急诊科鼻高流量氧疗的自动化滴定:一项随机对照试验
该开放标签、平行组随机对照试验在惠灵顿地区医院急诊科于2022年10月至2023年12月期间,比较了鼻高流量氧疗中自动化与手动氧滴定对低氧血症成人患者的效果。共筛查83例,随机52例,49例有主要终点数据。主要终点为达到至少30分钟治疗的患者中,SpO2处于目标范围(92%-96%,或高碳酸血症风险者为88%-92%)的时间比例。自动化组(n=25)中位时间比例为96.4%(IQR 92.5%-99.4%),手动组(n=24)为89.9%(IQR 69.8%-97.2%),差值8.0%(95%CI 1.7%-16.9%),p=0.01。未发现目标SpO2范围本身影响该比例(交互作用P=0.60)。结论认为自动化滴定显著增加目标SpO2范围内时间。
要点
- 急诊科鼻高流量氧疗成人患者中,自动化滴定组目标SpO2范围内时间比例中位数96.4%,手动组89.9%,差值8.0%(95%CI 1.7%-16.9%),p=0.01。
- 主要终点分析仅纳入达到至少30分钟治疗的患者,共49例有数据(自动化组n=25,手动组n=24)。
- 目标SpO2范围为92%-96%,高碳酸血症风险者为88%-92%,是否落在目标范围不取决于所选目标范围(交互作用P=0.60)。
- 研究为开放标签、平行组随机对照试验,在急诊科单中心进行,共随机52例。
适用人群与场景
- 惠灵顿地区医院急诊科的低氧血症成人患者,已开具目标SpO2范围且表现出最低氧需求者符合纳入条件。
- 主要终点分析限于达到至少30分钟鼻高流量氧疗的患者。
诊断标准与关键阈值
- 目标SpO2范围:92%-96%;高碳酸血症风险者为88%-92%。
- 纳入条件:已开具目标SpO2范围且表现出最低氧需求。
研究设计与关键结果
- 研究类型:开放标签、平行组随机对照试验;筛查83例,随机52例,49例有主要终点数据。
- 主要终点:达到至少30分钟治疗的患者中,SpO2处于目标范围的时间比例;自动化组(n=25)中位96.4%(IQR 92.5%-99.4%) vs 手动组(n=24)中位89.9%(IQR 69.8%-97.2%),差值8.0%(95%CI 1.7%-16.9%),p=0.01。
- 交互作用分析:目标SpO2范围(92%-96%或88%-92%)不影响目标范围内时间比例,P-interaction 0.60。
证据类型、等级与局限
- 证据类型:随机对照试验。
- 局限:原文未报告局限性。
对床旁实践的改变
- 在急诊科对低氧血症成人患者使用鼻高流量氧疗时,自动化氧滴定较手动滴定可增加SpO2处于目标范围(92%-96%,高碳酸血症风险者88%-92%)的时间比例,作者结论认为自动化滴定显著增加目标范围内时间。
提炼说明
- 本条依据论文摘要整理;摘要未涉及「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开
BACKGROUND: When delivering nasal high flow (NHF) therapy in a medical ward, a high dependency unit or an intensive care unit, automated oxygen titration increases time spent within a target oxygen saturation (SpO2) range compared with standard, manually titrated oxygen. This study explores whether this improvement is also seen when titrating oxygen with NHF in an emergency department (ED).
METHOD: This open-label, parallel groups, randomised controlled trial compared automated to manual oxygen titration using NHF therapy in hypoxaemic adult patients in the Wellington Regional Hospital ED between October 2022 and December 2023. Participants with a prescribed target SpO2 range who demonstrated a minimum oxygen requirement were eligible for inclusion. A rank-based comparison was used for the primary outcome, the proportion of time spent within the target SpO2 range of 92%-96%, or 88%-92% if at risk of hypercapnia, among participants achieving ≥30 min of therapy. An interaction term was applied to assess whether the proportion of time spent within target SpO2 range depended on the prescribed target range itself (SpO2 92%-96% or 88%-92%).
RESULTS: 83 participants were screened, 52 were randomised and 49 had data for the primary endpoint. Median (IQR) proportion of time spent within the target SpO2 range with automated oxygen (n=25) was 96.4% (92.5% to 99.4%) compared with 89.9% (69.8% to 97.2%) with manually adjusted oxygen (n=24); difference (95% CI) 8.0% (1.7% to 16.9%), p=0.01. There was no evidence that the proportion of time spent within target SpO2 range depended on the selected target SpO2 range, P-interaction 0.60.
CONCLUSION: Automatically titrated oxygen therapy significantly increased time spent within a target SpO2 range, compared with manual oxygen titration in adult patients receiving NHF therapy in the ED.