文献速递复苏、休克与脓毒症心脑血管急症

猪心脏骤停模型中连续多普勒监测颈动脉血流评价心肺复苏的新方法

A new method to evaluate carotid blood flow by continuous Doppler monitoring during cardiopulmonary resuscitation in a porcine model of cardiac arrest.
PubMed 文献 · Resuscitation · 2023-12-15
一句话结论猪心脏骤停模型中,可穿戴颈动脉多普勒贴片可连续监测颈动脉血流,按压深度与颈动脉最大流速和速度时间积分显著相关,并可用于识别自主循环恢复。

研究使用可穿戴颈动脉多普勒贴片在20只Landrace猪心脏骤停模型中连续监测颈动脉血流,捕捉不同心律失常及心肺复苏期间的颈动脉频谱波形。结果显示颈动脉血流波形随室性心律失常类型变化;心肺复苏时按压深度与颈动脉最大流速(Spearman r=0.682,P<0.001)和速度时间积分(r=0.794,P<0.001)显著相关。颈动脉最大流速和速度时间积分对生存有中等预测价值;自主循环恢复时可见朝向脑部的规则颈动脉血流模式,与按压波形同时出现;自主循环恢复后,速度时间积分和颈动脉搏动量与每搏量趋势相似,颈动脉分钟输出量与心输出量趋势相似。作者认为颈动脉血流监测可提供心律失常及心肺复苏质量的信息,及时有效识别自主循环恢复,并可能提供自主循环恢复后的血流动力学信息。

要点

  • 在猪心脏骤停模型中,心肺复苏时按压深度与颈动脉最大流速显著相关(Spearman r=0.682,P<0.001),与速度时间积分显著相关(r=0.794,P<0.001)。
  • 可穿戴颈动脉多普勒贴片可连续监测颈动脉血流,颈动脉血流波形随室性心律失常类型变化。
  • 自主循环恢复时可见朝向脑部的规则颈动脉血流模式,与按压波形同时出现,可用于识别自主循环恢复。
  • 颈动脉最大流速和速度时间积分对生存有中等预测价值。
  • 自主循环恢复后,速度时间积分和颈动脉搏动量与每搏量趋势相似,颈动脉分钟输出量与心输出量趋势相似。

适用人群与场景

  • 20只Landrace猪心脏骤停模型,按标准方案诱导心脏骤停。

研究设计与关键结果

  • 研究类型:猪心脏骤停模型的实验研究;样本量:20只Landrace猪;干预:使用可穿戴颈动脉多普勒贴片连续无创监测颈动脉血流,捕捉不同心律失常及心肺复苏期间的颈动脉频谱波形。
  • 颈动脉血流波形随室性心律失常类型变化。
  • 心肺复苏期间,按压深度与颈动脉最大流速显著相关(Spearman相关系数r=0.682,P<0.001),与速度时间积分显著相关(r=0.794,P<0.001)。
  • 颈动脉最大流速和速度时间积分对生存有中等预测价值。
  • 自主循环恢复时可见朝向脑部的规则颈动脉血流模式,与按压波形同时出现。
  • 自主循环恢复后,速度时间积分和颈动脉搏动量与每搏量趋势相似,颈动脉分钟输出量与心输出量趋势相似。

证据类型、等级与局限

  • 证据类型:动物实验(猪心脏骤停模型)。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 在猪心脏骤停模型中,可穿戴颈动脉多普勒贴片可连续监测颈动脉血流,提供不同心律失常及心肺复苏质量的信息。
  • 在猪心脏骤停模型中,颈动脉血流监测可及时有效识别自主循环恢复,并可能提供自主循环恢复后的血流动力学信息(作者认为)。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

AIM: We used a wearable carotid Doppler patch to study carotid blood flow patterns in a porcine model of cardiac arrest to identify return of spontaneous circulation (ROSC) and hemodynamics associated with different arrhythmias and the quality of compressions.

METHODS: Twenty Landrace pigs were used as models of cardiac arrest following a standard protocol. Carotid blood flow was monitored continuously using noninvasive ultrasound. Carotid spectral waveforms were captured during various arrhythmias and CPR. Typical carotid blood flow waveforms were recorded at the time of ROSC, and hemodynamic changes were compared with carotid blood flow parameters.

RESULTS: The results showed that the carotid blood flow waveforms varied with ventricular arrhythmia type. During CPR, compression depth correlated significantly with carotid maximal velocity (Vmax) (Spearman correlation coefficient (r) = 0.682, P < 0.001) and velocity-time integral (VTI) (r = 0.794, P < 0.001). Vmax and VTI demonstrated moderate predictive value for survival. The regular carotid blood flow pattern towards the brain was observed during ROSC, concurrent with compression waveforms. After ROSC, VTI and carotid pulse volume (cPV) showed similar trends as stroke volume (SV). The carotid minute volume (cMV) exhibited a similar trend as cardiac output (CO).

CONCLUSIONS: Carotid blood flow monitoring could provide valuable information about different arrhythmias as well as the quality of CPR. Carotid flow monitoring allows for timely and effective identification of ROSC. In addition, it may provide valuable hemodynamic information after ROSC.

来源:Resuscitation,查看原文。版权归原作者所有。

参考来源

  1. Zhao X, Wang S, Yuan W, et al. A new method to evaluate carotid blood flow by continuous Doppler monitoring during cardiopulmonary resuscitation in a porcine model of cardiac arrest. Resuscitation. 2023. PMID: 38104797. doi:10.1016/j.resuscitation.2023.110092

供执业医师抢救与诊疗参考,不替代临床判断;剂量请结合体重、肝肾功能和本院药品规格复核,以指南原文为准。