注射试验在确认食管球囊导管理想位置中的应用
该前瞻性研究纳入2015年5月至2017年3月首都医科大学附属北京天坛医院综合ICU接受有创机械通气的患者,对市售食管球囊导管进行改良以实施注射试验,逐步外撤导管直至出现干扰波,将该位置命名为0 cm,并在+15、+10、+5、0、-5、-10、-15 cm处分别行呼气末阻断,测量自主呼吸与被动通气下食管压(Pes)与气道压(Paw)变化比值(ΔPes/ΔPaw)。共纳入20例患者,15例完成自主呼吸与被动通气两部分,2例仅完成自主呼吸部分,3例仅完成被动通气部分。所有患者均可通过注射试验诱发出干扰波,研究期间无不良事件发生。自主呼吸下ΔPes/ΔPaw在0、-5、-10 cm处处于理想范围(0.8-1.2),以0 cm处最接近1(0.98±0.15);被动通气下ΔPes/ΔPaw在-5 cm和-10 cm处处于理想范围,以-10 cm处最接近1(0.94±0.12)。作者认为注射试验可快速、简便地确定食管球囊导管的理想位置,该方法安全且临床可行。
要点
- 注射试验在所有纳入患者中均可诱发出干扰波,研究期间无不良事件发生,作者认为该方法安全且临床可行。
- 自主呼吸下,ΔPes/ΔPaw在0、-5、-10 cm处处于理想范围(0.8-1.2),0 cm处最接近1(0.98±0.15),-15 cm处(0.66±0.26)显著低于0 cm处(P<0.05)。
- 被动通气下,ΔPes/ΔPaw在-5 cm和-10 cm处处于理想范围,-10 cm处最接近1(0.94±0.12);0 cm和-5 cm处(分别为1.43±0.31和1.12±0.14)显著高于-10 cm处(均P<0.01),-15 cm处(0.68±0.23)显著低于-10 cm处(P<0.01)。
- 无论自主呼吸还是被动通气,Pes均随食管导管逐步外撤而升高,在+5 cm处达最大值,继续外撤后下降。
- 自主呼吸与被动通气下干扰波平均深度分别为42.4±3.8 cm与41.8±3.3 cm,两者无显著差异(P=0.132)。
适用人群与场景
- 2015年5月至2017年3月首都医科大学附属北京天坛医院综合ICU收治的接受有创机械通气(MV)的患者。
诊断标准与关键阈值
- 以注射试验诱发出干扰波的位置为0 cm,提示球囊刚进入食管。
- ΔPes/ΔPaw理想范围为0.8-1.2。
处置流程与药物剂量
- 对市售食管球囊导管进行改良以实施注射试验。
- 逐步外撤导管并重复注射试验,直至出现干扰波,该位置命名为0 cm。
- 在+15、+10、+5、0、-5、-10和-15 cm位置分别行呼气末阻断,测量Pes和Paw变化并计算ΔPes/ΔPaw。
- 自主呼吸下理想位置为0、-5、-10 cm,以0 cm处ΔPes/ΔPaw最接近1(0.98±0.15)。
- 被动通气下理想位置为-5 cm和-10 cm,以-10 cm处ΔPes/ΔPaw最接近1(0.94±0.12)。
研究设计与关键结果
- 前瞻性研究,共纳入20例患者,15例完成自主呼吸与被动通气两部分,2例仅完成自主呼吸部分,3例仅完成被动通气部分。
- 所有患者均可通过注射试验诱发出干扰波,研究期间无不良事件发生。
- 自主呼吸下干扰波平均深度42.4±3.8 cm,被动通气下41.8±3.3 cm,两者无显著差异(P=0.132)。
- 自主呼吸下ΔPes/ΔPaw在0、-5、-10 cm处处于理想范围(0.8-1.2),0 cm处最接近1(0.98±0.15);-15 cm处(0.66±0.26)显著低于0 cm处(P<0.05)。
- 被动通气下ΔPes/ΔPaw在-5 cm和-10 cm处处于理想范围,-10 cm处最接近1(0.94±0.12);0 cm和-5 cm处(分别为1.43±0.31和1.12±0.14)显著高于-10 cm处(均P<0.01),-15 cm处(0.68±0.23)显著低于-10 cm处(P<0.01)。
- 无论自主呼吸还是被动通气,Pes均随食管导管逐步外撤而升高,在+5 cm处达最大值,继续外撤后下降。
证据类型、等级与局限
- 证据类型:前瞻性研究。
- 临床试验注册号:NCT02446938。
- 作者结论:注射试验可快速、简便地确定食管球囊导管的理想位置,该方法安全且临床可行。
- 局限:原文未报告局限性。
对床旁实践的改变
- 作者认为注射试验可快速、简便地确定食管球囊导管的理想位置,该方法安全且临床可行。
- 自主呼吸患者应用注射试验定位时,0 cm处ΔPes/ΔPaw最接近1(0.98±0.15),-15 cm处(0.66±0.26)显著低于0 cm处,不宜作为理想位置。
- 被动通气患者应用注射试验定位时,-10 cm处ΔPes/ΔPaw最接近1(0.94±0.12),0 cm和-5 cm处(分别为1.43±0.31和1.12±0.14)显著高于-10 cm处,不宜作为理想位置。
提炼说明
- 本条依据论文摘要整理;摘要未涉及「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开
OBJECTIVE: To evaluate the safety and feasibility of injection test which is used to locate esophageal balloon catheter.
