文献速递呼吸衰竭与机械通气

不同无创氧疗支持装置对潮气量的影响

Influence of different noninvasive oxygenation support devices on tidal volume.
PubMed 文献 · Ann Intensive Care · 2023-11-25
一句话结论无创氧疗装置显著影响潮气量:NIV潮气量最高,CPAP最低,且与吸气期口腔压力变化强相关。

该研究采用人体模型连接肺模拟器的台架实验,评估六种条件(无装置、储氧面罩、经鼻高流量氧疗、持续气道正压、面罩无创通气、头盔无创通气)在三种呼吸力学(正常、阻塞性、限制性)和三种模拟用力(低、中、呼吸窘迫)下的潮气量,并在健康志愿者中用电阻抗断层成像评估相同装置。结果:台架模型中,与储氧面罩(440±352 mL)相比,经鼻高流量氧疗(414±333 mL)和持续气道正压(377±297 mL)潮气量显著更低,而面罩无创通气(690±321 mL)和头盔无创通气(652±366 mL)显著更高;潮气量与各装置吸气期口腔压力时间乘积强相关。健康志愿者中,储氧面罩与持续气道正压潮气量相似(644 [571-764] mL 和 648 [586-770] mL),经鼻高流量氧疗、面罩无创通气、头盔无创通气更高(819 [609-918] mL、1110 [661-1305] mL、1086 [833-1243] mL)。结论:无创氧疗装置显著影响潮气量,NIV潮气量最高、CPAP最低,需临床研究明确其意义。

要点

  • 台架模型中,与储氧面罩(440±352 mL)相比,经鼻高流量氧疗(414±333 mL)和持续气道正压(377±297 mL)潮气量显著更低,而面罩无创通气(690±321 mL)和头盔无创通气(652±366 mL)显著更高。
  • 健康志愿者中,储氧面罩与持续气道正压潮气量相似(644 [571-764] mL 和 648 [586-770] mL),经鼻高流量氧疗、面罩无创通气、头盔无创通气更高(819 [609-918] mL、1110 [661-1305] mL、1086 [833-1243] mL)。
  • 潮气量与各装置吸气期口腔压力时间乘积强相关:经鼻高流量氧疗和持续气道正压为负值(-0.3 [-0.8 至 -0.2] 和 -0.7 [-2.2 至 -0.5] cmH2O·秒/周期),头盔无创通气和面罩无创通气为正值(4.5 [4.1-4.6] 和 6.1 [5.9-7.1] cmH2O·秒/周期)。
  • 台架模型模拟了正常、阻塞性、限制性三种呼吸力学和低、中、呼吸窘迫三种用力,提示装置对潮气量的影响在不同条件下均存在。
  • 作者认为需要临床研究来阐明这些效应的临床意义。

适用人群与场景

  • 台架模型:人体模型连接肺模拟器,模拟正常、阻塞性、限制性三种呼吸力学,以及低、中、呼吸窘迫三种用力。
  • 健康志愿者:同一组装置在健康志愿者中评估,潮气量用电阻抗断层成像评估。

研究设计与关键结果

  • 研究类型:台架研究(人体模型连接肺模拟器)加健康志愿者评估;台架评估六种条件:无装置、储氧面罩、经鼻高流量氧疗、持续气道正压、面罩无创通气、头盔无创通气。
  • 台架模型潮气量:储氧面罩 440±352 mL,经鼻高流量氧疗 414±333 mL,持续气道正压 377±297 mL,面罩无创通气 690±321 mL,头盔无创通气 652±366 mL;与储氧面罩相比,经鼻高流量氧疗和持续气道正压显著更低,面罩无创通气和头盔无创通气显著更高。
  • 台架模型吸气期口腔压力时间乘积:经鼻高流量氧疗 -0.3 [-0.8 至 -0.2] cmH2O·秒/周期,持续气道正压 -0.7 [-2.2 至 -0.5] cmH2O·秒/周期,头盔无创通气 4.5 [4.1-4.6] cmH2O·秒/周期,面罩无创通气 6.1 [5.9-7.1] cmH2O·秒/周期;潮气量与该压力时间乘积强相关。
  • 健康志愿者潮气量:储氧面罩 644 [571-764] mL,持续气道正压 648 [586-770] mL,经鼻高流量氧疗 819 [609-918] mL,面罩无创通气 1110 [661-1305] mL,头盔无创通气 1086 [833-1243] mL;储氧面罩与持续气道正压相似,经鼻高流量氧疗、面罩无创通气、头盔无创通气更高。

证据类型、等级与局限

  • 证据类型:台架研究加健康志愿者生理学评估。
  • 局限:原文未报告局限性;作者在结论中提出需要临床研究来阐明这些效应的临床意义。

对床旁实践的改变

  • 原文未据此给出推荐;作者认为需要临床研究来阐明这些效应的临床意义。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

BACKGROUND: Multiple devices are available for noninvasive oxygenation support, including non-rebreather oxygen mask (O2-mask), high-flow oxygen through nasal cannula (HFNC), continuous positive airway pressure (CPAP), mask noninvasive ventilation (Mask-NIV) and helmet NIV (Helmet-NIV). As tidal volume is a key determinant of efficacy and safety during ventilatory support, we assessed whether it was influenced by the type of noninvasive oxygenation device.

METHODS: A bench study using a manikin with a realistic face connected to a lung simulator was performed. Six conditions were assessed: no device, O2-mask, HFNC, CPAP, Mask-NIV and Helmet-NIV. Three respiratory mechanics were simulated (normal, obstructive, restrictive), at three simulated efforts (low, moderate, respiratory distress). Flow was recorded at the lung simulator inlet and mouth pressure into the manikin mouth. The same devices were evaluated on healthy volunteers with tidal volume assessed by electrical impedance tomography (EIT).

RESULTS: Tidal volume was significantly influenced by oxygenation devices in bench model. As compared to O2-mask, HFNC and CPAP delivered significantly lower tidal volumes (440 ± 352 mL, 414 ± 333 mL and 377 ± 297 mL, respectively), while Mask-NIV or Helmet-NIV were associated with significantly higher tidal volumes (690 ± 321 mL and 652 ± 366 mL, respectively). Tidal volume was strongly correlated with the specific effect of each device on mouth pressure during inspiration: HFNC and CPAP were characterized by a negative PTPmouth (- 0.3 [- 0.8 to - 0.2] and - 0.7 [- 2.2 to - 0.5] cmH2O.sec/cycle, respectively), while Helmet-NIV and Mask-NIV were associated with a positive PTPmouth (4.5 [4.1-4.6] and 6.1 [5.9-7.1] cmH2O.sec/cycle, respectively). Tidal volume was also significantly influenced by oxygenation devices in healthy volunteers, with similar tidal volumes between O2-mask and CPAP (644 [571-764] and 648 [586-770] mL) but higher with HFNC, Mask-NIV and Helmet-NIV (819 [609-918], 1110 [661-1305] and 1086 [833-1243] mL).

CONCLUSIONS: Tidal volume is significantly influenced by noninvasive oxygenation support devices, with a strong correlation with the pressure variation generated into the mouth during inspiration. NIV was associated with the highest tidal volumes and CPAP with the lowest ones. Clinical studies are needed to clarify the clinical implications of these effects.

来源:Ann Intensive Care,查看原文。版权归原作者所有。