一句话结论使用气道评估与喉镜镜片选择辅助工具可显著提高危重患者急诊气管插管一次成功率(91.3% vs 73.8%)。

广东省中医院医护人员设计了一种由两个直角手柄和一把直尺组成的气道评估与喉镜镜片选择辅助工具,并获国家实用新型专利(专利号:ZL 2021 2 1187307.9)。该工具手柄宽度基于「3-3-2」气道评估规则(LEMON规则),分别对应患者三指宽和两指宽,用于测量张口度(三指手柄)、颏-舌骨距离(三指手柄)、舌骨-甲状腺切迹距离(两指手柄)以初步识别困难气道;直尺用于精确测量耳垂至口角距离,为选择合适喉镜镜片长度提供客观依据。研究纳入2023年1—12月急诊和重症医学科需紧急气管插管的80例患者为观察组(使用该工具进行插管前评估),以2022年接受常规紧急气管插管的80例患者为对照组。结果显示观察组一次插管成功率为91.3%(73/80),显著高于对照组的73.8%(59/80)(χ2=8.485,P=0.004);并发症发生率虽低于对照组但差异无统计学意义。

要点

  • 观察组使用辅助工具后一次插管成功率为91.3%(73/80),显著高于对照组的73.8%(59/80)(χ2=8.485,P=0.004)。
  • 观察组中6例(7.5%)使用工具插管前识别为困难气道,均使用纤维支气管镜成功插管。
  • 观察组其余74例患者一次插管成功率为98.6%(73/74)。
  • 观察组牙齿脱落发生率2.5% vs 对照组5.0%(χ2=0.692,P=0.405),口腔出血6.3% vs 15.0%(χ2=3.224,P=0.073),差异均无统计学意义。
  • 该工具手柄宽度基于「3-3-2」气道评估规则(LEMON规则),直尺用于测量耳垂至口角距离以选择喉镜镜片长度。

适用人群与场景

  • 2023年1—12月广东省中医院急诊和重症医学科需紧急气管插管的80例患者(观察组)。
  • 2022年接受常规紧急气管插管的80例患者(对照组)。

诊断标准与关键阈值

  • 「3-3-2」气道评估规则(LEMON规则):三指手柄测量张口度和颏-舌骨距离,两指手柄测量舌骨-甲状腺切迹距离,用于初步识别困难气道。
  • 直尺测量耳垂至口角距离,为选择合适喉镜镜片长度提供客观依据。

处置流程与药物剂量

  • 插管前使用辅助工具进行气道评估:用三指手柄测量张口度和颏-舌骨距离,用两指手柄测量舌骨-甲状腺切迹距离,初步识别困难气道。
  • 用直尺精确测量耳垂至口角距离,据此选择合适喉镜镜片长度。
  • 观察组中工具识别为困难气道的6例(7.5%)均使用纤维支气管镜成功插管。

研究设计与关键结果

  • 研究类型为临床对照研究,观察组80例使用辅助工具进行插管前评估,对照组80例接受常规紧急气管插管。
  • 观察组一次插管成功率为91.3%(73/80),显著高于对照组的73.8%(59/80)(χ2=8.485,P=0.004)。
  • 观察组中6例(7.5%)使用工具插管前识别为困难气道,均使用纤维支气管镜成功插管;其余74例一次插管成功率为98.6%(73/74)。
  • 牙齿脱落发生率观察组2.5% vs 对照组5.0%(χ2=0.692,P=0.405),口腔出血发生率观察组6.3% vs 对照组15.0%(χ2=3.224,P=0.073),差异均无统计学意义。

证据类型、等级与局限

  • 证据类型:临床对照研究。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 作者认为该辅助工具能有效识别困难气道、指导准确选择喉镜镜片,并显著提高危重患者气管插管一次成功率。
  • 作者指出该工具设计简单,旨在缩短插管时间并间接提高一次插管成功率。
  • 该工具已获国家实用新型专利(专利号:ZL 2021 2 1187307.9)。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

Tracheal intubation is a key technology for establishing an artificial airway during the resuscitation of critically ill patients. Its success directly impacts patient life safety and prognosis. Identification of difficult airway and improper selection of laryngoscope blades are important factors leading to intubation failure and complications. To optimize this process, medical staff at Guangdong Provincial Hospital of Chinese Medicine have designed an auxiliary tool for airway assessment and laryngoscope blade selection, and have obtained a National Utility Model Patent of China (patent number: ZL 2021 2 1187307.9). The device is mainly composed of two right-angled handles and a ruler. The width of the handle is based on the "3-3-2" airway assessment rule (LEMON rule), corresponding to the patient's three-finger and two-finger widths, respectively, and is used to measure mouth opening (three-finger handle), chin-hyoid distance (three-finger handle), hyoid-thyroid notch distance (two-finger handle) to preliminarily identify difficult airways. The ruler is used to accurately measure the distance from the earlobe to the corner of the mouth, providing an objective basis for selecting the appropriate laryngoscope blade length. This tool is designed to be simple and is intended to shorten intubation time and indirectly improve first-time intubation success rate. To verify the clinical application effect of the auxiliary tool for airway assessment and laryngoscope blade selection, 80 patients who required emergency endotracheal intubation admitted to the department of emergency and intensive care unit of Guangdong Provincial Hospital of Chinese Medicine from January to December 2023 were enrolled, and the tool was used for pre-intubation assessment (observation group); and 80 patients who received routine emergency tracheal intubation in 2022 were used as the control (control group). The results showed that 6 cases (7.5%) in observation group had difficulty in identifying difficult airways before intubation using the tool, and all were successfully intubated using fiberoptic bronchoscopes. The success rate of one-time intubation for the remaining 74 patients was 98.6% (73/74). The success rate of one-time intubation in the observation group was significantly higher than that in the control group [91.3% (73/80) vs. 73.8% (59/80), χ2 = 8.485, P = 0.004]. Although the incidence of complications such as tooth loss and oral bleeding in the observation group was lower than that in the control group, the differences were not statistically significant (tooth loss: 2.5% vs. 5.0%, χ 2 = 0.692, P = 0.405; oral bleeding: 6.3% vs. 15.0%, χ 2 = 3.224, P = 0.073). The results showed that the auxiliary tool can effectively identify difficult airways, guide the accurate selection of laryngoscope blades, and significantly improve the one-time endotracheal intubation success rate of critically ill patients.