院前危重症环境中无创血压与动脉血压监测的配对比较

Non-invasive versus arterial pressure monitoring in the pre-hospital critical care environment: a paired comparison of concurrently recorded measurements.
PubMed 文献 · Scand J Trauma Resusc Emerg Med · 2024-08-27
一句话结论院前危重症患者无创血压常不准确,血流动力学不稳定时尤甚;临床医生解读NIBP应谨慎,条件允许时考虑直接动脉压监测。

本研究为回顾性队列研究,纳入2020年5月1日至2023年4月30日期间由院前危重症团队治疗、同时记录无创血压(NIBP)与有创动脉血压(IBP)的患者,共221例、2359对配对测量,中位年龄57岁,最常见转运原因为心脏骤停(79例,35.7%)。研究预设可接受差异为收缩压(SBP)和舒张压(DBP)<20mmHg、平均动脉压(MAP)<10mmHg,主要结局为配对一致率。Bland-Altman分析显示一致性界限过宽,低血压时NIBP高估SBP和MAP、高血压时低估。血流动力学休克(SBP<90mmHg)与SBP(aOR 0.52,95%CI 0.35-0.77)、DBP(aOR 0.65,95%CI 0.42-0.99)、MAP(aOR 0.53,95%CI 0.36-0.78)配对一致率降低独立相关;严重高血压(SBP>160mmHg)与SBP配对一致率降低独立相关(aOR 0.17,95%CI 0.11-0.27);转运状态与SBP、DBP、MAP的一致率无关联。结论认为院前危重症环境中NIBP常不准确,尤其血流动力学不稳定患者,临床医生应谨慎解读并考虑直接动脉压监测。

要点

  • 院前危重症患者中,NIBP与IBP配对一致率在血流动力学休克(SBP<90mmHg)时降低:SBP aOR 0.52(95%CI 0.35-0.77)、DBP aOR 0.65(95%CI 0.42-0.99)、MAP aOR 0.53(95%CI 0.36-0.78)。
  • 严重高血压(SBP>160mmHg)时SBP配对一致率降低(aOR 0.17,95%CI 0.11-0.27)。
  • Bland-Altman分析显示一致性界限过宽,低血压时NIBP高估SBP和MAP,高血压时低估。
  • 转运状态与SBP、DBP、MAP的配对一致率无关联。
  • 作者认为临床医生应谨慎解读NIBP,条件允许时考虑直接动脉压监测。

适用人群与场景

  • 2020年5月1日至2023年4月30日期间由院前危重症团队治疗、同时记录NIBP与有创动脉血压(IBP)的患者,共221例,中位年龄57岁。
  • 最常见转运原因为心脏骤停(79例,35.7%)。

诊断标准与关键阈值

  • 预设可接受差异:SBP和DBP<20mmHg,MAP<10mmHg。
  • 主要结局为配对一致率,即配对观察值落在可接受范围内的比例。
  • 血流动力学休克定义为SBP<90mmHg;严重高血压定义为SBP>160mmHg。

研究设计与关键结果

  • 回顾性队列研究,共221例患者、2359对配对测量,中位年龄57岁,最常见转运原因为心脏骤停(79例,35.7%)。
  • Bland-Altman分析显示一致性界限过宽,低血压时NIBP高估SBP和MAP,高血压时低估。
  • 血流动力学休克(SBP<90mmHg)与SBP(aOR 0.52,95%CI 0.35-0.77)、DBP(aOR 0.65,95%CI 0.42-0.99)、MAP(aOR 0.53,95%CI 0.36-0.78)配对一致率降低独立相关。
  • 严重高血压(SBP>160mmHg)与SBP配对一致率降低独立相关(aOR 0.17,95%CI 0.11-0.27)。
  • 转运状态与SBP、DBP、MAP的配对一致率无关联。

证据类型、等级与局限

  • 证据类型:回顾性队列研究。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 作者认为,院前危重症环境中NIBP测量常不准确,尤其血流动力学不稳定患者,临床医生应谨慎解读NIBP测量值。
  • 作者认为,条件允许时应考虑直接动脉压监测。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

BACKGROUND: Blood pressure monitoring is important in the pre-hospital management of critically ill patients. Non-invasive blood pressure (NIBP) measurements are commonly used but the accuracy of standard oscillometric cuff devices may be affected by extremes of physiology and adverse conditions (e.g. vibration) during transport. This study aimed to quantify the accuracy of NIBP measurements amongst patients requiring pre-hospital critical care.

METHODS: A retrospective cohort study was undertaken using data from patients treated by a pre-hospital critical team between 1st May 2020 and 30th April 2023 that had NIBP measured concurrently with invasive blood pressure (IBP) arterial manometry. An acceptable difference was determined a priori to be < 20mmHg for systolic blood pressure (SBP) and diastolic blood pressure (DBP), and < 10mmHg for mean arterial pressure (MAP). The primary outcome was "pairwise agreement", i.e. the proportion of paired observations that fell within this range of acceptability. Bland-Altman plots were constructed together with 95% limits of agreement to visualise differences between pairs of data. Associations with patient age, reason for critical care, transport status, haemodynamic shock, severe hypertension, and arterial catheter position were explored in univariate analyses and by fitting multivariable logistic regression models.

RESULTS: There were 2,359 paired measurements from 221 individual patients with a median age of 57. The most frequent reason for transport was cardiac arrest (79, 35.7%). Bland-Altman analyses suggested unacceptably wide limits of agreement with NIBP overestimating both SBP and MAP during hypotension and underestimating these values during hypertension. Haemodynamic shock (SBP < 90mmHg) was independently associated with reduced pairwise agreement for SBP (adjusted odds ratio [aOR] 0.52, 95% CI 0.35 to 0.77), DBP (aOR 0.65, 95% CI 0.42 to 0.99) and MAP (aOR 0.53, 95% CI 0.36 to 0.78) and severe hypertension (SBP > 160mmHg) with reduced pairwise agreement for SBP (aOR 0.17, 95% CI 0.11 to 0.27). There was no association between patient transport and agreement between the methods for SBP, DBP, or MAP.

CONCLUSIONS: Non-invasive blood pressure measurements are often inaccurate in the pre-hospital critical care setting, particularly in patients with haemodynamic instability. Clinicians should be cautious when interpreting NIBP measurements and consider direct arterial pressure monitoring when circumstances allow.