创伤低血红蛋白患者无创血红蛋白监测的评估
本研究纳入血红蛋白低于8 g/dL并计划手术的创伤患者,入院时及每输注一单位血后同时用Radical-7 Masimo设备无创测定Sp-Hb与实验室测定Lab-Hb。共获得184份时间匹配样本,Bland-Altman分析显示Sp-Hb与Lab-Hb平均偏差0.12 g/dL,一致性界限-0.56至0.79 g/dL;Pearson相关系数0.872;每单位输血后血红蛋白变化值Delta-Sp-Hb与Delta-Lab-Hb平均偏差-0.05,一致性界限-0.62至0.51。作者结论认为Sp-Hb在低血红蛋白创伤患者的绝对值与趋势值上均显示准确精度。
要点
- 在血红蛋白低于8 g/dL的创伤手术患者中,Sp-Hb与Lab-Hb平均偏差仅0.12 g/dL,一致性界限-0.56至0.79 g/dL。
- Sp-Hb与Lab-Hb的Pearson相关系数为0.872,原文报告为良好相关。
- 每输注一单位血后血红蛋白变化值Delta-Sp-Hb与Delta-Lab-Hb平均偏差-0.05,一致性界限-0.62至0.51。
- 共分析184份时间匹配样本,采血点为入院时及每输注一单位血后。
- 作者结论:Sp-Hb在低血红蛋白创伤患者的绝对值与趋势值上均显示准确精度。
适用人群与场景
- 血红蛋白低于8 g/dL、计划接受手术干预的创伤患者。
诊断标准与关键阈值
- 低血红蛋白定义为血红蛋白低于8 g/dL。
研究设计与关键结果
- 研究纳入血红蛋白低于8 g/dL并计划手术的创伤患者,入院时及每输注一单位血后同时用Radical-7 Masimo设备测定Sp-Hb与实验室测定Lab-Hb。
- 共184份时间匹配样本用于最终分析。
- Bland-Altman分析显示Sp-Hb与Lab-Hb平均偏差0.12 g/dL,一致性界限-0.56至0.79 g/dL。
- Sp-Hb与Lab-Hb的Pearson相关系数为0.872。
- Delta-Sp-Hb与Delta-Lab-Hb平均偏差-0.05,一致性界限-0.62至0.51。
证据类型、等级与局限
- 局限:原文未报告局限性。
对床旁实践的改变
- 作者结论认为Sp-Hb在低血红蛋白创伤患者的绝对值与趋势值上均显示准确精度。
提炼说明
- 本条依据论文摘要整理;摘要未涉及「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开
OBJECTIVE: Bleeding is a leading cause of death among trauma patients. Delayed assessment of blood hemoglobin level might result in either unnecessary blood transfusion in nonindicated patients or delayed blood transfusion in critically bleeding patients. In this study, we evaluate the precision of noninvasive hemoglobin monitoring in trauma patients with low hemoglobin levels.
METHODS: We included trauma patients with low hemoglobin levels (less than 8 g/dL) scheduled for surgical intervention. Blood samples were obtained on admission and after each blood unit with concomitant measurement of serum hemoglobin using radical-7 Masimo device. The change in blood hemoglobin after every transfused blood unit was also assessed by both methods (change in noninvasive Masimo hemoglobin [Delta-Sp-Hb] and change in laboratory hemoglobin [Delta-Lab-Hb]). The precision of Masimo hemoglobin level (Sp-Hb) compared with Laboratory hemoglobin level (Lab-Hb) was determined using both Bland-Altman and Pearson correlation analyses.
RESULTS: One hundred eighty-four time-matched samples were available for final analysis. Bland-Altman analysis showed excellent accuracy of Sp-Hb compared with Lab-Hb with mean bias of 0.12 g/dL and limits of agreement between -0.56 g/dL and 0.79 g/dL. Excellent correlation was reported between both measures with Pearson correlation coefficient of 0.872. Excellent agreement was also reported between both Delta-Sp-Hb and Delta-Lab-Hb with mean bias of -0.05 and limits of agreement from -0.62 to 0.51 CONCLUSIONS:: Sp-Hb showed accurate precision in both absolute values and trend values compared with Lab-Hb measurement in trauma patients with low hemoglobin levels.
参考来源
- Gamal M, Abdelhamid B, Zakaria D, et al. Evaluation of Noninvasive Hemoglobin Monitoring in Trauma Patients with Low Hemoglobin Levels. Shock. 2018. PMID: 28727608. doi:10.1097/SHK.0000000000000949
供执业医师抢救与诊疗参考,不替代临床判断;剂量请结合体重、肝肾功能和本院药品规格复核,以指南原文为准。