不同形状气囊气管导管在最低安全压力下封闭效果的体外研究

[In vitro study on the sealing effect of different shapes of cuff tracheal tubes under the lowest safe pressure].
PubMed 文献 · Zhonghua Wei Zhong Bing Ji Jiu Yi Xue · 2024-01-01
一句话结论持续测压装置可使锥形气囊在最低安全压力下维持有效封闭;间歇测压和/或圆柱形气囊时目标压力应设为25-30 cmH2O并定期调整。

该研究为体外实验,纳入2021年12月至2022年1月广西医科大学附属肿瘤医院ICU 24例气管插管患者,取其气囊上分泌物制作气管模型,比较圆柱形与锥形气囊导管在最低安全压力(20 cmH2O)下间歇测压与持续测压的封闭效果。共取180份滞留样本,两种测压方法各90份,每种黏度各30份。结果显示充气8小时时持续测压组不同形状气囊上声门下分泌物漏出量均较间歇测压组显著减少,III度黏稠分泌物在圆柱形气囊漏出量1.00(0.00, 1.25)mL vs. 2.00(1.00, 2.00)mL、锥形气囊1.00(0.00, 1.00)mL vs. 2.00(2.00, 2.00)mL,均P<0.01;持续测压组各时间点气囊压力均在设定范围20-21 cmH2O内,间歇测压组充气4小时气囊压力显著低于初始值(圆柱形18.3±0.6 cmH2O vs. 20.0±0.0 cmH2O,锥形18.4±0.6 cmH2O vs. 20.0±0.0 cmH2O,均P<0.01),且随充气时间延长呈显著下降趋势,但不同形状气囊导管间测压值差异无统计学意义。

要点

  • 体外实验中,充气8小时时持续测压组III度黏稠分泌物漏出量显著少于间歇测压组:圆柱形气囊1.00(0.00, 1.25)mL vs. 2.00(1.00, 2.00)mL,锥形气囊1.00(0.00, 1.00)mL vs. 2.00(2.00, 2.00)mL,均P<0.01。
  • 持续测压组不同形状气囊在各充气时间点测压值均在设定范围20-21 cmH2O内;间歇测压组充气4小时气囊压力显著低于初始值(圆柱形18.3±0.6 vs. 20.0±0.0 cmH2O,锥形18.4±0.6 vs. 20.0±0.0 cmH2O,均P<0.01)。
  • 间歇测压组充气4小时,圆柱形气囊I、II度分泌物样本全部漏出、III度3份漏出;锥形气囊I度28份漏出、II度仅2份漏出、III度无漏出。
  • 持续测压组充气4小时,圆柱形气囊I度全部漏出、II度29份漏出、III度无漏出;锥形气囊I度26份漏出、II度和III度无漏出;充气6小时锥形气囊III度仍无漏出。
  • 作者认为,使用改进测压法进行间歇测压和/或使用圆柱形气囊导管时,目标压力应设为25-30 cmH2O,并定期调整气囊压力。

适用人群与场景

  • 2021年12月至2022年1月广西医科大学附属肿瘤医院ICU收治的24例气管插管患者,取其气囊上分泌物进行体外实验。

诊断标准与关键阈值

  • 气囊上分泌物黏度分级:I度为水样声门下分泌物,II度为黏稠声门下分泌物,III度为凝胶样声门下分泌物。
  • 实验前气囊压力校准至20 cmH2O(1 cmH2O≈0.098 kPa)。

处置流程与药物剂量

  • 使用持续测压装置时,锥形气囊导管可在最低安全压力下维持有效封闭。
  • 使用改进测压法进行间歇测压和/或使用圆柱形气囊导管时,作者认为目标压力应设为25-30 cmH2O,并定期调整气囊压力。

研究设计与关键结果

  • 研究为体外实验,纳入24例气管插管患者,共取180份滞留样本,两种测压方法各90份,每种黏度各30份;采用自身对照,先对圆柱形和锥形气囊行间歇测压,再行持续测压,记录三种黏度声门下分泌物漏出量及充气4、6、8小时气囊测压值。
  • 充气8小时时,持续测压组声门下分泌物漏出量较间歇测压组显著减少:III度黏稠分泌物在圆柱形气囊为1.00(0.00, 1.25)mL vs. 2.00(1.00, 2.00)mL,在锥形气囊为1.00(0.00, 1.00)mL vs. 2.00(2.00, 2.00)mL,均P<0.01。
  • 间歇测压组充气4小时,圆柱形气囊I度、II度分泌物样本全部漏出,III度3份漏出;锥形气囊I度28份漏出,II度仅2份漏出,III度无漏出。充气6小时,不同形状气囊三种黏度分泌物样本全部漏出;漏出量随充气时间延长逐渐增加。
  • 持续测压组充气4小时,圆柱形气囊I度全部漏出、II度29份漏出、III度无漏出;锥形气囊I度26份漏出、II度和III度无漏出。充气6小时,锥形气囊III度仍无漏出。充气8小时,所有样本均漏出。
  • 持续测压组不同形状气囊在各充气时间点测压值均在设定范围20-21 cmH2O内;间歇测压组充气4小时气囊压力显著低于初始值(圆柱形18.3±0.6 cmH2O vs. 20.0±0.0 cmH2O,锥形18.4±0.6 cmH2O vs. 20.0±0.0 cmH2O,均P<0.01),且随充气时间延长呈显著下降趋势;不同形状气囊导管间测压值差异无统计学意义。

