新生儿复苏中背式与胸式心电电极放置:一项先导性随机对照试验

Back vs. chest ECG electrode placement in neonatal resuscitation: A pilot randomized controlled trial.
PubMed 文献 · Resuscitation · 2023-09-09
一句话结论先导性RCT显示背式与胸式心电电极在新生儿复苏中获取可靠心率的时间无显著差异(21秒 vs 26秒,p=0.91),背式电极在需高级复苏的新生儿中结果令人鼓舞,但需进一步研究。

该先导性随机对照试验将174名新生儿随机分配至背式(n=85)或胸式(n=89)心电电极放置组,比较从置于Panda warmer ResusView开始至检测到可靠心率的时间。两组在1分钟内可检测到心率的比例相似,中位获取时间胸式组26(13,38)秒 vs 背式组21(12,54)秒(p=0.91),无显著差异。胸式组中,活跃新生儿的心率获取时间短于非活跃新生儿(34±48秒 vs 50±44秒,p=0.049)。在非活跃且需高级复苏的新生儿中,两组失败率和心率获取时间相似(p=0.51)。作者认为背式预设心电电极在需高级复苏的新生儿中显示出令人鼓舞的结果,但需进一步研究以完善新生儿复苏指导。

要点

  • 背式与胸式心电电极获取可靠心率的中位时间无显著差异:胸式26(13,38)秒 vs 背式21(12,54)秒(p=0.91)。
  • 两组在1分钟内可检测到心率的比例相似。
  • 胸式组中活跃新生儿心率获取时间短于非活跃新生儿(34±48秒 vs 50±44秒,p=0.049)。
  • 非活跃且需高级复苏的新生儿中,背式与胸式组的失败率和心率获取时间相似(p=0.51)。
  • 作者认为背式预设心电电极在需高级复苏的新生儿中结果令人鼓舞,但需进一步研究。

适用人群与场景

  • 分娩室中接受复苏的新生儿,随机分配至背式(n=85)或胸式(n=89)心电电极放置组。
  • 纳入大量活跃新生儿;非活跃且需高级复苏的新生儿亚组也进行了分析。

诊断标准与关键阈值

  • 失败定义为5分钟内未检测到心率。
  • 时间测量从新生儿置于Panda warmer ResusView开始。

研究设计与关键结果

  • 研究类型:先导性随机对照试验;样本量174例(背式n=85,胸式n=89);人群为分娩室新生儿;干预为背式心电电极放置,对照为胸式(标准前位)放置;主要终点为心率信号获取时间。
  • 两组1分钟内可检测到心率的比例相似;中位(IQR)获取时间胸式组26(13,38)秒 vs 背式组21(12,54)秒(p=0.91)。
  • 胸式组中活跃新生儿心率获取时间短于非活跃新生儿(均值±标准差34±48秒 vs 50±44秒;p=0.049)。
  • 非活跃且需高级复苏的新生儿中,背式与胸式组的失败率和心率获取时间相似(p=0.51)。

证据类型、等级与局限

  • 证据类型:先导性随机对照试验。
  • 局限:原文未报告局限性;作者结论指出需进一步研究以完善新生儿复苏指导。

对床旁实践的改变

  • 作者认为背式预设心电电极在需高级复苏的新生儿中显示出令人鼓舞的结果,但需进一步研究以增强新生儿复苏期间的指导。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

BACKGROUND: The recent Neonatal Resuscitation Program advises the early utilization of an electrocardiogram (ECG) for non-vigorous newborns in the delivery room. However, placing ECG electrodes on the chest may delay obtaining a reliable heart rate (HR) and could interfere with chest compressions. Our previous study showed that preset ECG electrodes, attached to the back of the newborn, are quicker than a pulse oximeter (POX) for detecting HR.

AIM: To compare time to detect a reliable HR using back-placed ECG electrodes versus standard front placement.

METHODS: Infants were randomly assigned to back (n = 85) or chest (n = 89) electrode placement. Time measurement began upon placing infants on a Panda warmer ResusView. Failure was defined as no HR detected within 5 minutes. Intention-to-treat analysis compared HR signal acquisition time between groups.

RESULTS: Both groups showed similar proportions of detectable HR within the first minute. Median (IQR) time to obtain HR was 26 (13,38) seconds for the chest group and 21 (12,54) seconds for the back group (p = 0.91). A large number of vigorous infants were included. In the chest group, these vigorous infants had shorter HR acquisition times than non-vigorous infants (Mean ± SD of 34 ± 48 seconds vs. 50 ± 44 seconds respectively; p = 0.049). Failure rates and time to acquire a HR for infants who were non-vigorous and required advanced resuscitation were similar between the back and chest groups (p = 0.51).

CONCLUSION: Preset back ECG electrodes have shown encouraging results in neonates requiring advanced resuscitation. Further studies are needed to enhance guidance during neonatal resuscitation.