METHODS: A prospective study was conducted. The patients undergoing invasive mechanical ventilation (MV) admitted to general intensive care unit (ICU) of Beijing Tiantan Hospital Affiliated to Capital Medical University from May 2015 and March 2017 were enrolled. The commercially available esophageal balloon catheter was modified to perform injection test. The catheter was withdrawn step by step and the injection test was repeated until the presence disturbance wave presented, which indicated that the balloon had just entered the esophagus. The position where disturbance wave appears was named 0 cm. End-expiratory occlusions were performed at the positions of +15, +10, +5, 0, -5, -10 and -15 cm, respectively, and the changes of esophageal pressure (Pes) and airway pressures (Paw) were measured in the spontaneous breathing and passive ventilation, and the ratio between the changes (ΔPes/ΔPaw) was calculated.
RESULTS: A total of 20 patients were enrolled, of which 15 patients finished both the spontaneous and the passive ventilation parts, and 2 patients finished only the spontaneous part and 3 patients finished only passive part. (1) Disturbance waves could be induced by injection test in all patients. The average depth of disturbance wave in spontaneous breathing was deeper than that in passive ventilation (cm: 42.4±3.8 vs. 41.8±3.3), but there was no significant difference between the two ventilation settings (P = 0.132). No adverse events occurred during the study period. (2) Pes increased with the stepwise withdraw of esophageal catheter, reached the maximal value at +5 cm, and then decreased when the catheter was further withdrawn, no matter in the spontaneous or the passive ventilation. In spontaneous breathing, the ΔPes/ΔPaw was within the ideal range (0.8-1.2) at the positions of 0, -5 and -10 cm. The ΔPes/ΔPaw was closest to unity at the positions of 0 cm (0.98±0.15). The ΔPes/ΔPaw at -15 cm (0.66±0.26) was significantly lower than that at 0 cm (P < 0.05). For passive ventilation, the ΔPes/ΔPaw was within the ideal range at the positions of -5 cm and -10 cm, and the ΔPes/ΔPaw was closest to unity at the positions of -10 cm (0.94±0.12). The ΔPes/ΔPaw at 0 cm and -5 cm was significantly higher than that at -10 cm (1.43±0.31 and 1.12±0.14, respectively); while the ΔPes/ΔPaw at -15 cm (0.68±0.23) was significantly lower than that at -10 cm (all P < 0.01).
CONCLUSIONS: Ideal position of the esophageal balloon catheter could be determined quickly and easily by using injection test. The method is safe and clinically feasible.
CLINICAL TRIAL REGISTRATION: Clinical Trials, NCT02446938.