院外心脏骤停心室颤动:递增能量与固定能量除颤对比
该回顾性队列研究纳入中国48个城市2017至2023年342例院外心脏骤停且至少接受1次除颤的患者,共782次室颤除颤,比较递增高能量(200-300-360 J)与固定低能量(200-200-200 J)双相波自动体外除颤器方案。首次200 J除颤时两组短暂终止率无显著差异(94% vs 93%,P=0.64),但维持至下次心律分析的比例均仅约半数(49% vs 47%,P=0.68)。递增高能量组建立持续 organized rhythm 的比例更高(34% vs 25%,P=0.008,绝对差9%[95%CI 2%-16%]);难治性室颤中,第二次及以上除颤后(24% vs 13%,P=0.008,绝对差11%[95%CI 3%-19%])和第三次及以上除颤后(35% vs 18%,P=0.003,绝对差17%[95%CI 5%-27%])递增高能量组持续 organized rhythm 建立率均显著更高。作者认为递增高能量方案在所有除颤后与更好结局相关,尤其在难治性室颤患者中。
要点
- 首次200 J除颤时,递增高能量组与固定低能量组短暂终止室颤率无显著差异(94% vs 93%,P=0.64)。
- 首次除颤后仅约半数室颤维持终止至下次心律分析(递增组49% vs 固定组47%,P=0.68)。
- 递增高能量组建立持续 organized rhythm 的比例高于固定低能量组(34% vs 25%,P=0.008,绝对差9%[95%CI 2%-16%])。
- 难治性室颤中,第二次及以上除颤后递增组持续 organized rhythm 建立率24% vs 固定组13%(P=0.008,绝对差11%[95%CI 3%-19%])。
- 难治性室颤中,第三次及以上除颤后递增组持续 organized rhythm 建立率35% vs 固定组18%(P=0.003,绝对差17%[95%CI 5%-27%])。
适用人群与场景
- 中国48个城市2017至2023年院外心脏骤停心室颤动且至少接受1次除颤的患者,共342例(平均[SD]年龄57.2[20.6]岁;男性273例[79.8%])。
处置流程与药物剂量
- 递增高能量方案:首次200 J,第二次300 J,第三次360 J,依据可用的自动体外除颤器程序选择。
- 固定低能量方案:每次均为200 J(200-200-200 J),依据可用的自动体外除颤器程序选择。
研究设计与关键结果
- 研究类型:回顾性队列研究,纳入中国48个城市342例院外心脏骤停心室颤动患者,共782次室颤除颤。
- 分组:218例患者(63.8%)共480次室颤(61.4%)接受递增高能量方案自动体外除颤器。
- 首次200 J除颤短暂终止室颤:递增高能量组94% vs 固定低能量组93%(P=0.64)。
- 维持终止至下次心律分析:递增高能量组49% vs 固定低能量组47%(P=0.68)。
- 建立持续 organized rhythm:递增高能量组34% vs 固定低能量组25%(P=0.008;绝对差9%[95%CI 2%-16%])。
- 难治性室颤第二次及以上除颤后建立持续 organized rhythm:递增高能量组24% vs 固定低能量组13%(P=0.008;绝对差11%[95%CI 3%-19%])。
- 难治性室颤第三次及以上除颤后建立持续 organized rhythm:递增高能量组35% vs 固定低能量组18%(P=0.003;绝对差17%[95%CI 5%-27%])。
证据类型、等级与局限
- 证据类型:回顾性队列研究。
- 局限:原文未报告局限性。
对床旁实践的改变
- 作者认为,递增高能量方案(200-300-360 J)在所有除颤后与更好结局相关,尤其在难治性室颤患者中;固定低能量方案(200-200-200 J)在首次除颤短暂终止室颤方面同样有效。
提炼说明
- 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开
IMPORTANCE: There is limited evidence on whether higher-energy defibrillation is preferred in patients experiencing out-of-hospital cardiac arrest (OHCA) with shockable rhythms.
OBJECTIVE: To investigate the optimal energy regimen for initial and subsequent defibrillation delivered by biphasic waveform automated external defibrillators (AEDs) in OHCA ventricular fibrillation (VF).
DESIGN, SETTING, AND PARTICIPANTS: This cohort study was conducted in 48 cities across China, from 2017 to 2023, among 342 patients with OHCA who experienced at least 1 shock.
EXPOSURES: Escalating higher-energy (200-300-360 J) defibrillation or fixed low-energy (200-200-200 J) defibrillation according to the AED program available for use.
MAIN OUTCOME AND MEASURES: Sustained and transient termination of VF and establishment of an organized rhythm after defibrillations were the main clinical outcomes.
RESULTS: A total of 342 patients with OHCA were included (mean [SD] age, 57.2 [20.6] years; 273 male [79.8%]) with 782 VF defibrillations; 218 patients (63.8%) with a total of 480 instances (61.4%) of VF rhythm received AED with escalating higher-energy regimens. Most VF episodes were effectively terminated transiently at the first shock (200 J in both groups) (94% in the escalating higher-energy group vs 93% in the fixed lower-energy group; P = .64), but only half remained terminated until the next rhythm analysis (49% vs 47%; P = .68). Comparatively, VF that received escalating higher-energy regimens were more likely to establish sustained organized rhythm (34% vs 25%; P = .008; absolute difference, 9% [95% CI, 2% to 16%]). In refractory VF rhythms, the percentage of cases where sustained organized rhythms were established was significantly greater in escalating higher-energy regimens after second shocks and above (24% vs 13%; P = .008; absolute difference, 11% [95% CI, 3% to 19%]) and third shocks and above (35% vs 18%; P = .003; absolute difference 17% [95% CI, 5% to 27%]).
CONCLUSIONS AND RELEVANCE: In this retrospective cohort study of patients experiencing OHCA-VF, both the escalating higher-energy (200-300-360 J) regimen and the fixed low-energy (200-200-200 J) regimen were effective for transient VF termination at first shock, whereas the escalating higher-energy regimens were more likely to maintain termination and restore an organized rhythm. Higher-energy regimens were associated with better outcomes after all shocks, especially in patients with refractory VF.
参考来源
- Tang H, Wu R, Yin L, et al. Escalating vs Fixed Energy Defibrillation in Out-of-Hospital Cardiac Arrest Ventricular Fibrillation. JAMA Netw Open. 2025. PMID: 40299385. doi:10.1001/jamanetworkopen.2025.7411
供执业医师抢救与诊疗参考,不替代临床判断;剂量请结合体重、肝肾功能和本院药品规格复核,以指南原文为准。