气道闭合时食管压校准的错误

Erroneous calibration of esophageal pressure in case of airway closure.
PubMed 文献 · Crit Care · 2025-05-02
一句话结论气道闭合且总PEEP低于气道开放压时,呼气末正压阻断试验会明显低估ΔPaw/ΔPes;作者认为应在吸气末阻断时行正压阻断试验校准食管球囊。

研究在12具存在呼气末气道闭合的人类尸体上,于不同呼气末正压(PEEP)水平和吸气末进行食管球囊校准(正压阻断试验测ΔPaw/ΔPes)。总PEEP低于气道开放压(AOP)时,呼气末与吸气末ΔPaw/ΔPes估计均值分别为0.42 [0.40; 0.44] vs 0.95 [0.92; 0.97],P < 0.001;总PEEP高于AOP时分别为0.99 [0.92; 1.05] vs 0.99 [0.92; 1.06],P = 0.854。两例临床需食管压监测的患者的球囊校准观察佐证了上述结果。作者认为气道闭合时胸廓按压不能完全传至气道,在AOP以下行校准试验会明显低估ΔPaw/ΔPes,主张将吸气末阻断时的正压阻断试验作为食管球囊定位与校准的新标准操作。

要点

  • 气道闭合时,总PEEP低于AOP条件下呼气末ΔPaw/ΔPes估计均值0.42 [0.40; 0.44],明显低于吸气末的0.95 [0.92; 0.97],P < 0.001。
  • 总PEEP高于AOP时呼气末与吸气末ΔPaw/ΔPes估计均值分别为0.99 [0.92; 1.05] vs 0.99 [0.92; 1.06],P = 0.854,无显著差异。
  • 气道闭合时胸廓按压不能完全传至气道,可导致ΔPaw/ΔPes明显低估,使该监测技术在AOP以下行试验时可靠性差。
  • 作者认为应在吸气末阻断期间行正压阻断试验,作为食管球囊定位与校准的新标准操作。
  • 机械通气合并急性呼吸窘迫综合征、肥胖、静水压性肺水肿及心肺复苏期间可发生气道闭合,影响食管压校准。

适用人群与场景

  • 接受机械通气的急性呼吸窘迫综合征、肥胖、静水压性肺水肿患者及心肺复苏期间患者,这些情况下可发生气道闭合。
  • 研究校准试验在12具存在呼气末气道闭合的人类尸体上进行,并在两例临床需食管压监测的患者中观察佐证。

诊断标准与关键阈值

  • 气道闭合定义为近端与远端气道之间缺乏沟通,除非气道压(Paw)超过气道开放压(AOP)。
  • 校准试验采用正压阻断试验测ΔPaw/ΔPes,分别在呼气末和吸气末、不同水平PEEP下进行。

处置流程与药物剂量

  • 作者认为应在吸气末阻断期间行正压阻断试验,作为食管球囊定位与校准的新标准操作。
  • 总PEEP低于AOP时,呼气末校准试验会低估ΔPaw/ΔPes,可靠性差。

红旗征与不能漏的情况

  • 气道闭合且总PEEP低于AOP时,呼气末校准试验可明显低估ΔPaw/ΔPes,导致食管压监测不可靠。

研究设计与关键结果

  • 研究类型为在12具人类尸体上进行的校准试验观察,并在两例临床患者中佐证;人群为存在呼气末气道闭合的尸体。
  • 总PEEP低于AOP时,呼气末与吸气末ΔPaw/ΔPes估计均值分别为0.42 [0.40; 0.44] vs 0.95 [0.92; 0.97],P < 0.001。
  • 总PEEP高于AOP时,呼气末与吸气末ΔPaw/ΔPes估计均值分别为0.99 [0.92; 1.05] vs 0.99 [0.92; 1.06],P = 0.854。
  • 两例需食管压监测的临床患者的食管球囊校准观察佐证了上述结果。

证据类型、等级与局限

  • 证据类型:在12具人类尸体上进行的校准试验观察,附两例临床患者观察佐证。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 作者认为气道闭合时在AOP以下行呼气末校准试验会低估ΔPaw/ΔPes,使食管压监测可靠性差。
  • 作者认为应将吸气末阻断期间的正压阻断试验作为食管球囊定位与校准的新标准操作。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

Airway closure results in a lack of communication between proximal and distal airways unless the airway pressure (Paw) overcomes the airway opening pressure (AOP). This has been described in patients undergoing mechanical ventilation with acute respiratory distress syndrome, obesity, hydrostatic pulmonary edema and during cardiopulmonary resuscitation. In these categories of patients, esophageal pressure (Pes) can guide the personalization of mechanical ventilation and calibration of the esophageal balloon is necessary to obtain reliable Pes measurements. The impact of airway closure has never been envisaged. This study investigated the impact of airway closure on the calibration of the esophageal balloon by the ∆Paw/∆Pes following a positive pressure occlusion test during passive mechanical ventilation. The calibration test was performed in twelve human cadavers with airway closure at end-expiration at different levels of positive end-expiratory pressure (PEEP) and at end-inspiration. The ∆Paw/∆Pes measured at end-expiration and at end-inspiration were significantly different when total PEEP was lower than AOP (estimated means 0.42 [0.40; 0.44] vs. 0.95 [0.92; 0.97], P < 0.001), while this difference was not observed when total PEEP was higher than AOP (estimated means 0.99 [0.92; 1.05] vs. 0.99 [0.92; 1.06], P = 0.854). These results were corroborated by observations during esophageal balloon calibration in two patients requiring Pes monitoring for clinical management. In case of airway closure, compression of the chest is not fully transmitted to the airways. This can lead to a conspicuous underestimation of the ∆Paw/∆Pes and poor reliability of this monitoring technique when the test takes place below AOP. Our results favor a positive pressure occlusion test performed during an end-inspiratory occlusion as the new standard of operative procedures for positioning and calibrating the esophageal balloon.

来源:Crit Care,查看原文。版权归原作者所有。