文献速递心脑血管急症

即时高敏肌钙蛋白能否提升急诊效率:美国多中心研究

Can point-of-care high sensitivity troponin improve efficiency: a multisite US study.
PubMed 文献 · Am J Emerg Med · 2026-08-19
一句话结论在无STEMI且至少开具一次hs-cTnI的成人急诊胸痛患者中,即时hs-cTnI将周转时间缩短36分钟,并可能将处置决策时间缩短51分钟、急诊停留时间缩短62分钟。

该前瞻性队列研究在美国3家急诊科(2025年2-9月)纳入602例无STEMI且至少开具一次hs-cTnI的成人患者,同时采集血样分别用i-STAT 1即时检测和中心实验室检测hs-cTnI。即时检测中位周转时间16分钟,中心实验室52分钟(差36分钟,p<0.001);基于即时检测的潜在处置决策时间中位222分钟 vs 273分钟(差51分钟,p<0.001);潜在急诊停留时间中位325分钟 vs 387分钟(差62分钟,p<0.001)。研究表明即时hs-cTnI可缩短周转时间并产生实质性的潜在处置决策和急诊停留时间缩短。

要点

  • 纳入标准:成人、无STEMI、至少开具一次hs-cTnI;共602例,48%(286/602)女性,中位年龄60岁(IQR 50-70)。
  • 即时hs-cTnI(i-STAT 1)中位周转时间16分钟(IQR 15-17),中心实验室(Beckman Coulter)52分钟(IQR 45-61),差36分钟(p<0.001)。
  • 基于即时检测的潜在处置决策时间中位222分钟(IQR 144-320),中心实验室273分钟(IQR 219-382),差51分钟(p<0.001)。
  • 潜在急诊停留时间中位325分钟(IQR 224-513)vs 387分钟(IQR 276-564),潜在缩短62分钟(p<0.001)。
  • 处置决策时间为基于即时检测结果对提供者进行问卷调查得到的潜在决策,非实际决策时间,需前瞻性验证。

适用人群与场景

  • 美国3家急诊科(2025年2-9月)纳入的成人患者,无STEMI且至少开具一次hs-cTnI;共602例,48%(286/602)女性,中位年龄60岁(IQR 50-70)。
  • 排除STEMI患者;原文未提及其他排除标准。

诊断标准与关键阈值

  • 原文未提及

处置流程与药物剂量

  • 同时采集血样,分别用i-STAT 1分析仪(Abbott Laboratories)行即时hs-cTnI检测、用Beckman Coulter行中心实验室hs-cTnI检测。
  • 即时hs-cTn检测后对提供者进行问卷调查,询问潜在处置和治疗决策,并与实际处置时间比较。
  • 原文未提及具体药物剂量、给药途径、频次或目标值。

红旗征与不能漏的情况

  • 原文未提及

研究设计与关键结果

  • 研究类型:前瞻性队列研究,美国3家急诊科,2025年2-9月,纳入602例无STEMI且至少开具一次hs-cTnI的成人患者。
  • 干预与对照:同时采集血样行即时hs-cTnI(i-STAT 1)与中心实验室hs-cTnI(Beckman Coulter)检测。
  • 主要终点:即时hs-cTnI中位周转时间16分钟(IQR 15-17)vs 中心实验室52分钟(IQR 45-61),差36分钟(p<0.001)。
  • 潜在处置决策时间中位222分钟(IQR 144-320)vs 273分钟(IQR 219-382),差51分钟(p<0.001)。
  • 潜在急诊停留时间中位325分钟(IQR 224-513)vs 387分钟(IQR 276-564),潜在缩短62分钟(p<0.001)。

证据类型、等级与局限

  • 证据类型:前瞻性队列研究(多中心,美国3家急诊科),注册号NCT06899776。
  • 局限:处置决策时间为基于即时检测结果的提供者问卷调查得到的潜在决策,非实际决策时间;原文未报告不良事件;样本量602例。

对床旁实践的改变

  • 在无STEMI且需检测hs-cTnI的成人急诊患者中,即时hs-cTnI可将周转时间从52分钟缩短至16分钟,可能加速处置决策和减少急诊停留时间。
  • 处置决策时间和急诊停留时间的缩短为潜在估计(基于问卷调查),实际临床推行前需前瞻性验证。
  • 原文未提及其他适用/不适用人群或常见陷阱。

指南或共识出处

  • 原文未提及

以上提炼依据论文摘要(全文未开放获取,或许可不允许翻译转载);全文请点「阅读原文」。

英文原文(摘要)· 点击展开

BACKGROUND: New point-of-care (POC) high sensitivity troponin (hs-cTn) assays could improve chest pain care efficiency. This study seeks to determine whether implementation of POC hs-cTn reduces key Emergency Department (ED) time metrics.

METHODS: A prospective cohort study was conducted at three U.S. EDs (February-September 2025). Adults without STEMI and at least one hs-cTnI ordered were accrued. Blood samples were collected simultaneously for POC hs-cTnI measurement on an i-STAT 1 analyzer (Abbott Laboratories) and central laboratory hs-cTnI measurement (Beckman Coulter). Following POC hs-cTn measures, providers were surveyed regarding potential disposition and treatment decisions which were compared to actual disposition times. Turnaround time, time-to-disposition, and ED length of stay for POC versus central laboratory measures were compared using Wilcoxon signed-rank tests.

RESULTS: During the study period, 602 patients were accrued, of which 48% (286/602) were female and median age was 60 years (IQR: 50-70). POC hs-cTnI had a median turnaround time of 16 min (IQR: 15-17) compared to 52 min (IQR: 45-61) for the central laboratory, a 36-min difference (p < 0.001). Median time to potential disposition decision using POC was 222 min (IQR: 144-320) versus 273 min (IQR: 219-382) for central laboratory measures, a difference of 51 min (p < 0.001). The median potential ED length of stay was 325 min (IQR: 224-513) for POC versus 387 min (IQR: 276-564) for the central laboratory, yielding a potential reduction of 62 min (p < 0.001).

CONCLUSIONS: POC hs-cTnI measurement reduced turnaround time and produced substantive potential reductions in time-to-disposition decision and ED length of stay.

TRIAL REGISTRATION: NCT06899776.

来源:Am J Emerg Med,查看原文。版权归原作者所有。