人工颅内压记录与计算机高分辨率系统相比的准确性:一项CENTER-TBI分析
该研究纳入100例患者,比较入住后7天内人工记录(MR)的整点颅内压(ICP)和脑灌注压(CPP)与计算机高分辨率(HR)系统连续记录值的差异,并分析基于整点测量计算的压力-时间剂量能否反映高ICP(>20 mm Hg)和低CPP(<60 mm Hg)负荷。结果显示MR与HR整点ICP平均差0.02 mm Hg(SD 3.86),CPP平均差1.54 mm Hg(SD 8.81);MR漏检ICP>20 mm Hg和CPP<60 mm Hg的比例分别为1.6%和5.8%。但基于人工记录计算的负荷在42%和28%的患者中严重低估ICP和CPP负荷。提示整点人工记录不能可靠替代连续监测评估颅内高压和低CPP负担。
要点
- 人工整点ICP与HR系统整点ICP平均差0.02 mm Hg(SD 3.86),CPP平均差1.54 mm Hg(SD 8.81),整点值本身较接近。
- 同步测量中,人工记录漏检ICP>20 mm Hg的比例为1.6%,漏检CPP<60 mm Hg的比例为5.8%。
- 基于人工整点记录计算的压力-时间剂量在42%的患者中严重低估ICP负荷,在28%的患者中严重低估CPP负荷。
- 对重症TBI患者,仅凭整点人工记录评估颅内高压和低CPP负担可能不足,存在严重低估风险。
适用人群与场景
- 纳入100例患者,比较入住后前7天内人工记录与计算机高分辨率系统记录的ICP和CPP数据。
诊断标准与关键阈值
- 高ICP定义为ICP>20 mm Hg,低CPP定义为CPP<60 mm Hg。
研究设计与关键结果
- 研究类型为CENTER-TBI分析,纳入100例患者,比较人工记录(MR)整点ICP和CPP与计算机高分辨率(HR)系统连续记录值。
- MR与HR整点值平均差:ICP 0.02 mm Hg(SD 3.86 mm Hg),CPP 1.54 mm Hg(SD 8.81 mm Hg)。
- 同步测量中,MR未检出ICP>20 mm Hg的比例为1.6%,未检出CPP<60 mm Hg的比例为5.8%。
- 基于人工记录计算的压力-时间剂量在42%的患者中严重低估ICP负荷,在28%的患者中严重低估CPP负荷。
证据类型、等级与局限
- 证据类型:CENTER-TBI分析(观察性研究)。
- 局限:原文未报告局限性。
对床旁实践的改变
- 作者认为,人工整点记录能较好代表整点HR ICP和CPP,但与计算机系统相比可能不够充分,存在严重低估ICP和CPP负荷的风险。
提炼说明
- 本条依据论文摘要整理;摘要未涉及「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开
BACKGROUND: Monitoring intracranial pressure (ICP) and cerebral perfusion pressure (CPP) is crucial in the management of the patient with severe traumatic brain injury (TBI). In several institutions ICP and CPP are summarized hourly and entered manually on bedside charts; these data have been used in large observational and interventional trials. However, ICP and CPP may change rapidly and frequently, so data recorded in medical charts might underestimate actual ICP and CPP shifts. The aim of this study was to evaluate the accuracy of manual data annotation for proper capturing of ICP and CPP. For this aim, we (1) compared end-hour ICP and CPP values manually recorded (MR) with values recorded continuously by computerized high-resolution (HR) systems and (2) analyzed whether MR ICP and MR CPP are reliable indicators of the burden of intracranial hypertension and low CPP.
METHODS: One hundred patients were included. First, we compared the MR data with the values stored in the computerized system during the first 7 days after admission. For this point-to-point analysis, we calculated the difference between end-hour MR and HR ICP and CPP. Then we analyzed the burden of high ICP (> 20 mm Hg) and low CPP (< 60 mm Hg) measured by the computerized system, in which continuous data were stored, compared with the pressure-time dose based on end-hour measurements.
RESULTS: The mean difference between MR and HR end-hour values was 0.02 mm Hg for ICP (SD 3.86 mm Hg) and 1.54 mm Hg for CPP (SD 8.81 mm Hg). ICP > 20 mm Hg and CPP < 60 mm Hg were not detected by MR in 1.6% and 5.8% of synchronized measurements, respectively. Analysis of the pathological ICP and CPP throughout the recording, however, indicated that calculations based on manual recording seriously underestimated the ICP and CPP burden (in 42% and 28% of patients, respectively).
CONCLUSIONS: Manual entries fairly represent end-hour HR ICP and CPP. However, compared with a computerized system, they may prove inadequate, with a serious risk of underestimation of the ICP and CPP burden.