文献速递呼吸衰竭与机械通气重症综合管理与镇静镇痛

呼吸机显示屏同时显示估算的肌肉压力曲线对医护人员正确识别患者-呼吸机不同步能力的影响:一项随机研究(Pmus研究)

Impact on the ability of healthcare professionals to correctly identify patient-ventilator asynchronies of the simultaneous visualization of estimated muscle pressure curves on the ventilator display: a randomized study (Pmus study).
PubMed 文献 · Crit Care · 2023-03-30
一句话结论在模拟场景中,同时显示估算的Pmus波形使ICU医生和呼吸治疗师识别不同步的敏感性从52.94%提高到65.8%(p<0.001),但作者指出这些发现需要临床验证。

这项前瞻性单中心平行随机研究纳入98名ICU医生和呼吸治疗师,随机分入对照组或干预组,每组49人,分析用ASL-5000肺模拟器制作的49个场景的压力和流量波形。干预组额外显示通过人工智能算法估算的Pmus波形。主要结局为平均不同步检出率(敏感性),Pmus组显著高于对照组(65.8±16.2 vs. 52.94±8.42,p<0.001),按不同步类型分层后该效应仍存在。作者结论为显示Pmus波形提高了医护人员通过目视呼吸机波形识别患者-呼吸机不同步的能力,但这些发现需要临床验证。

要点

  • 在模拟场景中,同时显示估算的Pmus波形使ICU医生和呼吸治疗师识别不同步的敏感性从52.94±8.42提高到65.8±16.2(p<0.001)。
  • 该效应在按不同步类型分层后仍然存在。
  • 研究为前瞻性单中心随机平行设计,共纳入98名参与者,每组49人,分析49个模拟场景。
  • 作者指出这些发现需要临床验证。

适用人群与场景

  • 在重症监护室工作的医生和呼吸治疗师。

研究设计与关键结果

  • 研究类型:前瞻性单中心随机平行研究,评估显示估算的Pmus波形是否提高模拟临床场景中不同步的正确识别。
  • 样本量与分组:共纳入98名参与者,随机分为对照组和干预组,每组49人。
  • 干预与对照:两组均分析用ASL-5000肺模拟器制作的49个不同场景的压力和流量波形;干预组额外显示估算的Pmus波形。
  • 主要终点:Pmus组每名参与者的不同步识别敏感性显著高于对照组(65.8±16.2 vs. 52.94±8.42,p<0.001)。
  • 次要终点:按不同步类型分层后该效应仍然存在。

证据类型、等级与局限

  • 证据类型:前瞻性单中心随机研究。
  • 局限:原文未报告局限性。
  • 作者结论:这些发现需要临床验证。

对床旁实践的改变

  • 作者认为,显示估算的Pmus波形可能提高医护人员通过目视呼吸机波形识别患者-呼吸机不同步的能力,但这些发现需要临床验证。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

BACKGROUND: Patient-ventilator asynchronies are usually detected by visual inspection of ventilator waveforms but with low sensitivity, even when performed by experts in the field. Recently, estimation of the inspiratory muscle pressure (Pmus) waveforms through artificial intelligence algorithm has been proposed (Magnamed®, São Paulo, Brazil). We hypothesized that the display of these waveforms could help healthcare providers identify patient-ventilator asynchronies.

METHODS: A prospective single-center randomized study with parallel assignment was conducted to assess whether the display of the estimated Pmus waveform would improve the correct identification of asynchronies in simulated clinical scenarios. The primary outcome was the mean asynchrony detection rate (sensitivity). Physicians and respiratory therapists who work in intensive care units were randomized to control or intervention group. In both groups, participants analyzed pressure and flow waveforms of 49 different scenarios elaborated using the ASL-5000 lung simulator. In the intervention group the estimated Pmus waveform was displayed in addition to pressure and flow waveforms.

RESULTS: A total of 98 participants were included, 49 per group. The sensitivity per participant in identifying asynchronies was significantly higher in the Pmus group (65.8 ± 16.2 vs. 52.94 ± 8.42, p < 0.001). This effect remained when stratifying asynchronies by type.

CONCLUSIONS: We showed that the display of the Pmus waveform improved the ability of healthcare professionals to recognize patient-ventilator asynchronies by visual inspection of ventilator tracings. These findings require clinical validation.

TRIAL REGISTRATION: ClinicalTrials.gov: NTC05144607. Retrospectively registered 3 December 2021.

来源:Crit Care,查看原文。版权归原作者所有。

参考来源

  1. Silva DO, de Souza PN, de Araujo Sousa ML, et al. Impact on the ability of healthcare professionals to correctly identify patient-ventilator asynchronies of the simultaneous visualization of estimated muscle pressure curves on the ventilator display: a randomized study (Pmus study). Crit Care. 2023. PMID: 36998022. doi:10.1186/s13054-023-04414-9

供执业医师抢救与诊疗参考,不替代临床判断;剂量请结合体重、肝肾功能和本院药品规格复核,以指南原文为准。