便携式早产儿转运呼吸装置的设计及其在基层医院极早产儿院内转运中的应用
该前瞻性随机对照试验纳入2023年5月至10月出生并转入新生儿重症监护病房(NICU)接受氧疗支持的极早产儿73例,随机分为对照组38例和观察组35例。对照组使用传统T组合复苏器连接纯氧进行呼吸支持与院内转运,观察组使用医护人员设计制作的便携式早产儿转运呼吸装置(304不锈钢材质,含T组合复苏器、空气氧气混合器、空气罐、纯氧钢瓶、减压阀、伸缩杆、托盘、挂钩、底板和四个移动轮,可实现转运中FiO2精准控制)。结果显示,观察组转入NICU后目标SpO2首次达标率显著高于对照组[94.29%(33/35)vs. 26.32%(10/38),P<0.05],PaO2控制范围更优[85.50±6.36 mmHg vs. 103.00±2.83 mmHg,P<0.05],人力支出和转运所需时间均显著减少[人力支出(人数):2.14±0.35 vs. 3.17±0.34,转运所需时间(分钟):10.42±0.76 vs. 15.54±0.34,均P<0.05]。作者认为该装置可改善目标SpO2达标率、将PaO2控制在目标范围内、避免转运中缺氧或高氧,且小巧便携、节省人力与时间、成本效益好,适合基层医院广泛使用。
要点
- 观察组转入NICU后目标SpO2(0.90-0.95)首次达标率94.29%(33/35),显著高于对照组26.32%(10/38),P<0.05。
- 观察组PaO2为85.50±6.36 mmHg,对照组为103.00±2.83 mmHg,观察组控制范围更优,P<0.05。
- 观察组转运人力支出2.14±0.35人,对照组3.17±0.34人,观察组显著减少,P<0.05。
- 观察组转运所需时间10.42±0.76分钟,对照组15.54±0.34分钟,观察组显著缩短,P<0.05。
- 该装置由304不锈钢制成,含T组合复苏器、空气氧气混合器、空气罐、纯氧钢瓶、减压阀等,可实现转运中FiO2精准控制。
适用人群与场景
- 2023年5月至10月出生并转入NICU接受氧疗支持的极早产儿,在基层医院进行院内转运。
诊断标准与关键阈值
- 目标SpO2达标定义为SpO2达到0.90-0.95。
处置流程与药物剂量
- 对照组:使用传统T组合复苏器连接纯氧进行呼吸支持并实施院内转运。
- 观察组:使用医护人员设计制作的便携式早产儿转运呼吸装置(304不锈钢材质,主要含T组合复苏器、空气氧气混合器、空气罐、纯氧钢瓶、减压阀、伸缩杆、托盘、挂钩、底板和四个移动轮)进行呼吸支持并实施院内转运,可实现转运中FiO2精准控制。
研究设计与关键结果
- 研究类型:前瞻性随机对照试验;样本量:共73例极早产儿,对照组38例,观察组35例;两组性别、出生胎龄、出生体重、分娩方式、出生后1分钟和5分钟Apgar评分及转运期间氧疗情况无显著差异。
- 主要终点:观察组转入NICU后目标SpO2首次达标率94.29%(33/35)vs. 对照组26.32%(10/38),P<0.05。
- 次要终点:观察组PaO2控制范围85.50±6.36 mmHg vs. 对照组103.00±2.83 mmHg,P<0.05;观察组人力支出2.14±0.35人 vs. 对照组3.17±0.34人,P<0.05;观察组转运所需时间10.42±0.76分钟 vs. 对照组15.54±0.34分钟,P<0.05。
证据类型、等级与局限
- 证据类型:前瞻性随机对照试验。
- 局限:原文未报告局限性。
对床旁实践的改变
- 作者认为该便携式早产儿转运呼吸装置用于基层医院极早产儿转运中的呼吸支持,可提高目标SpO2达标率、将PaO2控制在目标范围内、避免转运中缺氧或高氧。
- 作者认为该装置小巧便携、可节省转运中的人力资源与时间、成本效益好,适合基层医院广泛使用。
提炼说明
- 本条依据论文摘要整理;摘要未涉及「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开
OBJECTIVE: To design a portable respiratory device for transporting premature infants and explore its application effect in the in-hospital transportation of extremely premature infants in primary hospitals.
METHODS: A prospective randomized controlled trial was conducted. The extremely premature infants born and transferred to neonatal intensive care unit (NICU) with oxygen therapy support from May to October in 2023 were selected and randomly divided into control group and observation group. The infants in the control group received respiratory support and in-hospital transportation using a traditional T-combination resuscitator connected to pure oxygen, and those in the observation group used a portable premature infant transport respiratory device designed and manufactured by medical staff to provide respiratory support and implement in-hospital transportation. The respiratory device for transporting premature infants is made of 304 stainless steel material, mainly consisting of a T-combination resuscitator, an air oxygen mixer, an air tank, a pure oxygen cylinder, a pressure reducing valve, a telescopic rod, a tray, a hook, a bottom plate, and four moving wheels, which can achieve precise control of the fraction of inspired oxygen (FiO2) during transportation. The achievement rate of first-time target pulse oxygen saturation (SpO2, achieving a target SpO2 of 0.90-0.95 was considered as meeting the standard) and arterial partial pressure of oxygen (PaO2) after being transferred to the NICU, as well as the manpower expenditure and time required for transportation of pediatric patients between the two groups were observed.
RESULTS: A total of 73 extremely premature infants were enrolled, including 38 in the control group and 35 in the observation group. There was no significant difference in the gender, gestational age at birth, birth weight, mode of delivery, Apgar score at 1 minute and 5 minutes after birth, and oxygen therapy during the transportation between the two groups. The achievement rate of first-time target SpO2 after NICU in the observation group was significantly higher than that in the control group [94.29% (33/35) vs. 26.32% (10/38), P < 0.05], the PaO2 control range was better [mmHg (1 mmHg = 0.133 kPa): 85.50±6.36 vs. 103.00±2.83, P < 0.05], manpower expenditure and time required for transportation were significantly reduced [manpower expenditure (number): 2.14±0.35 vs. 3.17±0.34, time required for transportation (minutes): 10.42±0.76 vs. 15.54±0.34, both P < 0.05].
CONCLUSIONS: The portable respiratory device for transporting premature infants is used for respiratory support during the transportation of extremely premature infants in primary hospitals. It can improve the achievement rate of target SpO2, control PaO2 within the target range, and avoid hypoxia or hyperoxia during transportation. The breathing apparatus is compact, easy to carry, can save labor resources and time during transport, is cost-effective, and is suitable for widespread application in primary hospitals.
参考来源
- Zhang L, Gu S, Zheng P, et al. [Design of portable respiratory device for transporting premature infants and application in the in-hospital transportation of extremely premature infants in primary hospitals]. Zhonghua Wei Zhong Bing Ji Jiu Yi Xue. 2025. PMID: 41017185. doi:10.3760/cma.j.cn121430-20240809-00676