文献速递呼吸衰竭与机械通气重症综合管理与镇静镇痛

电阻抗断层成像区域通气动态与四维计算机断层扫描的验证:单中心前瞻性观察研究

Regional ventilation dynamics of electrical impedance tomography validated with four-dimensional computed tomography: single-center, prospective, observational study.
PubMed 文献 · Crit Care · 2024-10-16
一句话结论在被动机械通气成人中,电阻抗断层成像各区域动态通气与四维计算机断层扫描高度相关,但区域通气分布<2%时相关性降低。

本研究为单中心、前瞻性、观察性研究,于2021年7月至2024年2月在综合重症监护病房纳入接受机械通气的成人患者,患者接受被动机械通气并在同一天进行电阻抗断层成像与四维计算机断层扫描。共分析45例患者,右肺腹侧、中央、背侧区域动态通气的相关系数分别为0.963、0.963、0.835,左肺分别为0.947、0.927、0.823。当电阻抗断层成像检测的区域通气分布<2%时相关系数较低;排除9例区域通气分布<2%的患者后,右肺腹侧、中央、背侧相关系数分别为0.963、0.963、0.946,左肺分别为0.942、0.924、0.951。作者认为电阻抗断层成像在动态通气中的区域通气与四维计算机断层扫描高度准确且时相一致,两者高度相关可为进一步研究区域通气动态提供帮助。

要点

  • 45例被动机械通气成人同日行电阻抗断层成像与四维计算机断层扫描,右肺腹侧、中央、背侧区域动态通气相关系数分别为0.963、0.963、0.835。
  • 左肺腹侧、中央、背侧区域动态通气相关系数分别为0.947、0.927、0.823。
  • 电阻抗断层成像检测的区域通气分布<2%时相关系数较低。
  • 排除9例区域通气分布<2%的患者后,右肺腹侧、中央、背侧相关系数分别为0.963、0.963、0.946,左肺分别为0.942、0.924、0.951。
  • 作者认为两种方法高度相关,可为进一步研究区域通气动态提供帮助。

适用人群与场景

  • 2021年7月至2024年2月在一家综合重症监护病房接受机械通气的成人患者,接受被动机械通气。

研究设计与关键结果

  • 单中心、前瞻性、观察性研究,共分析45例接受被动机械通气的成人患者,同日进行电阻抗断层成像与四维计算机断层扫描。
  • 右肺腹侧、中央、背侧区域动态通气相关系数分别为0.963、0.963、0.835;左肺分别为0.947、0.927、0.823。
  • 电阻抗断层成像检测的区域通气分布<2%时相关系数较低。
  • 排除9例区域通气分布<2%的患者后,右肺腹侧、中央、背侧相关系数分别为0.963、0.963、0.946,左肺分别为0.942、0.924、0.951。

证据类型、等级与局限

  • 单中心、前瞻性、观察性研究。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 作者认为电阻抗断层成像在动态通气中的区域通气与四维计算机断层扫描高度准确且时相一致,两者高度相关可为进一步研究区域通气动态提供帮助。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

BACKGROUND: The dynamic regional accuracy of electrical impedance tomography has not yet been validated. We aimed to compare the regional accuracy of electrical impedance tomography with that of four-dimensional computed tomography during dynamic ventilation.

METHODS: This single-center, prospective, observational study conducted in a general intensive care unit included adult patients receiving mechanical ventilation from July 2021 to February 2024. The patients were mechanically ventilated passively and underwent electrical impedance tomography and four-dimensional computed tomography on the same day.

RESULTS: Overall, 45 patients were analyzed. The correlation coefficients in regional dynamic ventilation between four-dimensional computed tomography and electrical impedance tomography in each region were 0.963, 0.963, 0.835 (ventral, central, and dorsal, respectively) in the right lung and 0.947, 0.927, 0.823 (ventral, central, and dorsal, respectively) in the left lung. The correlation coefficient was low when the regional ventilation distribution detected by the electrical impedance tomography was < 2%. After excluding nine patients with a regional ventilation distribution of < 2%, the ventral, central, and dorsal correlation coefficients were 0.963, 0.963, and 0.946 in the right lung and 0.942, 0.924, and 0.951, respectively, in the left lung.

CONCLUSIONS: Regional ventilation using electrical impedance tomography during dynamic ventilation was highly accurate and consistent with the time phase compared to four-dimensional computed tomography. Given the high correlation between these modalities, they can contribute significantly to further studies on regional ventilation dynamics. Trial registration number ClinicalTrials.gov (No. UMIN00044386).

来源:Crit Care,查看原文。版权归原作者所有。