文献速递呼吸衰竭与机械通气重症综合管理与镇静镇痛

超声引导下基于膈肌的电阻抗断层成像电极带定位新方法

A novel method for diaphragm-based electrode belt position of electrical impedance tomography by ultrasound.
PubMed 文献 · J Intensive Care · 2023-09-25
一句话结论超声定位膈肌后放置EIT电极带在健康志愿者中最高潮气阻抗变异位于L6cm,在>50%的危重患者中该位置与常规第4-6肋间不一致,高BMI患者更明显。

该研究介绍了一种基于超声确定膈肌位置的电阻抗断层成像(EIT)电极带放置方法,并在8名健康志愿者和50名机械通气危重患者中比较其与常规放置的差异。健康志愿者仰卧位下在7个不同水平采集EIT数据,最高潮气阻抗变异(TV)出现在L6cm,最低在L-2cm,故将L6cm作为超声确定的基于膈肌的电极带水平。50例患者中23例(Agreed组)基于膈肌的电极带平面与常规平面(第4-6肋间)一致,其余为Disagreed组(高于第4肋间);Disagreed组BMI显著更高,且基于膈肌的电极带平面的全局TV更低。作者认为该方法可行,在>50%的受试者中导致与常规位置不同的电极带位置,尤其在高BMI患者中,需进一步研究验证其对EIT图像和临床管理的影响。

要点

  • 超声确定膈肌位置(L0)为零参考,头侧为正、足侧为负;健康志愿者仰卧位7个水平中最高TV在L6cm、最低在L-2cm。
  • 8名健康志愿者中,以最高潮气阻抗变异(TV)确定的基于膈肌的EIT电极带水平为L6cm。
  • 50名机械通气危重患者中,23例基于膈肌的电极带平面与常规平面(第4-6肋间)一致(Agreed组),其余为Disagreed组(高于第4肋间)。
  • Disagreed组较Agreed组BMI显著更高,且基于膈肌的电极带平面的全局TV显著更低。
  • 作者认为超声引导基于膈肌的电极带定位可行,在>50%的受试者中导致与常规位置不同的电极带位置,尤其在高BMI患者中;需进一步研究验证其对EIT图像和临床管理的影响。

适用人群与场景

  • 健康志愿者8名(仰卧位,用于确定基于膈肌的EIT电极带水平)。
  • 机械通气危重患者50名(用于比较基于膈肌的电极带水平与传统电极带水平)。

诊断标准与关键阈值

  • 以超声确定的膈肌位置(L0)为零参考水平,头侧方向为正、足侧方向为负。
  • 基于膈肌的EIT电极带水平(Lxcm)定义为达到最高潮气阻抗变异(TV)的水平。
  • 常规EIT电极带平面为第4-6肋间;基于膈肌的电极带平面与常规平面一致者归为Agreed组,其余归为Disagreed组(高于第4肋间)。

处置流程与药物剂量

  • 健康志愿者仰卧位下分别在L-2 cm、L0、L2cm、L4cm、L6cm、L8cm、L10cm共7个水平采集EIT数据。
  • 在50名机械通气危重患者中,分别在基于膈肌的EIT电极带水平和传统EIT电极带水平进行EIT测量。

研究设计与关键结果

  • 研究类型为方法学比较研究;健康志愿者8名,机械通气危重患者50名;干预为超声确定膈肌位置后放置EIT电极带,对照为传统EIT电极带位置。
  • 健康志愿者中最高TV出现在L6cm,最低出现在L-2cm,故将L6cm作为超声确定的基于膈肌的EIT电极带水平。
  • 50例患者中23例基于膈肌的电极带平面与常规平面(第4-6肋间)一致(Agreed组),其余为Disagreed组(高于第4肋间)。
  • Disagreed组较Agreed组BMI显著更高,且基于膈肌的电极带平面的全局TV显著更低。
  • 基于膈肌的电极带位置在>50%的受试者中与常规位置不同,尤其在高BMI患者中。

证据类型、等级与局限

  • 证据类型:方法学比较研究(健康志愿者+危重患者)。
  • 局限:作者指出需进一步研究验证该新方法对EIT图像和临床管理的影响。

对床旁实践的改变

  • 作者认为超声引导下基于膈肌的EIT电极带定位可行,在>50%的受试者中导致与常规位置不同的电极带位置,尤其在高BMI患者中。
  • 作者指出需进一步研究验证该新方法对EIT图像和临床管理的影响。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

BACKGROUND: This aim of study was to introduce a diaphragm-based EIT-belt placement method based on diaphragm position by ultrasound, and to evaluate the difference between diaphragm-based EIT-belt placement and conventional EIT-belt placement.

METHOD: The diaphragm position (L0) determined by ultrasound was taken as zero reference level. The direction of headward is defined as positive, and toward feet is negative. For EIT data collection, the electrode belt was placed at 7 different levels, respectively (denoted as L-2 cm, L0, L2cm, L4cm, L6cm, L8cm, L10cm) at supine position in healthy volunteers. The diaphragm-based EIT-belt level (Lxcm) was defined where highest tidal impedance variation (TV) was achieved. Subsequently, EIT measurements were conducted at diaphragm-based EIT-belt levels and traditional EIT-belt level in 50 critically ill patients under mechanical ventilation.

RESULT: The highest TV was achieved at L6cm and the smallest at L-2 cm., so the L6cm were taken as diaphragm-based EIT-belt level by ultrasound in 8 healthy volunteers. In 23 patients, the diaphragm-based EIT-belt plane agreed with the conventional planes (4th-6th ICS), which was defined as the Agreed group. Other patients were classified to the Disagreed group (above 4th ICS). The Disagreed group has a significantly higher BMI and lower global TV at the diaphragm-based EIT-belt plane compared to the Agreed group.

CONCLUSIONS: The diaphragm-based EIT-belt position by ultrasound was feasible and resulted in different belt positions compared to the conventional position in > 50% of the examined subjects, especially in patients with higher BMI. Further study is required to validate the impact on EIT images with this novel method on clinical management.

来源:J Intensive Care,查看原文。版权归原作者所有。