院内真实心脏骤停中专业人员对自动体外除颤器语音提示的依从性:设备数据分析

Adherence to automated external defibrillator instructions during real-life in-hospital cardiac arrest: an analysis of device data.
PubMed 文献 · Resuscitation · 2026-01-29
一句话结论院内真实心脏骤停中,专业人员对「开始CPR」「按下放电键」依从性约八成,但对「按压再用力」依从性仅15.3%,依从该提示与高质量按压相关。

该回顾性观察研究分析德国一家医院2007至2024年466份自动体外除颤器(AED)数据中的419例院内心脏骤停案例,评估专业人员对AED语音提示的依从性及其对胸外按压质量的影响。结果显示不同提示依从率从15.3%到81.1%不等(P<0.001),「开始CPR」和「按下放电键」依从率分别为81.1%和77.7%,而「按压再用力」仅15.3%;「按压再用力」提示后平均按压深度从3.26 cm增至3.66 cm、3.61 cm、3.51 cm(P<0.001),平均按压频率仅在第一个间期从89.8次/分改善至102.2次/分(P<0.001)。对「按压再用力」的依从与高质量胸外按压相关(54.6%,P<0.001)。作者建议面向专业人员的AED应加入分析前倒计时或提示音,并提供激励-纠正性指令及提示及时更换施救者。

要点

  • 419例院内心脏骤停中,AED提示依从率从15.3%到81.1%不等(P<0.001),纠正性提示「按压再用力」依从率最低(15.3%)。
  • 「开始CPR」和「按下放电键」依从率分别为81.1%和77.7%,核心提示依从性较高。
  • 「按压再用力」提示后平均按压深度从3.26 cm增至3.66 cm、3.61 cm、3.51 cm(P<0.001);平均按压频率仅在第一个间期从89.8次/分改善至102.2次/分(P<0.001)。
  • 对「按压再用力」的依从与高质量胸外按压相关者为54.6%(P<0.001)。
  • 作者建议面向专业人员的AED应加入分析前倒计时或提示音,并提供激励-纠正性指令及提示及时更换施救者。

适用人群与场景

  • 德国一家医院2007至2024年院内心脏骤停的真实AED设备数据,共分析419例(来自466份AED文件)。

研究设计与关键结果

  • 回顾性观察研究,分析德国一家医院2007至2024年466份AED文件中的419例院内心脏骤停案例,数据包括节律分析周期、语音提示和胸外按压质量指标的时间戳记录。
  • 依从性定义为按预设标准及时执行提示;各提示依从率范围为15.3%至81.1%(P<0.001)。
  • 「开始CPR」依从率81.1%,「按下放电键」依从率77.7%,「按压再用力」依从率15.3%。
  • 「按压再用力」提示后平均按压深度从3.26 cm增至3.66 cm、3.61 cm、3.51 cm(P<0.001)。
  • 平均按压频率仅在第一个间期从89.8次/分改善至102.2次/分(P<0.001)。
  • 对「按压再用力」的依从与高质量胸外按压相关者为54.6%(P<0.001)。

证据类型、等级与局限

  • 证据类型:回顾性观察研究。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 作者认为,面向专业人员的AED应加入分析前倒计时或提示音以提高使用者依从性。
  • 作者认为,设备应提供激励-纠正性指令并提示及时更换施救者,以确保按压质量持续。
  • 数据提示,专业人员对中断性及纠正性AED提示的依从性低于核心提示,且对「按压再用力」的依从与高质量胸外按压相关。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

BACKGROUND: High-quality chest compression (CC) and timely defibrillation are crucial to survive in-hospital cardiac arrest (IHCA). Automated external defibrillators (AED) are implemented in hospitals to minimize defibrillation delay. This study examined whether AED prompts were followed by professionals during IHCA, and analyzed whether adherence influenced CC quality.

METHODS: A retrospective observational study was conducted on real AED data from a German hospital between 2007 and 2024. Data included time-stamped documentation of rhythm analysis cycles, voice prompts, and CC quality metrics. Adherence was defined as the timely implementation of prompts according to predefined criteria. Statistical analysis included descriptive statistics, non-parametric comparisons, Generalized Estimating Equations for adherence rates, and chi-square testing.

RESULTS: Among 466 AED files, 419 cases were analyzed. Prompt adherence ranged from 15.3% to 81.1% (p < 0.001). The prompts "start CPR" and "press shock button" were followed in 81.1% and 77.7% of cases, while "push harder" was followed in 15.3%. Mean CC depth increased after the "push harder" prompt (from 3.26 cm to 3.66 cm, 3.61 cm, 3.51 cm; p < 0.001). Mean CC rate improved only within the first interval (from 89.8 cpm to 102.2 cpm; p < 0.001). Adherence to "push harder" was associated with high-quality CC in 54.6% of cases (p < 0.001).

CONCLUSION: Adherence to core AED prompts was high, whereas interruptive and corrective instructions were disregarded more frequently. AEDs for professionals should incorporate pre-analysis countdowns or alert tones to enhance user compliance. To ensure sustained high-quality compressions, devices should provide motivational-corrective instructions and prompt timely rescuer switches.