文献速递呼吸衰竭与机械通气重症综合管理与镇静镇痛

气管导管套囊压力管理模式的自对照研究

[A self-controlled study on endotracheal tube cuff pressure management modes].
PubMed 文献 · Zhonghua Wei Zhong Bing Ji Jiu Yi Xue · 2025-01-01
一句话结论动态监测套囊上方CO2压力指导套囊压力调节,可在较低套囊压力负荷下实现更好的套囊密封。

该前瞻性自对照研究纳入11例气管插管机械通气患者,在建立人工气道后24小时内随机顺序应用四种套囊压力管理模式各24小时:模式I自动套囊压力管理(安全范围20-35 cmH2O,每5分钟检测套囊上方CO2压力并自动调节)、模式II恒定25 cmH2O、模式III恒定30 cmH2O、模式IV护士每6-8小时手动测压维持25-30 cmH2O。每种模式检测264次,模式I总漏气和大漏气次数均显著低于模式II-IV;模式I套囊压力介于模式II和III之间,套囊压力>50 cmH2O的极端值次数显著低于模式II和III。结论:动态监测套囊上方CO2压力指导套囊压力调节可实现更好密封且套囊压力负荷相对较低。

要点

  • 模式I(自动套囊压力管理,安全范围20-35 cmH2O,每5分钟检测套囊上方CO2压力并自动调节)总漏气30例(11.36%),显著低于模式II 81例(30.68%)、模式III 70例(26.52%)、模式IV 103例(39.02%)(均P<0.05)。
  • 模式I大漏气15例(5.68%),显著低于模式II 50例(18.94%)、模式III 48例(18.18%)、模式IV 66例(25.00%)(均P<0.05)。
  • 模式II与模式III之间套囊漏气次数无显著差异,模式IV漏气最严重。
  • 模式I套囊压力为27.09(26.10,28.14)cmH2O,介于模式II 26.60(25.92,27.47)cmH2O和模式III 31.01(30.33,31.88)cmH2O之间(均P<0.01)。
  • 模式I套囊压力>50 cmH2O的极端值次数为19 900例(0.105%),显著低于模式II 22 297例(0.117%)和模式III 27 618例(0.145%)(均P<0.05)。

适用人群与场景

  • 2020年10月1日至2022年4月1日入住中国人民解放军总医院第二医学中心、接受气管插管和机械通气并使用自动气道管理系统(AGs)的11例患者。

诊断标准与关键阈值

  • 套囊漏气定义:套囊上方CO2压力=0为无漏气,0<CO2压力<2 mmHg为小漏气,CO2压力≥2 mmHg为大漏气。
  • 模式I套囊压力安全范围设定为20-35 cmH2O(1 cmH2O≈0.098 kPa)。

处置流程与药物剂量

  • 模式I(自动套囊压力管理):套囊压力安全范围设为20-35 cmH2O,AGs每5分钟自动检测套囊上方CO2压力以判断套囊密封状态,并自动调节套囊压力。
  • 模式II(恒定套囊压力25 cmH2O管理):AGs通过压力传感器监测套囊压力,并通过压力泵将套囊压力维持于25 cmH2O。
  • 模式III(恒定套囊压力30 cmH2O管理):AGs通过压力传感器监测套囊压力,并通过压力泵将套囊压力维持于30 cmH2O。
  • 模式IV(手动套囊压力管理):护士每6-8小时使用套囊压力表手动测量套囊压力,充气后使套囊压力保持在25-30 cmH2O。
  • 四种模式在建立人工气道和机械通气后24小时内按随机顺序依次对每例患者各应用24小时。

研究设计与关键结果

  • 前瞻性自对照研究,纳入11例气管插管机械通气患者,每例患者按随机顺序接受四种套囊压力管理模式各24小时。
  • 每例患者每种模式共进行24次CO2压力测量,每种模式共264次检测。
  • 总漏气次数:模式I 30例(11.36%) vs 模式II 81例(30.68%)、模式III 70例(26.52%)、模式IV 103例(39.02%),模式I均显著低于其余三组(均P<0.05)。
  • 大漏气次数:模式I 15例(5.68%) vs 模式II 50例(18.94%)、模式III 48例(18.18%)、模式IV 66例(25.00%),模式I均显著低于其余三组(均P<0.05)。
  • 模式II与模式III之间套囊漏气次数无显著差异,模式IV漏气最严重。
  • 套囊压力仅在模式I-III中分析(模式IV因需阻断套囊管与AGs管连接、AGs套囊压力管理模块未实际工作而无法实时监测套囊压力)。每例患者每种模式进行24小时套囊压力监测,每种模式监测19 008 000次。
  • 套囊压力:模式I 27.09(26.10,28.14)cmH2O,模式II 26.60(25.92,27.47)cmH2O,模式III 31.01(30.33,31.88)cmH2O;模式I介于模式II和III之间(均P<0.01)。
  • 套囊压力>50 cmH2O的极端值次数:模式I 19 900例(0.105%) vs 模式II 22 297例(0.117%)、模式III 27 618例(0.145%),模式I均显著低于模式II和III(均P<0.05)。

