用无创方法探索脑脊液代偿带
本研究纳入61例接受脑室颅内压(ICP)监测的颅脑创伤患者,同步用外部颅骨微动力学传感器(brain4care)记录替代波形参数P2/P1比值和达峰时间(TTP),并按有创RAP指数将患者分为颅内顺应性(ICC)充足、受损、耗竭三组。结果显示三组的P2/P1比值和TTP随ICC恶化而升高,无创参数与RAP指数分级的关联具有统计学意义(p<0.0001)。作者认为这些无创指标可能作为ICC状态的可靠指标,按RAP确定的临界ICP为19.45 mmHg,低于现行TBI指南升级治疗的阈值,但尚需进一步前瞻性验证。
要点
- 61例颅脑创伤脑室ICP监测患者中,ICC充足组P2/P1比值1.06±0.3、TTP 0.18±0.09 s,ICC受损组P2/P1比值1.15±0.32、TTP 0.23±0.07 s,ICC耗竭组P2/P1比值1.31±0.26、TTP 0.25±0.05 s。
- 三组ICP中位数分别为ICC充足组12.3±5.4 mmHg、ICC受损组13±6.4 mmHg、ICC耗竭组19.45±5.9 mmHg。
- 无创参数与RAP指数-based ICC分级的回归分析关联具有统计学意义(p<0.0001)。
- 按RAP确定的临界ICP为19.45 mmHg,低于现行TBI指南升级治疗的阈值。
- 作者认为该方法可能促进ICC恶化前更早干预并实现无创监测,但尚需进一步前瞻性验证。
适用人群与场景
- 接受脑室颅内压(ICP)监测的颅脑创伤患者。
诊断标准与关键阈值
- 按有创RAP指数将患者分为三组:ICC充足、ICC受损、ICC耗竭。
- 按RAP确定的临界ICP为19.45 mmHg。
研究设计与关键结果
- 单中心研究,共纳入61例颅脑创伤脑室ICP监测患者,同步用外部颅骨微动力学传感器(brain4care)记录P2/P1比值和TTP,并按RAP指数分组分析。
- ICC充足组:ICP中位数12.3±5.4 mmHg,P2/P1比值1.06±0.3,TTP 0.18±0.09 s。
- ICC受损组:ICP中位数13±6.4 mmHg,P2/P1比值1.15±0.32,TTP 0.23±0.07 s。
- ICC耗竭组:ICP中位数19.45±5.9 mmHg,P2/P1比值1.31±0.26,TTP 0.25±0.05 s。
- 回归分析显示无创参数与RAP指数-based ICC分级之间的关联具有统计学意义(p<0.0001)。
- 按RAP确定的临界ICP为19.45 mmHg,低于现行TBI指南升级治疗的阈值。
证据类型、等级与局限
- 证据类型:临床试验(注册号NCT03144219)。
- 局限:原文报告尚需进一步前瞻性验证。
对床旁实践的改变
- 作者认为,无创ICP波形衍生参数(P2/P1比值和TTP)可能作为ICC状态的可靠指标。
- 作者认为,该方法有潜力促进ICC恶化前更早干预并实现无创监测,可能改善神经重症预后,但尚需进一步前瞻性验证。
提炼说明
- 本条依据论文摘要整理;摘要未涉及「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开
BACKGROUND: Intracranial compliance (ICC) reflects the balance among intracranial volume components. Recent technological advances enable continuous, noninvasive assessment of ICC in neurocritical care settings. In this study, we aimed to correlate noninvasive ICC parameters derived from intracranial pressure (ICP) waveform morphology with the established amplitude-pressure index (RAP index), which is calculated using invasive ICP monitoring.
METHODS: Patients with traumatic brain injury underwent ventricular ICP monitoring. Simultaneously, ICP values and waveform characteristics were recorded using an external skull microdynamics sensor (brain4care) that provides surrogate waveform parameters, including the P2/P1 ratio and time-to-peak (TTP). The RAP index was calculated using dedicated software based on ICP values and pulse amplitude and was used to categorize patients into three groups: (1) adequate ICC, (2) compromised ICC, and (3) exhausted ICC. Noninvasive parameters (P2/P1 ratio and TTP) were then analyzed in relation to RAP index groupings.
RESULTS: A total of 61 patients were included. Group 1 (adequate ICC) had a median ICP of 12.3 ± 5.4 mm Hg, a P2/P1 ratio of 1.06 ± 0.3, and a TTP of 0.18 ± 0.09 s. Group 2 (compromised ICC) had a median ICP of 13 ± 6.4 mm Hg, a P2/P1 ratio of 1.15 ± 0.32, and a TTP of 0.23 ± 0.07 s. Group 3 (exhausted ICC) had a median ICP of 19.45 ± 5.9 mm Hg, a P2/P1 ratio of 1.31 ± 0.26, and a TTP of 0.25 ± 0.05 s. Regression analysis revealed a statistically significant association between the noninvasive parameters and RAP index-based ICC classification (p < 0.0001).
CONCLUSIONS: This study demonstrates a significant correlation between the RAP index and noninvasive ICP waveform-derived parameters, such as the P2/P1 ratio and TTP. These findings suggest that such noninvasive measures may serve as reliable indicators of ICC status. The critical ICP cut-off per RAP was 19.45 mmHg, below the current threshold for therapy escalation according to TBI guidelines. Although further prospective validation is required, this approach has the potential to facilitate earlier intervention before ICC deterioration and enable noninvasive monitoring, possibly improving outcomes in neurocritical care.
TRIAL REGISTRATION: NCT03144219. Registered 15 June 2017, http://www.
CLINICALTRIALS: gov/NCT03144219 .
CLINICAL TRIAL REGISTRATION: ClinicalTrials.gov identifier: NCT03144219.