颈动脉超声与人工触诊法在心肺复苏中脉搏检查的比较
该前瞻性观察研究在大学的医院急诊医学科危重症区开展,对49例非创伤性心搏骤停并接受心肺复苏的患者,同时用一侧颈动脉超声与另一侧人工触诊法进行脉搏检查,共568次测量。以监护心律、人工股动脉触诊、呼气末二氧化碳和心脏超声综合判断的自主循环恢复为金标准,人工法敏感度80%、特异度91%(+PV 35%,-PV 64%),颈动脉超声敏感度100%、特异度98%(+PV 84%,-PV 100%),两者敏感度差-0.0704(95%CI -0.0965~-0.0466)、特异度差0.0106(95%CI 0.0006~0.0222),差异有统计学意义。人工法判断自主循环恢复用时3±0.17 s,颈动脉超声2.8±0.15 s,差异有统计学意义。作者认为颈动脉超声脉搏检查在快速、准确决策方面可能优于人工法。
要点
- 颈动脉超声判断自主循环恢复的敏感度100%、特异度98%,+PV 84%、-PV 100%,人工触诊法分别为80%、91%、35%、64%。
- 两种方法敏感度差-0.0704(95%CI -0.0965~-0.0466),特异度差0.0106(95%CI 0.0006~0.0222),差异均有统计学意义。
- 判断自主循环恢复用时:人工法3±0.17 s,颈动脉超声2.8±0.15 s,差异有统计学意义。
- 研究共对49例非创伤性心搏骤停患者完成568次脉搏测量,两种方法分别取自两侧颈动脉。
- 作者认为颈动脉超声脉搏检查在快速、准确决策方面可能优于人工触诊法。
适用人群与场景
- 在大学的医院急诊医学科危重症区接受心肺复苏的非创伤性心搏骤停患者。
诊断标准与关键阈值
- 以团队领导者结合监护心律、人工股动脉触诊、呼气末二氧化碳和心脏超声作出的临床判断作为判断自主循环恢复的金标准。
研究设计与关键结果
- 研究类型为前瞻性观察研究,在大学的医院急诊医学科危重症区开展,对49例非创伤性心搏骤停患者共完成568次脉搏测量,同一患者一侧颈动脉用颈动脉超声、另一侧用人工触诊法检查。
- 人工触诊法预测自主循环恢复的敏感度80%、特异度91%(+PV 35%,-PV 64%);颈动脉超声敏感度100%、特异度98%(+PV 84%,-PV 100%)。
- 颈动脉超声与人工触诊法的敏感度差为-0.0704(95%CI -0.0965~-0.0466),特异度差为0.0106(95%CI 0.0006~0.0222),以团队领导者综合多种仪器的临床判断为金标准分析,差异有统计学意义。
- 人工触诊法作出自主循环恢复判断用时3±0.17 s,颈动脉超声2.8±0.15 s,差异有统计学意义。
证据类型、等级与局限
- 证据类型:前瞻性观察研究。
- 局限:原文未报告局限性。
对床旁实践的改变
- 作者认为,在心肺复苏中,颈动脉超声脉搏检查在快速、准确决策方面可能优于人工触诊法。
提炼说明
- 本条依据论文摘要整理;摘要未涉及「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开
OBJECTIVES: The success of the manual pulse check method frequently employed during cardiopulmonary resuscitation (CPR) is controversial due to its subjective, patient- and operator-dependent, and time-consuming nature. Carotid ultrasound (c-USG) has recently emerged as an alternative, although there are still insufficient studies on the subject. The purpose of the present study was to compare the success of the manual and c-USG pulse check methods during CPR.
METHODS: This prospective observational study was conducted in the critical care area of a university hospital emergency medicine clinic. Pulse checks in patients with non-traumatic cardiopulmonary arrest (CPA) undergoing CPR were performed using the c-USG method from one carotid artery and the manual method from the other. The gold standard in the decision regarding return of spontaneous circulation (ROSC) was the clinical judgment made using the rhythm on the monitor, manual femoral pulse check, end tidal carbon dioxide (ETCO2), and cardiac USG instruments. The success in predicting ROSC and measurement times of the manual and c-USG methods were compared. The success of both methods was calculated as sensitivity and specificity, and the clinical significance of the difference between the methods' sensitivity and specificity was evaluated Newcombe's method.
RESULTS: A total of 568 pulse measurements were performed on 49 CPA cases using both c-USG and the manual method. The manual method exhibited 80% sensitivity and 91% specificity in predicting ROSC (+PV: 35%, -PV: 64%), while c-USG exhibited 100% sensitivity and 98% specificity (+PV: 84%, -PV: 100%). The difference in sensitivities between the c-USG and manual methods was -0.0704 (95% CI: -0.0965; -0.0466), and the difference between their specificities was 0.0106 (95% CI: 0.0006; 0.0222). The difference between the specificities and sensitivities was statistically significant at analysis performed adopting the clinical judgment of the team leader using multiple instruments as the gold standard. The manual method yielded an ROSC decision in 3 ± 0.17 s and c-USG in 2.8 ± 0.15 s, the difference being statistically significant.
CONCLUSION: According to the results of this study, the pulse check method with c-USG may be superior to the manual method in terms of fast and accurate decision making in CPR.