文献速递复苏、休克与脓毒症心脑血管急症

除颤电流与电极片位置对难治性室颤自主循环恢复的影响

The impact of defibrillation current and pad position on return of spontaneous circulation during refractory ventricular fibrillation.
PubMed 文献 · Resuscitation · 2026-03-17
一句话结论在难治性室颤中,前后位电极片位置(而非实际交付电流)与自主循环恢复独立相关(aOR 2.01)。

该研究为DOSE VF试验的二次按方案分析,纳入342例难治性室颤患者,计算前六次除颤的经胸电流,比较标准前侧位、向量改变前后位及双序贯体外除颤的电流交付。前后位除颤(向量改变19.6 A、双序贯19.2 A)的平均电流显著高于标准前侧位(16.6 A)和双序贯前侧位(17.3 A)(P<0.001)。广义估计方程模型显示,前后位位置与自主循环恢复独立相关(aOR 2.01;95%CI 1.12-3.59),而交付电流与自主循环恢复无关。作者认为,额外的前后位电极片位置可能是向量改变和双序贯体外除颤获益的关键机制因素。

要点

  • 在难治性室颤患者中,前后位除颤(向量改变19.6 A、双序贯19.2 A)的平均电流显著高于标准前侧位(16.6 A)和双序贯前侧位(17.3 A)(P<0.001)。
  • 广义估计方程模型显示,前后位电极片位置与自主循环恢复独立相关(aOR 2.01;95%CI 1.12-3.59)。
  • 交付电流与自主循环恢复无独立关联,提示电极片位置本身而非电流大小可能是获益机制。
  • 作者认为,额外的前后位电极片位置可能是向量改变和双序贯体外除颤获益的关键机制因素。

适用人群与场景

  • 纳入DOSE VF试验的难治性室颤患者,共342例(平均年龄63.6岁;85%男性)。

研究设计与关键结果

  • 研究类型:DOSE VF试验的二次按方案分析;样本量342例;人群为难治性室颤患者;干预为前六次除颤中标准前侧位、向量改变前后位及双序贯体外除颤(前侧位+前后位)的电流交付比较。
  • 共分析849次干预性除颤(51.6%标准、25.4%向量改变、22.9%双序贯)。
  • 平均电流:向量改变前后位19.6 A、双序贯前后位19.2 A,均显著高于标准前侧位16.6 A和双序贯前侧位17.3 A(P<0.001)。
  • 广义估计方程模型显示,前后位位置与自主循环恢复独立相关(aOR 2.01;95%CI 1.12-3.59),交付电流与自主循环恢复无关。

证据类型、等级与局限

  • 证据类型:DOSE VF试验的二次按方案分析。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 作者认为,额外的前后位电极片位置可能是向量改变和双序贯体外除颤获益的关键机制因素。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

BACKGROUND: Although pad position influences current delivery, the relationship between transthoracic current, pad position, and shock efficacy during external defibrillation for refractory ventricular (VF) has not been examined.

OBJECTIVE: To assess the relationship between pad position, current delivery and return of spontaneous circulation (ROSC) among patients enrolled in the Double Sequential External Defibrillation for Refractory VF (DOSE VF) trial.

METHODS: We conducted a secondary, per-protocol analysis of patients enrolled in DOSE VF. Transthoracic current was calculated for the first six shocks using defibrillator-derived measures of shock energy and impedance, applying manufacturer-provided formulas. Current delivery was compared across standard anterolateral (AL), vector-change (VC) anteroposterior (AP), and double sequential external defibrillation (DSED; AL + AP) shocks. Multivariable logistic regression and generalized estimating equation (GEE) models were used to examine associations between current delivery, pad position and ROSC.

RESULTS: 342 patients were included (mean age 63.6 years; 85% male), contributing 849 interventional shocks (51.6% standard, 25.4% VC and 22.9% DSED). Mean current was significantly greater for AP shocks in both VC (19.6 A) and DSED (19.2 A), compared with standard AL (16.6 A) and DSED AL shocks (17.3 A) (P < 0.001). In GEE modeling, AP positioning was independently associated with ROSC compared to AL positioning (aOR 2.01; 95% CI: 1.12-3.59). Delivered current was not associated with ROSC.

CONCLUSIONS: AP delivered higher current than AL positioning; however, pad position rather than delivered current was independently associated with ROSC. These findings suggest the additional AP pad position may be a key mechanistic contributor to the benefit of VC and DSED.

来源:Resuscitation,查看原文。版权归原作者所有。

参考来源

  1. Cheskes S, Drennan IR, Turner L, et al. The impact of defibrillation current and pad position on return of spontaneous circulation during refractory ventricular fibrillation. Resuscitation. 2026. PMID: 41856454. doi:10.1016/j.resuscitation.2026.111061

供执业医师抢救与诊疗参考,不替代临床判断;剂量请结合体重、肝肾功能和本院药品规格复核,以指南原文为准。