Reducing Physical Restraint in the ICU: Multicenter Phase II Randomized Trial of a Novel Device Versus Traditional Wrist Restraints.
OBJECTIVES: To evaluate a novel device, developed under a National Institutes of Health Small Business Technology Transfer program grant, designed as an alternative to traditional wrist restraints. DESIGN: Within-patient crossover phase II randomized trial. SETTING: Four U.S.…
英文原文(摘要)· 点击展开
OBJECTIVES: To evaluate a novel device, developed under a National Institutes of Health Small Business Technology Transfer program grant, designed as an alternative to traditional wrist restraints.
DESIGN: Within-patient crossover phase II randomized trial.
SETTING: Four U.S. academic medical centers.
PATIENTS: ICU patients ( n = 54) who were verbally responsive with a provider order for wrist restraints and expected stay greater than or equal to 2 days.
INTERVENTIONS: Alternating use of novel device vs. traditional restraint during 4-hour assessment periods over 2 days (four assessment periods), with randomization of initial allocation to novel device vs. restraint.
MEASUREMENTS AND MAIN RESULTS: Upper extremity activity (measured via wrist actigraphy) was the primary outcome, with safety, sedation, delirium, satisfaction (via Quebec User Evaluation of Satisfaction with Assistive Technology [QUEST] survey), and qualitative analyses of survey data as secondary outcomes. Analysis of actigraphy data demonstrated similar mean activity counts per 4-hour session for the novel device vs. traditional restraints, with a range of mean activity across the four sessions of 2930-4395 vs. 3898-4910, respectively, with relative mean total activity count per session of 0.93 (95% CI, 0.76-1.13; p = 0.45). Across 98 sessions using the novel device, 2 (2%) potential safety events were reported. No statistically significant differences were observed in sedation/agitation, or delirium status and severity. Of the QUEST satisfaction surveys completed by patient/family ( n = 25) and clinicians ( n = 79), 58% and 51%, respectively, indicated high satisfaction with qualitative analyses demonstrating four positive and six constructive themes.
CONCLUSIONS: No differences in upper extremity activity, delirium, sedation, or agitation were demonstrated between the novel device and traditional wrist restraints. The novel device had overall moderate patient/family and clinician satisfaction, and qualitative analyses provide important insights for advancing the development of novel restraint alternative devices.
参考来源
- Stapleton RD, Kamdar BB, Colantuoni EA, et al. Reducing Physical Restraint in the ICU: Multicenter Phase II Randomized Trial of a Novel Device Versus Traditional Wrist Restraints. Crit Care Med. 2026. PMID: 42206929. doi:10.1097/CCM.0000000000007172
供执业医师抢救与诊疗参考,不替代临床判断;剂量请结合体重、肝肾功能和本院药品规格复核,以指南原文为准。