无线心率监测仪(NeoBeat)与标准心电图在极早产儿复苏过程中心率监测准确性的比较

Comparative accuracy of a wireless heart rate monitor (NeoBeat) versus standard EKG during resuscitation of very premature neonates.
PubMed 文献 · Resuscitation · 2025-08-30
一句话结论在极早产儿复苏中,NeoBeat无线监测仪与标准三导联心电图心率高度相关(r=0.96),平均绝对差3.76次/分,支持其用于极早产儿(含250克)的实时监测。

本前瞻性观察研究在三级医疗中心纳入30例接受常规复苏的极早产儿(胎龄26.6±1.9周,出生体重范围250-1310克),同步用NeoBeat无线监测仪与标准三导联心电图记录心率,共分析1690对心率数据。结果显示两者强相关(Pearson r=0.96,Spearman ρ=0.95),平均绝对差3.76次/分;在不同复苏干预(插管、正压通气)亚组中相关性依然稳健(插管r=0.93 vs 未插管r=0.90,p=0.207;正压通气r=0.93 vs 无正压通气r=0.87,p=0.726)。作者认为NeoBeat在极早产儿中达到96%精确度(r=0.96)并保持3.76次/分的平均绝对差,支持其用于极早产儿实时监测的可靠性与准确性,但250克左右极低体重儿统计效能有限;建议开展随机对照研究比较NeoBeat与标准心电图以确认这些发现并探索NeoBeat是否改善临床结局。

要点

  • 极早产儿(胎龄26.6±1.9周)复苏中,NeoBeat无线监测仪与标准三导联心电图心率强相关(Pearson r=0.96,Spearman ρ=0.95),平均绝对差3.76次/分。
  • 共分析30例极早产儿、1690对心率数据,出生体重范围250-1310克,相关性在插管和正压通气等复苏干预亚组中保持稳健(插管r=0.93 vs 未插管r=0.90,p=0.207;正压通气r=0.93 vs 无正压通气r=0.87,p=0.726)。
  • 作者认为NeoBeat达到96%精确度(r=0.96),支持其用于极早产儿复苏实时监测的可靠性与准确性,包括体重低至250克的婴儿,但该体重范围统计效能有限。
  • 作者建议开展随机对照研究比较NeoBeat与标准心电图,以确认在极早产儿中的发现并探索NeoBeat是否改善临床结局。

适用人群与场景

  • 在三级医疗中心接受常规复苏的极早产儿(胎龄26.6±1.9周,出生体重范围250-1310克)。
  • 研究为前瞻性观察性研究,纳入30例极早产儿。

研究设计与关键结果

  • 研究类型:前瞻性观察性研究,单中心三级医疗中心(原文写明tertiary care center),样本量30例极早产儿(胎龄26.6±1.9周,出生体重范围250-1310克),共分析1690对心率数据。
  • 干预与对照:NeoBeat无线心率监测仪与标准三导联心电图同步记录心率。
  • 主要结果:NeoBeat与心电图心率强相关(Pearson r=0.96,Spearman ρ=0.95),平均绝对差3.76次/分。
  • 亚组结果:插管者r=0.93 vs 未插管者r=0.90(p=0.207);正压通气者r=0.93 vs 无正压通气者r=0.87(p=0.726),相关性均保持稳健。
  • 结论:作者认为NeoBeat达到96%精确度(r=0.96),平均绝对差3.76次/分,支持其用于极早产儿实时监测的可靠性与准确性,包括体重低至250克的婴儿,但该体重范围统计效能有限。

证据类型、等级与局限

  • 证据类型:前瞻性观察性研究(原文prospective observational study)。
  • 局限:作者指出在体重低至250克的婴儿范围统计效能有限,并建议开展随机对照研究以确认这些发现并探索NeoBeat是否改善临床结局。

对床旁实践的改变

  • 作者认为NeoBeat无线监测仪与标准三导联心电图在极早产儿复苏中心率高度相关,支持其用于极早产儿实时监测的可靠性与准确性,包括体重低至250克的婴儿。
  • 作者建议开展随机对照研究比较NeoBeat与标准心电图,以确认在极早产儿中的发现并探索NeoBeat是否改善临床结局。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

BACKGROUND: Rapid and accurate heart rate assessment is essential in neonatal resuscitation, particularly for very premature infants where delays can impact timely intervention. A wireless heart rate monitor (NeoBeat) device has been validated in term babies but not in smaller preterm infants. We sought to validate the use of the Neobeat device compared to standard EKG in extremely preterm neonates during resuscitation.

METHODS: A prospective observational study of very premature neonates undergoing routine resuscitation was conducted at a tertiary care center. Heart rate data were simultaneously recorded using the NeoBeat wireless monitor and a standard three-lead EKG.

RESULTS: Thirty very preterm infants (gestational age, (GA) 26.6 ± 1.9 weeks and birth weight range (250-1310 g)) were enrolled. A total of 1690 heart rate pairs were analyzed. A strong correlation and tracking observed between NeoBeat and EKG-derived heart rates (Pearson r = 0.96, Spearman ρ = 0.95), with a mean absolute difference of 3.76 beats per minute (bpm). Correlation remained robust across different resuscitation interventions (Intubation: r = 0.93 vs. No Intubation: r = 0.90, p = 0.207, PPV: r = 0.93 vs. No PPV: r = 0.87, p = 0.726).

CONCLUSION: In very preterm infants, NeoBeat wireless monitor achieved 96 % precision (r = 0.96) and maintained a mean absolute difference of 3.76 bpm compared to EKG, supporting its reliability and accuracy for real time neonatal monitoring, including infants weighing as little as 250 g, though statistical power in this range was limited. Randomized studies comparing NeoBeat to standard EKG are warranted to confirm these findings in extremely preterm infants and explore whether NeoBeat improves clinical outcomes.

来源:Resuscitation,查看原文。版权归原作者所有。