操作者独立连续超声监测膈肌位移预测机械通气脱机成功:一项前瞻性观察性研究
该前瞻性观察性研究纳入49例机械通气患者,在30分钟自主呼吸试验期间用操作者独立的连续超声设备(CUSdi)测量膈肌位移(EXdi)与收缩峰值速度(PCVdi)。CUSdi与标准超声测量EXdi的一致性偏倚为0.1 cm(95%CI -0.7–0.9 cm),两者正相关(Rho=0.84,p<0.001)。54%患者脱机失败;自主呼吸试验开始后1、2、3分钟,脱机成功组EXdi均高于失败组;2分钟时EXdi<1.1 cm预测脱机失败敏感度0.83、特异度0.68、阳性预测值0.76、阴性预测值0.24。EXdi与膈肌压力峰值(r=0.22,95%CI 0.15–0.28)及压力时间乘积(r=0.13,95%CI 0.06–0.20)仅弱相关。
要点
- 49例机械通气患者30分钟自主呼吸试验中,CUSdi与标准超声测EXdi一致性偏倚0.1 cm(95%CI -0.7–0.9 cm),Rho=0.84(p<0.001)。
- 脱机失败率54%;自主呼吸试验开始后2分钟EXdi<1.1 cm预测脱机失败,敏感度0.83、特异度0.68、阳性预测值0.76、阴性预测值0.24。
- 自主呼吸试验开始后1、2、3分钟,脱机成功组EXdi均高于脱机失败组。
- EXdi与膈肌压力峰值(r=0.22,95%CI 0.15–0.28)及压力时间乘积(r=0.13,95%CI 0.06–0.20)仅弱相关,PCVdi结果类似。
适用人群与场景
- 49例机械通气患者,在30分钟自主呼吸试验期间接受CUSdi监测。
诊断标准与关键阈值
- 自主呼吸试验开始后2分钟,EXdi<1.1 cm作为预测脱机失败的切点值。
研究设计与关键结果
- 研究类型为前瞻性观察性研究,纳入49例机械通气患者,在30分钟自主呼吸试验中记录CUSdi并测量EXdi与PCVdi。
- CUSdi与标准超声测量EXdi的一致性偏倚为0.1 cm(95%CI -0.7–0.9 cm);Passing-Bablok回归显示两者无系统差异,呈正相关(Rho=0.84,p<0.001)。
- 54%患者脱机失败;自主呼吸试验开始后1、2、3分钟,脱机成功组EXdi高于失败组。
- 自主呼吸试验开始后2分钟EXdi<1.1 cm预测脱机失败:敏感度0.83、特异度0.68、阳性预测值0.76、阴性预测值0.24。
- EXdi与膈肌压力峰值(r=0.22,95%CI 0.15–0.28)及压力时间乘积(r=0.13,95%CI 0.06–0.20)呈弱相关;PCVdi结果类似。
证据类型、等级与局限
- 证据类型:前瞻性观察性研究。
- 局限:原文未报告局限性。
对床旁实践的改变
- 作者认为,操作者独立的连续膈肌监测可可靠量化EXdi,并以1.1 cm为切点值预测脱机失败。
提炼说明
- 本条依据论文摘要整理;摘要未涉及「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开
BACKGROUND: In mechanically ventilated patients, diaphragm ultrasound can identify diaphragm weakness and predict weaning failure. We evaluated whether a novel operator-independent ultrasound-based medical device allowing continuous monitoring of the diaphragm (CUSdi) could reliably (1) measure diaphragm excursion (EXdi) and peak contraction velocity (PCVdi), (2) predict weaning outcome, and (3) approximate transdiaphragmatic pressure (Pdi).
METHODS: In 49 mechanically ventilated patients, CUSdi was recorded during a 30-min spontaneous breathing trial (SBT), and EXdi and PCVdi were measured. In subgroups of patients, standard ultrasound measurement of EXdi and PCVdi was performed (n = 36), and Pdi derived parameters (peak and pressure time product, n = 30) were measured simultaneously.
RESULTS: The agreement bias between standard ultrasound and CUSdi for EXdi was 0.1 cm (95% confidence interval -0.7-0.9 cm). The regression of Passing-Bablok indicated a lack of systematic difference between EXdi measured with standard ultrasound and CUSdi, which were positively correlated (Rho = 0.84, p < 0.001). Weaning failure was observed in 54% of patients. One, two and three minutes after the onset of the SBT, EXdi was higher in the weaning success group than in the failure group. Two minutes after the onset of the SBT, an EXdi < 1.1 cm predicted weaning failure with a sensitivity of 0.83, a specificity of 0.68, a positive predictive value of 0.76, and a negative predictive value of 0.24. There was a weak correlation between EXdi and both peak Pdi (r = 0.22, 95% confidence interval 0.15 - 0.28) and pressure time product (r = 0.13, 95% confidence interval 0.06 - 0.20). Similar results were observed with PCVdi.
CONCLUSIONS: Operator-independent continuous diaphragm monitoring quantifies EXdi reliably and can predict weaning failure with an identified cut-off value of 1.1 cm. Trial registration clinicaltrial.gov, NCT04008875 (submitted 12 April 2019, posted 5 July 2019) and NCT03896048 (submitted 27 March 2019, posted 29 March 2019).