文献速递心脑血管急症

Apple Watch 心电图输出识别心律失常的诊断准确性:与急诊科12导联心电图的比较

Diagnostic accuracy of apple watch ECG outputs in identifying dysrhythmias: A comparison with 12-Lead ECG in emergency department.
PubMed 文献 · Am J Emerg Med · 2024-02-04
一句话结论在急诊重症区721例患者中,Apple Watch 区分正常与异常心律的能力与标准心电图导联I相似(p=0.52),但识别异常心电图表现存在显著差异(p<0.05)。

这项单中心前瞻性观察研究于2023年10月1日至31日在三级急诊科重症区连续纳入721例因各种临床原因接受标准12导联心电图的患者,将Apple Watch佩戴于腕部获取单导联心电图,与标准心电图导联I比较心率、心律及异常表现,由两名急诊医学专家判读。Apple Watch区分正常与异常心律的能力与标准心电图相似(p=0.52),但识别异常心电图表现存在显著差异(p<0.05)。Bland-Altman分析显示心率绝对平均差为0.81±6.12次/分(r=0.94),700次心率测量中658次(94%)一致性强。作者认为Apple Watch有潜力检测房颤以外的心律,可能有助于患者到达急诊科前评估心律,但需更大队列进一步研究,尤其针对特定诊断。

要点

  • 在急诊重症区721例患者中,Apple Watch区分正常与异常心律的能力与标准心电图导联I相似(p=0.52)。
  • Apple Watch与标准心电图在识别导联I异常心电图表现上存在显著差异(p<0.05)。
  • 心率评估的绝对平均差为0.81±6.12次/分(r=0.94),700次测量中658次(94%)一致性强。
  • 作者认为Apple Watch有潜力检测房颤以外的心律,可能有助于患者到达急诊科前评估心律。
  • 作者指出需更大患者队列进一步研究,尤其针对特定诊断。

适用人群与场景

  • 2023年10月1日至31日在三级急诊科重症区接受评估、因各种临床原因行标准12导联心电图的所有患者,连续纳入。

研究设计与关键结果

  • 单中心前瞻性观察研究,纳入721例急诊重症区患者,Apple Watch佩戴于腕部获取单导联心电图,与标准心电图导联I比较,由两名急诊医学专家判读。
  • Apple Watch区分正常与异常心律的能力与标准心电图相似(p=0.52)。
  • Apple Watch与标准心电图在识别导联I异常心电图表现上存在显著差异(p<0.05)。
  • Bland-Altman分析显示心率绝对平均差为0.81±6.12次/分(r=0.94)。
  • 700次心率测量中658次(94%)一致性强。

证据类型、等级与局限

  • 单中心前瞻性观察研究。
  • 局限:作者指出需更大患者队列进一步研究,尤其针对特定诊断。

对床旁实践的改变

  • 作者认为Apple Watch有潜力检测房颤以外的心律,Apple Watch获取的心电描记可能有助于患者到达急诊科前评估心律。
  • 作者指出需更大患者队列进一步研究,尤其针对特定诊断。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

BACKGROUND: Wearable devices, particularly smartwatches like the Apple Watch (AW), can record important cardiac information, such as single‑lead electrocardiograms (ECGs). Although they are increasingly used to detect conditions such as atrial fibrillation (AF), research on their effectiveness in detecting a wider range of dysrhythmias and abnormal ECG findings remains limited. The primary objective of this study is to evaluate the accuracy of the AW in detecting various cardiac rhythms by comparing it with standard ECG's lead-I.

METHODS: This single-center prospective observational study was conducted in a tertiary care emergency department (ED) between 1.10.2023 and 31.10.2023. The study population consisted of all patients assessed in the critical care areas of the ED, all of whom underwent standard 12‑lead ECGs for various clinical reasons. Participants in the study were included consecutively. An AW was attached to patients' wrists and an ECG lead-I printout was obtained. Heart rate, rhythm and abnormal findings were evaluated and compared with the lead-I of standard ECG. Two emergency medicine specialists performed the ECG evaluations. Rhythms were categorized as normal sinus rhythm and abnormal rhythms, while ECG findings were categorized as the presence or absence of abnormal findings. AW and 12‑lead ECG outputs were compared using the McNemar test. Predictive performance analyses were also performed for subgroups. Bland-Altman analysis using absolute mean differences and concordance correlation coefficients was used to assess the level of heart rate agreement between devices.

RESULTS: The study was carried out on 721 patients. When analyzing ECG rhythms and abnormal findings in lead-I, the effectiveness of AW in distinguishing between normal and abnormal rhythms was similar to standard ECGs (p = 0.52). However, there was a significant difference between AW and standard ECGs in identifying abnormal findings in lead-I (p < 0.05). Using Bland-Altman analysis for heart rate assessment, the absolute mean difference for heart rate was 0.81 ± 6.12 bpm (r = 0.94). There was strong agreement in 658 out of 700 (94%) heart rate measurements.

CONCLUSION: Our study indicates that the AW has the potential to detect cardiac rhythms beyond AF. ECG tracings obtained from the AW may help evaluate cardiac rhythms prior to the patient's arrival in the ED. However, further research with a larger patient cohort is essential, especially for specific diagnoses.

来源:Am J Emerg Med,查看原文。版权归原作者所有。