高分辨率直流脑电图无创检测皮层扩散性去极化的概念验证及未来设备建议

Proof-of-Concept Validation of Noninvasive Detection of Cortical Spreading Depolarization with High Resolution Direct Current-Electroencephalography with Future Device Recommendations.
指南 / 共识 · medRxiv · 2025-05-10
一句话结论在1例蛛网膜下腔出血患者中,头皮高密度直流脑电图阵列检测的直流偏移与侵入性皮层脑电图记录的扩散性去极化时间进程高度一致(24/26个可读偏移决定系数>0.80)。

本研究为概念验证研究,使用29个6 mm直径Ag/AgCl电极(间距1 cm)的热塑性弹性体阵列贴附于前额,在正常志愿者和神经重症监护病房的严重急性脑损伤患者中记录直流脑电图,并与金标准皮层脑电图对比。在1例蛛网膜下腔出血患者11天记录到的140个皮层脑电图编码的扩散性去极化中,26个位于可读、无伪差的头皮记录段,其中24个头皮与皮层脑电图的直流偏移时间进程决定系数>0.80(中位数0.98,IQR 0.02)。头皮直流电位热图显示5 cm范围、-460 μV峰值区域持续约70秒。作者认为无创扩散性去极化检测可行,但高伪差负担和低可用性记录是限制性阴性结果,为未来重新设计设备提供了建议。

要点

  • 在1例蛛网膜下腔出血患者11天记录中,皮层脑电图编码140个扩散性去极化,其中26个位于可读、无伪差头皮记录段。
  • 26个可读直流偏移中24个头皮与皮层脑电图时间进程决定系数>0.80,中位数0.98(IQR 0.02)。
  • 头皮直流电位热图重建显示5 cm范围、-460 μV峰值区域持续约70秒;头皮直流偏移峰值-457±69 μV(均值±标准差),半高全宽70.9±5.92秒,面积18.7±2.76 cm²。
  • 高伪差负担和低可用性记录掩盖了该患者大部分头皮原型设备数据,为限制性阴性结果。
  • 作者建议未来重新设计设备以减少伪差、提高可用性,从而实现直流脑电图扩散性去极化检测。

适用人群与场景

  • 正常志愿者和神经重症监护病房的严重急性脑损伤患者,部分有、部分无侵入性皮层脑电图电极。

诊断标准与关键阈值

  • 扩散性去极化在皮层脑电图上根据直流偏移伴高频抑制和传播进行评分。
  • 头皮与皮层脑电图直流偏移时间进程相似性通过六参数高斯加二次基线模型拟合参数比较,决定系数>0.80确认相似。

研究设计与关键结果

  • 研究类型:概念验证验证研究;样本:正常志愿者和神经重症监护病房严重急性脑损伤患者,其中1例蛛网膜下腔出血患者记录了11天。
  • 设备:83×58 mm热塑性弹性体阵列,29个6 mm直径Ag/AgCl电极,间距1 cm,即CerebroPatch™概念验证原型,粘附于前额并加电极凝胶界面记录直流脑电图。
  • 头皮和皮层脑电图电压由Moberg® Advanced ICU Amplifier采集;伪差视觉识别并记录可用性问题。
  • 该蛛网膜下腔出血患者11天记录到140个皮层脑电图编码的扩散性去极化,其中26个位于可读、无伪差头皮记录段,24个头皮与皮层脑电图直流偏移决定系数>0.80(中位数0.98,IQR 0.02)。
  • 头皮直流电位热图重建显示5 cm范围、-460 μV峰值区域持续约70秒;头皮直流偏移峰值-457±69 μV(均值±标准差),半高全宽70.9±5.92秒,面积18.7±2.76 cm²。
  • 作者认为这些头皮直流偏移代表无创检测到的扩散性去极化。

证据类型、等级与局限

  • 证据类型:概念验证研究。
  • 局限:高伪差负担数据和低可用性记录是限制性阴性结果;作者建议未来重新设计设备以减少伪差、提高可用性。

对床旁实践的改变

  • 作者认为使用高空间分辨率脑电图阵列的头皮直流电位信号可实现无创扩散性去极化检测。
  • 作者建议未来重新设计设备以减少伪差、提高可用性,从而实现直流脑电图扩散性去极化检测,以支持继发性脑损伤的多模态监测。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

BACKGROUND/OBJECTIVE: Cortical spreading depolarization (SD) is increasingly recognized as a major contributor to secondary brain injury. Noninvasive SD monitoring would enable the institution of SD-based therapeutics. Our primary objective is to establish proof-of-concept validation that scalp DC-potentials can provide noninvasive SD detection by comparing scalp direct-current (DC)-shifts from a high-density electrode array to SDs detected by gold-standard electrocorticography (ECoG). Our secondary objective is to assess usability and artifact tolerance.

METHODS: An 83×58 mm thermoplastic elastomer array with 29 6-mm diameter Ag/AgCl 1-cm spaced electrodes, the CerebroPatch™ Proof-of-Concept Prototype, was adhesively placed on the forehead with an intervening electrode gel interface to record DC-electroencephalography in normal volunteers and severe acute brain injury patients in the neuro-intensive care unit some with and some without invasive ECoG electrodes. The scalp and ECoG voltages were collected by a Moberg® Advanced ICU Amplifier. Artifacts were visually identified and usability issues were recorded. SD was scored on ECoG based on DC-shifts with associated high-frequency suppression and propagation. A six-parameter Gaussian plus quadratic baseline model was used to estimate ECoG and scalp electrode time-courses and scalp-voltage heat-map movies. The similarity of the noninvasive scalp and invasive ECoG DC-shift time-courses was compared via the Gaussian fit parameters and confirmed if the Coefficient-of-Determination was >0.80.

RESULTS: Usability and artifact issues obscured most scalp Prototype device data of the 140 ECoG-coded SDs during 11 days in one sub-arachnoid hemorrhage patient. Twenty-six of these DC-shifts were in readable, artifact-free portions of scalp recordings and 24 of these had a >0.80 Coefficient-of-Determination (0.98[0.02], median[IQR]) between invasive ECoG and noninvasive Prototype device DC-shifts. Reconstructed heat-map movies of the scalp DC-potentials showed a 5-cm extent, -460 μV peak region that persisted for ~70 sec. These data suggest that these scalp DC-shifts (peak -457±69 μV [mean±StD], full-width-half maximum 70.9±5.92 sec, area 18.7±2.76 cm2) depicted in the heat-map movies represent noninvasively detected SDs.

CONCLUSIONS: These results using 26 SDs as the observational units suggest that noninvasive SD detection is possible using scalp DC-potential signals with a high spatial resolution EEG array. Although the high artifact burden data and low usability records were limiting, negative results, they serve as an important entrepreneurial recipe that provides suggestions for a future, re-designed device that would reduce artifacts and improve usability for DC-EEG SD detection needed to enable multi-modal monitoring for secondary brain injury.

来源:medRxiv,查看原文。版权归原作者所有。