自动体外除颤器启动后电击输送的操作瓶颈识别

Identifying the operational bottleneck to shock delivery after automated external defibrillator activation.
PubMed 文献 · Resuscitation · 2026-10-20
一句话结论AED启动至电极片贴放完成是最大可测量瓶颈,中位耗时42秒,贴片完成≥51秒与首次电击延迟超过60秒相关。

该回顾性研究分析了2018年1月至2026年2月间远程监控的Nihon Kohden AED的8115次激活事件的时间戳。结果显示,从开机到电极片贴放完成的中位时间为42秒(IQR 30-56),该间期占开机至首次分析总间期的中位81%(IQR 75-85)。在900次初始建议电击且记录到电击的事件中,电极片贴放至电击的中位时间为15秒(IQR 13-17)。探索性ROC分析得出开机至贴片完成≥51秒与首次电击在开机后60秒以上相关,建议贴片完成在50秒内作为探索性流程目标。

要点

  • AED开机至电极片贴放完成的中位时间为42秒(IQR 30-56),是开机至首次分析间期的主要组成部分(中位81%)。
  • 在初始建议电击且记录到电击的900次事件中,电极片贴放至电击的中位时间为15秒(IQR 13-17)。
  • 探索性ROC分析提示开机至贴片完成≥51秒与首次电击延迟超过60秒相关,贴片完成≤50秒可作为探索性流程目标。
  • 当首次分析发生在开机60秒后时,贴片间期占总间期的比例升至87%(IQR 85-90),凸显贴片环节的瓶颈效应。

适用人群与场景

  • 使用远程监控的Nihon Kohden AED的激活事件,排除检查记录;限初始建议电击且记录到电击输送的事件用于电击时序和ROC分析。

诊断标准与关键阈值

  • 探索性ROC分析得出开机至电极片贴放完成≥51秒的阈值,与首次记录电击在AED开机后超过60秒相关。

处置流程与药物剂量

  • 原文未提及

红旗征与不能漏的情况

  • 原文未提及

研究设计与关键结果

  • 回顾性分析8115次AED激活事件,中位开机至贴片完成42秒(IQR 30-56),贴片至首次分析9秒(IQR 9-11),开机至首次分析终点52秒(IQR 40-66)。
  • 贴片间期占开机至首次分析总间期的中位81%(IQR 75-85),当首次分析>60秒时升至87%(IQR 85-90)。
  • 在900次初始建议电击且记录电击的事件中,贴片至电击中位15秒(IQR 13-17);ROC分析开机至贴片≥51秒与首次电击>60秒相关。

证据类型、等级与局限

  • 回顾性研究,基于远程监控AED时间戳数据,样本量大但为单设备品牌、观察性设计,探索性阈值需前瞻性验证。

对床旁实践的改变

  • 对于使用AED的 cardiac arrest 场景,应关注开机至电极片贴放完成的时间,将其作为流程改进目标,争取贴片完成≤50秒以促进60秒内首次电击。

指南或共识出处

  • 原文未提及

以上提炼依据论文摘要(全文未开放获取,或许可不允许翻译转载);全文请点「阅读原文」。

英文原文(摘要)· 点击展开

INTRODUCTION: Cardiac arrest registries record automated external defibrillator (AED) use and shock delivery but not post-activation operational timing. We aimed to identify the largest measurable bottleneck and an exploratory pad-application target for shock delivery within 60 s of AED power-on.

METHODS: We retrospectively analyzed timestamps from remotely monitored Nihon Kohden AEDs between January 2018 and February 2026. After excluding inspection records, we grouped records into activation events and quantified intervals. Shock-timing and receiver operating characteristic (ROC) analyses were restricted to initially shock-advised events with recorded shock delivery.

RESULTS: We analyzed 8115 activation events. Median times were 42 s (interquartile range [IQR], 30-56) from power-on to completion of pad application, nine seconds (IQR, 9-11) from pad application to the first analysis, and 52 s (IQR, 40-66) from power-on to the first analysis endpoint. The power-on-to-pad-application interval accounted for a median of 81% (IQR, 75-85) of the total power-on-to-first-analysis interval, increasing to 87% (IQR, 85-90) when first analysis occurred more than 60 s after power-on. Among 900 initially shock-advised events with recorded shock delivery, the median pad-application-to-shock interval was 15 s (IQR 13-17). Exploratory ROC analysis yielded a power-on to pad-application threshold of ≥51 s in relation to first recorded shock delivery occurring more than 60 s after AED power-on.

CONCLUSIONS: The interval from AED power-on to electronically detected completion of pad application, encompassing all intervening actions, was the largest measurable post-activation bottleneck. Completion within 50 s may serve as an exploratory process target for first recorded shock delivery within 60 s among initially shock-advised events.