证据类型、等级与局限

  • 证据类型:体外研究。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 作者认为,使用持续测压装置时可维持锥形气囊导管在最低安全压力下的有效封闭;使用改进测压法进行间歇测压和/或使用圆柱形气囊导管时,目标压力应设为25-30 cmH2O,并定期调整气囊压力。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

OBJECTIVE: To compare the effectiveness of cylindrical-shaped and conical-shaped cuff catheters for airway closure using different pressure measurement methods at the lowest safe pressure and to guide the clinical application.

METHODS: Twenty-four patients with endotracheal intubation admitted to the intensive care unit (ICU) of Guangxi Medical University Cancer Hospital from December 2021 to January 2022 were enrolled. Leakage test in vitro was performed on the secretion on the patients' cuff. The needle and plunger from 20 mL syringe was separated, the syringe was sealed with adhesive, and the syringe nozzle was filled thoroughly to create a tracheal model. Consecutively, both cylindrical-shaped and conical-shaped cuff catheters were inserted into the simulated trachea, and the cuff pressure was calibrated to 20 cmH2O (1 cmH2O ≈ 0.098 kPa) before commencing the experiment. The viscosity of the secretion on the patients' cuff was classified (grade I was watery subglottic secretion, grade II was thick subglottic secretion, grade III was gel-like subglottic secretion), and the same viscosity secretion was injected into the catheter cuff. Utilizing a self-control approach, intermittent pressure measurement was initially conducted on both the cylindrical-shaped and conical-shaped cuff by improved pressure measurement method (intermittent pressure measurement group), followed by continuous pressure measurement experiment (continuous pressure measurement group). The leakage volume of the three viscosity subglottic secretions and the values of cuff pressure measurement of different shaped cuff catheters at 4, 6, 8 hours of inflation were recorded.

RESULTS: A total of 180 retention samples were extracted from 24 patients with tracheal intubation during ventilation, with 90 samples in each of the two groups using different pressure measurement methods, and 30 samples of retention materials with different viscosities in each group. In the intermittent pressure measurement group, at 4 hours of inflation, all samples of secretion with grade I and grade II on cylindrical-shaped cuff leaked, while 3 samples of secretion with grade III also leaked. For conical-shaped cuff, 28 samples of secretion with grade I leaked, only 2 samples of secretion with grade II leaked, and there was no leak for secretion with grade III. At 6 hours of inflation, all samples of the three viscosity secretions on different shaped cuffs leaked. The leakage was gradually increased with the prolongation of inflation time. In the continuous pressure measurement group, at 4 hours of inflation, all samples of secretion with grade I on cylindrical-shaped cuff leaked, while 29 samples of secretion with grade II leaked, and there was no leak for secretion with grade III. For the conical-shaped cuff, 26 samples of secretion with grade I leaked, and there was no leak for secretion with grade II and grade III. At 6 hours of inflation, the conical-shaped cuff still had no leak for secretion with grade III. As the inflation time prolonged, the leakage of subglottic secretion on different shaped cuffs in both groups was gradually increased. At 8 hours of inflation, all samples experienced leakage, but the leakage of subglottic secretion on different shaped cuffs in the continuous pressure measurement group was significantly reduced as compared with the intermittent pressure measurement group [leakage for secretion with grade III (mL): 1.00 (0.00, 1.25) vs. 2.00 (1.00, 2.00) on the cylindrical-shaped cuff, 1.00 (0.00, 1.00) vs. 2.00 (2.00, 2.00) on the conical-shaped cuff, both P < 0.01]. The values of pressure measurement of cuffs with different shapes at different time points of inflation in the continuous pressure measurement group were within the set range (20-21 cmH2O). The cuff pressure at 4 hours of inflation in the intermittent pressure measurement group was significantly lower than the initial value (cmH2O: 18.3±0.6 vs. 20.0±0.0 in the cylindrical-shaped cuff, 18.4±0.6 vs. 20.0±0.0 in the conical-shaped cuff, both P < 0.01), and the cuff pressure in both shaped cuffs showed a significant decrease tendency as inflation time prolonged. However, there was no statistically significant difference in values of pressure measurement between the different shaped cuff catheters.

CONCLUSIONS: Continuous pressure monitoring devices can maintain the effective sealing of conical-shaped cuff catheters at the lowest safe pressure. When using an improved pressure measurement method for intermittent pressure measurement and/or using a cylindrical cuff catheter, the target pressure should be set at 25-30 cmH2O, and the cuff pressure should be adjusted regularly.

来源:Zhonghua Wei Zhong Bing Ji Jiu Yi Xue,查看原文。版权归原作者所有。