ARDS患者中通过计算机断层扫描和电阻抗断层成像估算的动态相对区域肺应变
本研究纳入14例ARDS患者,在两个PEEP水平(基于ARDS-net策略的低PEEP和根据电阻抗断层成像滴定的高PEEP)下进行计算机断层扫描和电阻抗断层成像采集,比较三种基于电阻抗断层成像的动态相对区域肺应变方法与基于计算机断层扫描的相对应变。结果显示,基于(ΔZml低PEEP/EELIml低PEEP)/(ΔZml高PEEP/EELIml高PEEP)和ΔZAU低PEEP/ΔZAU高PEEP的方法与CT方法关系良好且一致性较好,而非依赖区R²分别为0.9050和0.8679,依赖区R²分别为0.8373和0.6588;基于EELIAU的方法关系弱且一致性差。结论提示电阻抗断层成像可用于监测ARDS患者PEEP试验中动态相对区域肺应变的变化,ΔZAU相对变化无需CT数据校准。
要点
- 在14例ARDS患者中,基于(ΔZml低PEEP/EELIml低PEEP)/(ΔZml高PEEP/EELIml高PEEP)的动态相对区域肺应变与CT方法关系良好:非依赖区R²=0.9050,依赖区R²=0.8373。
- 基于ΔZAU低PEEP/ΔZAU高PEEP的方法与CT方法关系良好:非依赖区R²=0.8679,依赖区R²=0.6588,且无需CT数据校准。
- 基于EELIAU的方法与CT方法呈弱负相关且一致性差:非依赖区和依赖区R²约0.3,偏倚分别为3.11和2.08。
- 两种表现良好的电阻抗断层成像方法偏倚范围为-0.11至0.51,一致性界限范围为-0.73至1.16(涵盖两种方法和两个肺区域)。
- 作者认为,ARDS患者PEEP试验中动态相对区域肺应变的变化可通过基于ΔZmL/EELIml和ΔZAU的电阻抗断层成像监测。
适用人群与场景
- 14例ARDS患者,在低PEEP(基于ARDS-net策略)和高PEEP(根据电阻抗断层成像滴定)两个水平下接受计算机断层扫描和电阻抗断层成像采集。
研究设计与关键结果
- 研究类型为在14例ARDS患者中比较电阻抗断层成像方法与计算机断层扫描评估动态相对区域肺应变的研究;在低PEEP(基于ARDS-net策略)和高PEEP(根据电阻抗断层成像滴定)两个水平下进行计算机断层扫描和电阻抗断层成像采集(吸气末和呼气末)。
- 基于(ΔZml低PEEP/EELIml低PEEP)/(ΔZml高PEEP/EELIml高PEEP)的动态相对区域肺应变与CT方法关系良好:非依赖区R²=0.9050,依赖区R²=0.8373。
- 基于ΔZAU低PEEP/ΔZAU高PEEP的动态相对区域肺应变与CT方法关系良好:非依赖区R²=0.8679,依赖区R²=0.6588。
- 上述两种方法与CT方法的偏倚范围为-0.11至0.51,一致性界限范围为-0.73至1.16(涵盖两种方法和两个肺区域)。
- 基于EELIAU的方法([ΔZAU低PEEP/EELIAU低PEEP]/[ΔZAU高PEEP/EELIAU高PEEP])与CT方法在非依赖区和依赖区均呈弱负相关且一致性差:R²约0.3;偏倚分别为3.11和2.08,一致性界限分别为-2.13至8.34和-1.49至5.64。
证据类型、等级与局限
- 证据类型:在14例ARDS患者中比较电阻抗断层成像与计算机断层扫描评估动态相对区域肺应变的研究。
- 局限:原文未报告局限性。
对床旁实践的改变
- 作者认为,ARDS患者PEEP试验中动态相对区域肺应变的变化可通过基于ΔZmL/EELIml和ΔZAU的电阻抗断层成像监测。
- 作者指出,ΔZAU的相对变化具有无需CT数据校准的优势。
提炼说明
- 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开
BACKGROUND: In the acute distress respiratory syndrome (ARDS), specific lung regions can be exposed to excessive strain due to heterogeneous disease, gravity-dependent lung collapse and injurious mechanical ventilation. Computed tomography (CT) is the gold standard for regional strain assessment. An alternative tool could be the electrical impedance tomography (EIT). We aimed to determine whether EIT-based methods can predict the dynamic relative regional strain (DRRS) between two levels of end-expiratory pressure (PEEP) in gravity-non-dependent and dependent lung regions.
METHODS: Fourteen ARDS patients underwent CT and EIT acquisitions (at end-inspiratory and end-expiratory) at two levels of PEEP: a low-PEEP based on ARDS-net strategy and a high-PEEP titrated according to EIT. Three EIT-based methods for DRRS were compared to relative CT-based strain: (1) the change of the ratio between EIT ventilation and end-expiratory lung impedance in arbitrary units ([ΔZAU low-PEEP/EELIAU low-PEEP]/[ΔZAU high-PEEP/EELIAU high-PEEP]), (2) the change of ΔZ/EELI ratio calibrated to mL ([ΔZml low-PEEP/EELIml low-PEEP]/[ΔZml high-PEEP/EELIml high-PEEP]) using CT data, and (3) the relative change of ∆ZAU (∆ZAU low-PEEP/∆ZAU high-PEEP). We performed linear regressions analysis and calculated bias and limits of agreement to assess the performance of DRRS by EIT in comparison with CT.
RESULTS: The DRRS assessed by (ΔZml low-PEEP/EELIml low-PEEP)/(ΔZml high-PEEP/EELIml high-PEEP) and ∆ZAU low-PEEP/∆ZAU high-PEEP showed good relationship and agreement with the CT method (R2 of 0.9050 and 0.8679, respectively, in non-dependent region; R2 of 0.8373 and 0.6588, respectively, in dependent region; biases ranging from - 0.11 to 0.51 and limits of agreement ranging from - 0.73 to 1.16 for both methods and lung regions). Conversely, DRRS based on EELIAU ([ΔZAU low-PEEP/EELIAU low-PEEP]/[ΔZAU high-PEEP/EELIAU high-PEEP]) exhibited a weak negative relationship and poor agreement with the CT method for both non-dependent and dependent regions (R2 ~ 0.3; bias of 3.11 and 2.08, and limits of agreement of - 2.13 to 8.34 and from - 1.49 to 5.64, respectively).
CONCLUSION: Changes in DRRS during a PEEP trial in ARDS patients could be monitored using EIT, based on changes in ΔZmL/EELIml and ∆ZAU. The relative change ∆ZAU offers the advantage of not requiring CT data for calibration.