46 (1.46 to 3.46) mg/dL, BUN, 1.69 (0.95 to 2.42) mg/dL, creatinine 0.13 (0.08 to 0.17) mg/dL, and hematocrit -0.39 (-0.93 to 0.15) %. In 3 of 400 measurements, the difference between POC and core lab exceeded the maximal clinically acceptable deviation based on physician surveys.
CONCLUSIONS: Bedside POC by ED nurses is reliable and accurate and does not deviate significantly from core laboratory testing by trained technicians.