猪心脏骤停模型中EOlife X装置在常规与复苏期间同步通气时潮气量和呼吸频率的测量:方法比较研究
该前瞻性实验方法比较研究在猪心脏骤停模型中,将EOlife X(一种用于训练的呼吸反馈装置)的吸气潮气量(Vti)、呼气潮气量(Vte)和呼吸频率(f)与基于传感器的参考计算值进行比较。分析每只动物心脏骤停前2分钟和CPR期间2分钟的数据,动物经气管插管机械通气。采用Bland-Altman分析(线性混合模型校正重复测量)评估一致性,预设临床可接受差异为潮气量±50 mL、呼吸频率±3/min。结果:11只动物共821对通气(581次常规,240次复苏中)。常规通气时所有指标均在预设可接受差异内;复苏通气时所有指标均超出:Vti偏差-26 mL(一致性限-355至304 mL),Vte偏差-281 mL(-599至37 mL),f偏差23.8/min(-50.2至97.8/min)。超出可接受差异的比例:Vti 37.1%,Vte 86.2%,f 60.4%。波形提示胸外按压所致反向气流影响基于气流的测量。
要点
- 在猪心脏骤停模型中,胸外按压期间EOlife X的呼气潮气量偏差达-281 mL(一致性限-599至37 mL),86.2%测量值超出预设±50 mL可接受范围。
- 复苏期间呼吸频率偏差23.8/min(一致性限-50.2至97.8/min),60.4%测量值超出预设±3/min可接受范围。
- 复苏期间吸气潮气量偏差-26 mL(一致性限-355至304 mL),37.1%测量值超出预设±50 mL可接受范围。
- 常规通气时EOlife X的Vti、Vte和f均在预设临床可接受差异内(潮气量±50 mL,呼吸频率±3/min)。
- 波形分析提示胸外按压引起的反向气流影响基于气流的潮气量和呼吸频率计算,复苏条件下使用的通气监测设备需专门验证。
适用人群与场景
- 猪心脏骤停模型,经气管插管机械通气,分析心脏骤停前2分钟(常规通气)和CPR期间2分钟(复苏通气)的数据。
- 原文未提及人类患者或临床场景的纳入排除标准。
诊断标准与关键阈值
- 预设临床可接受差异:潮气量±50 mL,呼吸频率±3/min。
- 采用Bland-Altman分析,以线性混合模型校正重复测量得出偏差和一致性限。
处置流程与药物剂量
- 原文未提及
红旗征与不能漏的情况
- 胸外按压期间EOlife X的呼气潮气量、呼吸频率和吸气潮气量测量均可能超出临床可接受差异,波形提示按压所致反向气流干扰基于气流的测量。
研究设计与关键结果
- 研究类型:前瞻性实验方法比较研究,猪心脏骤停模型,11只动物,821对通气(581次常规,240次复苏中)。
- 常规通气时EOlife X所有指标均在预设可接受差异内;复苏通气时所有指标均超出:Vti偏差-26 mL(LoA -355至304 mL),Vte偏差-281 mL(LoA -599至37 mL),f偏差23.8/min(LoA -50.2至97.8/min)。
- 复苏通气时超出可接受差异的比例:Vti 37.1%,Vte 86.2%,f 60.4%。
- 波形检查提示胸外按压引起的反向气流影响基于气流的测量。
证据类型、等级与局限
- 证据类型:前瞻性实验方法比较研究(动物模型)。
- 局限:猪心脏骤停模型,样本量11只动物,非临床患者;原文未提及证据等级或推荐强度。
对床旁实践的改变
- EOlife X在常规通气时测量准确,但在胸外按压期间不准确,不能直接用于复苏中通气监测;用于CPR条件的通气监测设备应专门验证。
- 原文未提及对人类急诊/ICU实践的具体改变。
指南或共识出处
- 原文未提及
以上提炼依据论文摘要(全文未开放获取,或许可不允许翻译转载);全文请点「阅读原文」。
英文原文(摘要)· 点击展开
AIM: To evaluate the accuracy of the EOlife X, a ventilation feedback device for training, during regular ventilation and during intra-arrest ventilation with ongoing chest compressions.
METHODS: In this prospective experimental method-comparison study in a porcine cardiac arrest model, EOlife X-derived inspiratory tidal volume (Vti), expiratory tidal volume (Vte), and respiratory rate (f) were compared with sensor-based reference computations. Data were extracted by optical character recognition from screen recording of the EOlife X. Reference metrics were computed from continuous airflow and airway pressure measurements. For each animal, a two-minute-long period before cardiac arrest and a two-minute-long period of cardiopulmonary resuscitation (CPR) were analysed. The animals were mechanically ventilated via an endotracheal tube. Agreement was assessed using Bland-Altman analysis with linear mixed model derived bias and limits of agreement to account for repeated measures. Clinically acceptable differences were prespecified as ±50 mL for tidal volumes and ±3 /min for respiratory rate.
RESULTS: Data from 11 animals, yielding 821 paired ventilations (581 regular, 240 intra-arrest), were analysed. During regular ventilation, EOlife X measurements were within the prespecified clinically acceptable difference for all metrics, while exceeding it for all metrics during intra-arrest ventilation: Bias for Vtiwas -26 mL with limits of agreement (LoA) ranging from -355 to 304 mL, for Vtebias was -281 mL (LoA -599 to 37 mL), and for f23.8 /min (LoA -50.2 to 97.8 /min). The proportions of measurements outside the prespecified clinically acceptable difference during intra-arrest ventilation were 37.1% for Vti, 86.2% for Vte, and 60.4% for f. Waveform inspection suggested that chest compression-induced reverse airflow affects airflow-based measurements.
CONCLUSION: EOlife X showed agreement within predefined clinically acceptable limits during regular ventilation, but not during intra-arrest ventilation with ongoing chest compressions. Chest compression-induced reverse airflow appears to compromise airflow-based calculation of tidal volume and respiratory rate. Devices for ventilatory monitoring should be validated specifically under CPR conditions.
参考来源
- Orlob S, Purkarthofer D, Kern WJ, et al. Measurement of tidal volumes and respiratory rate during regular and asynchronous intra-arrest ventilation using the EOlife X device: A method-comparison study in a porcine cardiac arrest model. Resuscitation. 2026. PMID: 42492618. doi:10.1016/j.resuscitation.2026.111217