证据类型、等级与局限

  • 证据类型:前瞻性自对照研究。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 作者认为,动态监测套囊上方CO2压力以指导气管导管套囊压力调节,可在相对较低的套囊压力负荷下实现更好的套囊密封。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

OBJECTIVE: To explore the effects of different endotracheal tube cuff pressure management modes on cuff sealing and the pressure exerted on the tracheal wall.

METHODS: A prospective self-controlled study was conducted. Eleven patients undergoing endotracheal intubation and mechanical ventilation with an automatic airway management system (AGs) admitted to the Second Medical Centre of the Chinese People's Liberation Army General Hospital from October 1, 2020, to April 1, 2022, were enrolled as the study subjects. Within 24 hours after the establishment of artificial airway and mechanical ventilation, four cuff pressure management modes were randomly applied to each patient for 24 hours in sequence: automatic cuff pressure management mode [modeI: the safe range of cuff pressure was set at 20-35 cmH2O (1 cmH2O≈0.098 kPa), and the CO2 pressure above the endotracheal tube cuff was automatically detected by AGs every 5 minutes to determine the cuff sealing status, and the cuff pressure was automatically adjusted], constant cuff pressure (25 cmH2O) management mode (mode II: the cuff pressure was monitored by AGs through a pressure sensor, and the cuff pressure was maintained at 25 cmH2O via a pressure pump), constant cuff pressure (30 cmH2O) management mode (mode III: the cuff pressure was monitored by AGs through a pressure sensor, and the cuff pressure was maintained at 30 cmH2O via a pressure pump), and manual cuff pressure management mode (mode IV: the cuff pressure was manually measured by nurses every 6-8 hours using a cuff pressure gauge to keep the cuff pressure at 25-30 cmH2O after inflation). The CO2 pressure above the endotracheal tube cuff (at 60-minute intervals) and the cuff pressure changes (at 50-ms intervals) were recorded to compare the differences in number of cuff leaks [no leak was defined as CO2 pressure = 0, small leak as 0 < CO2 pressure < 2 mmHg (1 mmHg≈0.133 kPa), and large leak as CO2 pressure ≥ 2 mmHg] and cuff pressure among modesI-IV.

RESULTS: A total of 24 CO2 pressure measurements were taken per patient across the four modes, resulting in a total of 264 detections for each mode. Regarding the cuff leak, the total number of leak and large leak in modeIwas significantly lower than that in modes II-IV [total leak: 30 cases (11.36%) vs. 81 cases (30.68%), 70 cases (26.52%), 103 cases (39.02%); large leak: 15 cases (5.68%) vs. 50 cases (18.94%), 48 cases (18.18%), 66 cases (25.00%), all P < 0.05]. There was no significant difference in the number of cuff leak between modes II and III, and mode IV had the most severe cuff leak. In terms of cuff pressure, since mode IV required blocking the cuff tube from the AGs tube and the AGs cuff pressure management module did not actually work, real-time monitoring of cuff pressure was not possible. Therefore, cuff pressure changes were only analyzed in modes I-III. Each of the 11 patients underwent 24-hour cuff pressure monitoring under modes I-III, with 19 008 000 monitoring times for each mode. The cuff pressure in mode I was between that in modes II and III [cmH2O: 27.09 (26.10, 28.14) vs. 26.60 (25.92, 27.47), 31.01 (30.33, 31.88), both P < 0.01]. Moreover, the number of extreme values of cuff pressure > 50 cmH2O in mode I was significantly lower than that in modes II and III [19 900 cases (0.105%) vs. 22 297 cases (0.117%), 27 618 cases (0.145%), both P < 0.05].

CONCLUSION: Dynamically monitoring the CO2 pressure above the cuff to guide the adjustment of endotracheal tube cuff pressure can achieve better cuff sealing with a relatively lower cuff pressure load.

来源:Zhonghua Wei Zhong Bing Ji Jiu Yi Xue,查看原文。版权归原作者所有。