文献速递重症综合管理与镇静镇痛消化、代谢与肾脏急症

体外膜氧合联合肾脏替代治疗连接方式比较:范围综述

Comparison of the approaches to combine Renal Replacement Therapy with Extracorporeal Membrane Oxygenation - a Scoping Review.
PubMed 文献 · Ann Intensive Care · 2026-09-21
一句话结论ECMO-RRT不同连接方式在效率与临床结局上未见一致差异,整合方式滤器寿命更长或相当,但证据有限且异质,需进一步系统比较。

该范围综述检索了PubMed、Cochrane Library、Web of Science、Embase和ClinicalTrials.gov(最终检索2026年2月9日),纳入20项比较至少两种ECMO-RRT连接方式的研究,其中临床研究12项(含1项随机对照试验)、体外研究3项、计算机模拟1项、问卷调查4项。自2013年以来临床研究描述了超过8种ECMO-RRT连接方案,各方案间效率或临床结局未见一致差异,整合方式与滤器寿命更长或相当相关。问卷调查证实医疗实践异质,并报告了2项安全顾虑(不可停止的压力报警和空气栓塞)。作者结论认为证据有限且异质,需通过临床前和临床研究对所有连接方式进行系统严格比较以统一医疗实践并为指南制定提供依据。

要点

  • 纳入20项研究比较至少2种ECMO-RRT连接方式,其中仅1项为随机对照试验,证据整体有限且异质。
  • 自2013年以来临床研究描述了超过8种ECMO-RRT连接方案,各方案间效率或临床结局未见一致差异。
  • 整合式ECMO-RRT方式与滤器寿命更长或相当相关。
  • 问卷调查证实ECMO-RRT医疗实践异质,报告了2项安全顾虑:不可停止的压力报警和空气栓塞。
  • 作者认为需通过临床前和临床研究对所有连接方式进行系统严格比较,以统一医疗实践并为指南制定提供依据。

适用人群与场景

  • 接受体外膜氧合(ECMO)治疗心脏或呼吸衰竭、合并急性肾损伤(AKI)和液体过载需要肾脏替代治疗(RRT)的患者。

处置流程与药物剂量

  • ECMO与RRT可独立并行运行,或将RRT血滤器或设备整合入ECMO回路;自2013年以来临床研究已描述超过8种ECMO-RRT连接方案。

红旗征与不能漏的情况

  • 问卷调查报告了2项安全顾虑:不可停止的压力报警和空气栓塞。

研究设计与关键结果

  • 范围综述共检索6806条记录,最终纳入20项研究:临床研究12项(含1项随机对照试验)、体外研究3项、计算机模拟研究1项、问卷调查4项。
  • 自2013年以来临床研究描述了超过8种ECMO-RRT连接方案。
  • 所比较的各ECMO-RRT连接方式之间,效率或临床结局未见一致差异。
  • 整合式ECMO-RRT方式与滤器寿命更长或相当相关。
  • 问卷调查证实ECMO-RRT医疗实践异质,并报告了2项安全顾虑:不可停止的压力报警和空气栓塞。

证据类型、等级与局限

  • 证据类型:范围综述,纳入20项研究(临床研究12项含1项随机对照试验、体外研究3项、计算机模拟研究1项、问卷调查4项)。
  • 局限:原文报告证据有限且异质;作者指出需通过临床前和临床研究对所有连接方式进行系统严格比较以统一医疗实践并为指南制定提供依据。

对床旁实践的改变

  • 原文未据此给出推荐;作者结论认为ECMO-RRT连接方式证据有限且异质,未来需通过临床前和临床研究对所有连接方式进行系统严格比较,以统一医疗实践并为指南制定提供依据。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

INTRODUCTION: Extracorporeal membrane oxygenation (ECMO) is increasingly used for cardiac and respiratory failure and is associated with high rates of acute kidney injury (AKI) and fluid overload requiring renal replacement therapy (RRT). ECMO and RRT can be run independently in parallel, or RRT hemofilter or device can be integrated into the ECMO circuit. Several parallel and integrated ECMO-RRT configurations have been described, but the optimal approach remains uncertain.This scoping review mapped the available literature comparing at least two ECMO-RRT configurations, summarized reported outcomes, and identified knowledge gaps.

METHODS: We searched PubMed, Cochrane Library, Web of Science, Embase, and ClinicalTrials.gov without restrictions, with the final search conducted on February 9, 2026. Eligible studies compared at least 2 ECMO-RRT configurations linked to reported outcomes. Data on study characteristics, configurations, outcomes, and anticoagulation were extracted and synthesized.

RESULTS: Of 6,806 records, 20 studies were included: 12 were clinical including 1 randomized controlled trial, 3 in vitro, 1 in silico, and 4 surveys. More than 8 ECMO-RRT connection schemes have been described by clinical studies since 2013. No consistent differences in efficiency or clinical outcomes were observed between the configurations compared. Integrated methods were associated with longer or comparable hemofilter lifespan. Surveys confirmed heterogeneous medical practices, and 2 safety concerns (unstoppable pressure alarms and air embolism) were reported.

CONCLUSION: Evidence on ECMO-RRT configurations is limited and heterogeneous. Future systematic and rigorous comparison of all the configurations through preclinical and clinical studies is needed to harmonize medical practices and inform the development of guidelines.

来源:Ann Intensive Care,查看原文。版权归原作者所有。