无手柄颈动脉多普勒超声在院内心搏骤停中实时检测自主循环恢复的多中心随机对照研究

Evaluating hands-free carotid Doppler ultrasound for real-time return of spontaneous circulation detection in in-hospital cardiac arrest: a multicenter randomized controlled study.
PubMed 文献 · Resuscitation · 2026-02-18
一句话结论院内心搏骤停患者中,无手柄颈动脉多普勒超声(CADFlow)较手动触诊评估自主循环恢复的时间更短(2.9秒 vs 7.2秒),自主循环恢复成功率无显著差异(43% vs 33%)。

该多中心随机对照研究纳入5所医科大学附属教学医院的341例院内心搏骤停成人患者,比较无手柄颈动脉超声(CADFlow)与手动颈动脉触诊评估自主循环恢复的效果。CADFlow组自主循环恢复成功率为43%,对照组为33%,差异未达统计学显著(P=0.053),绝对增加10个百分点。CADFlow组评估颈动脉搏动所需平均时间显著缩短(2.9秒 vs 7.2秒,P<0.001)。胸外按压深度与CADFlow参数呈强相关,提示该系统可有效评估心肺复苏按压质量。作者认为CADFlow系统改进了自主循环恢复检测、缩短评估时间并提供心肺复苏质量优化的实时反馈。

要点

  • 院内心搏骤停成人患者中,CADFlow组自主循环恢复成功率为43%,手动触诊组为33%,差异未达统计学显著(P=0.053),绝对增加10个百分点。
  • CADFlow组评估颈动脉搏动的平均时间显著短于手动触诊组(2.9秒(标准差0.8) vs 7.2秒(标准差1.3),P<0.001)。
  • 胸外按压深度与CADFlow参数之间观察到强相关,提示该系统可有效评估心肺复苏按压质量。
  • 作者认为CADFlow无手柄颈动脉多普勒超声系统改进了自主循环恢复检测、缩短评估时间并提供心肺复苏质量优化的实时反馈。

适用人群与场景

  • 5所医科大学附属教学医院的院内心搏骤停成人患者,共341例。

研究设计与关键结果

  • 多中心随机对照研究,纳入5所医科大学附属教学医院的341例院内心搏骤停成人患者,使用PALMCPR系统监测胸外按压频率和深度。
  • CADFlow组使用无手柄颈动脉超声监测颈动脉搏动,对照组依靠手动触诊颈动脉评估自主循环恢复。
  • 主要终点:CADFlow组自主循环恢复成功率为43%,对照组为33%,差异未达统计学显著(P=0.053),绝对增加10个百分点。
  • CADFlow组评估颈动脉搏动的平均时间显著短于对照组(2.9秒(标准差0.8) vs 7.2秒(标准差1.3),P<0.001)。
  • 胸外按压深度与CADFlow参数之间观察到强相关,提示该系统可有效评估心肺复苏按压质量。

证据类型、等级与局限

  • 证据类型:多中心随机对照研究。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 作者认为CADFlow无手柄颈动脉多普勒超声系统可改进院内心搏骤停患者的自主循环恢复检测、缩短评估时间并提供心肺复苏质量优化的实时反馈。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

OBJECTIVE: This study investigated the effectiveness and accuracy of hands-free carotid ultrasound (CADFlow, Continuous Auto Doppler Flow, Sensus Medical, Suzhou, China) for monitoring carotid blood flow in assessing the return of spontaneous circulation (ROSC), in comparison to manual palpation of major arterial pulses.

METHODS: A total of 341 adult patients with in-hospital cardiac arrest were enrolled from five affiliated teaching hospitals of medical universities. Chest compression rate and depth were monitored using the Prehospital Advanced Life Support Manual Cardiopulmonary Resuscitation (PALMCPR, SunLife Science, Suzhou, China) system. In the CADFlow group, hands-free carotid ultrasound was used to monitor carotid pulsation, while the control group relied on manual palpation of the carotid artery to assess ROSC. The groups were compared in terms of ROSC success rates, time to ROSC determination, and ROSC misjudgment rates. Additionally, the relationships between chest compression parameters (depth and rate) and carotid blood flow were analyzed.

RESULTS: The ROSC success rate was non-significantly higher in the CADFlow group than in the control group (43% vs. 33%, P = 0.053), corresponding to a 10-percentage-point absolute increase The mean time required to assess carotid pulsation was significantly shorter in the CADFlow group (2.9 (SD 0.8) s vs. 7.2 (SD 1.3) s, P < 0.001). Strong correlations were observed between chest compression depth and CADFlow parameters, suggesting that the system can effectively assess Cardiopulmonary Resuscitation (CPR) compression quality.

CONCLUSIONS: The CADFlow hands-free carotid Doppler ultrasound system improves ROSC detection, shortens assessment time, and provides real-time feedback for CPR quality optimisation.

TRIAL REGISTRATION: This clinical trial is registered with the Chinese Clinical Trial Registry (ChiCTR), with the registration number ChiCTR2200064633. The trial was registered on October 13, 2022. The full trial protocol is available upon request from the corresponding author.

来源:Resuscitation,查看原文。版权归原作者所有。