文献速递呼吸衰竭与机械通气重症综合管理与镇静镇痛

体外膜氧合中离心泵功能障碍的急诊处理

[Emergency treatment of centrifugal pump dysfunction in extracorporeal membrane oxygenation].
PubMed 文献 · Zhonghua Wei Zhong Bing Ji Jiu Yi Xue · 2022-01-01
一句话结论VV-ECMO运行第120小时转速3822 r/min而流量仅0.2 L/min且升速无反应,快速排查判定离心泵功能障碍并成功更换。

本文为单病例分析,对象为2021年9月25日皖南医学院第一附属医院重症医学科收治的1例14岁男孩,诊断为重症支原体肺炎、重症病毒性肺炎及急性呼吸窘迫综合征(ARDS)。有创机械通气下氧合难以维持、肺实变进展严重,评估后实施静脉-静脉体外膜氧合(VV-ECMO),氧合改善。VV-ECMO建立后第120小时血流突然下降,转速3822 r/min时流量仅0.2 L/min,升速后流量无变化;此前转速3530 r/min、流量3.4 L/min且稳定。快速筛查判定为离心泵功能障碍,成功更换后流量满意。作者指出多数ECMO中心不常规监测泵前后压力,缺乏判断泵功能的直观定量技术或指标及相应临床经验,本文总结该例排查处理过程以供参考。

要点

  • VV-ECMO运行中血流突然下降、转速3822 r/min而流量仅0.2 L/min、升速后流量无变化,应快速筛查并判断离心泵功能障碍。
  • 该例此前ECMO转速3530 r/min、流量3.4 L/min且稳定,前后对比是识别泵功能障碍的关键。
  • 判定离心泵功能障碍后应更换ECMO,该例更换后流量满意。
  • 作者指出多数ECMO中心不常规监测泵前后压力,缺乏判断泵功能的直观定量技术或指标,也缺乏相应处理经验。

适用人群与场景

  • 1例14岁男孩,诊断为重症支原体肺炎、重症病毒性肺炎及急性呼吸窘迫综合征(ARDS),有创机械通气下氧合难以维持、肺实变进展严重,评估后实施静脉-静脉体外膜氧合(VV-ECMO)。

处置流程与药物剂量

  • VV-ECMO运行中血流突然下降时,快速筛查以判断是否为离心泵功能障碍;该例判定后更换ECMO,更换后流量满意。

红旗征与不能漏的情况

  • VV-ECMO运行中血流突然下降:该例转速3822 r/min而流量仅0.2 L/min,升速后流量无变化,需快速筛查离心泵功能障碍。

研究设计与关键结果

  • 单病例分析:1例14岁男孩因重症支原体肺炎、重症病毒性肺炎及ARDS接受VV-ECMO,氧合改善。
  • VV-ECMO建立后第120小时血流突然下降,转速3822 r/min时流量仅0.2 L/min,升速后流量无变化;此前转速3530 r/min、流量3.4 L/min且稳定。
  • 快速筛查判定离心泵功能障碍,成功更换ECMO后流量满意。

证据类型、等级与局限

  • 证据类型:病例分析(单病例)。
  • 局限:作者指出多数ECMO中心不常规监测泵前后压力,缺乏判断泵功能的直观定量技术或指标,也缺乏相应的处理临床经验。

对床旁实践的改变

  • 作者认为本文总结的该例ECMO泵功能障碍排查与处理过程可供参考。
  • 作者指出目前多数ECMO中心不常规监测泵前后压力,缺乏判断泵功能的直观定量技术或指标及相应临床经验。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

OBJECTIVE: To summarize the treatment process of rapid decrease in blood flow due to centrifugal pump dysfunction during extracorporeal membrane oxygenation (ECMO) and its related thinking.

METHODS: On September 25, 2021, the ECMO treatment of a 14-year-old boy with severe mycoplasma pneumonia, severe viral pneumonia and acute respiratory distress syndrome (ARDS) admitted to the department of critical care medicine of the First Affiliated Hospital of Wannan Medical College was analyzed.

RESULTS: Oxygenation of the child was difficult to maintain under invasive mechanical ventilation, and lung consolidation progressed seriously. After evaluation, venous-venous ECMO (VV-ECMO) was implemented, then oxygenation was improved. In the 120th hour after VV-ECMO establishment, the blood flow sudden decreased, the speed was 3 822 r/min, while the flow was only 0.2 L/min, more over there was no change in the flow when the speed was increased. Before that, the ECMO speed was 3 530 r/min, and the flow was up to 3.4 L/min and stable. After rapid screening, it was determined that the centrifugal pump was dysfunction. ECMO was successfully replaced and the flow was satisfactory.

CONCLUSIONS: At present, most ECMO centers do not routinely monitor the pressure before and after the pump. There is a lack of visual and quantitative techniques or indicators to judge the pump's function, and there is also a lack of corresponding clinical experience in treatment. This paper summarizes the investigation and treatment process of ECMO pump dysfunction of this case to provide reference.

来源:Zhonghua Wei Zhong Bing Ji Jiu Yi Xue,查看原文。版权归原作者所有。

参考来源

  1. Fang K, Wang J, Sun R, et al. [Emergency treatment of centrifugal pump dysfunction in extracorporeal membrane oxygenation]. Zhonghua Wei Zhong Bing Ji Jiu Yi Xue. 2022. PMID: 36473569. doi:10.3760/cma.j.cn121430-20220126-00100

供执业医师抢救与诊疗参考,不替代临床判断;剂量请结合体重、肝肾功能和本院药品规格复核,以指南原文为准。