超声比较两种机械按压装置的血流量:动物实验
该研究在8只雄性杂种猪中经氯化钾诱发心脏骤停,随机分为 LUCAS™ 3 组与 Easy Pulse 组,用超声测量股动脉收缩期峰值流速(PSV)与收缩期、舒张期血管直径并计算血流量,同时测呼气末二氧化碳(EtCO2)与按压深度。LUCAS™ 3 组按压深度更深(6.80 cm vs 3.279 cm,P<0.001),但 EtCO2 更低(19.8 mmHg vs 33.4 mmHg,P<0.001),PSV 更高(67.6 cm/s vs 55.0 cm/s,P<0.001),股动脉收缩期与舒张期直径在 Easy Pulse 组更大(收缩期 2.0 cm vs 1.5 cm,舒张期 0.8 cm vs 0.4 cm,均 P<0.001),股动脉血流量在 LUCAS™ 3 组更低(32.55 cm³/s vs 61.35 cm³/s,P<0.001)。作者认为 Easy Pulse 可能产生并维持更有效的胸腔内压。
要点
- 猪心脏骤停模型中,LUCAS™ 3 按压深度 6.80 cm 明显深于 Easy Pulse 的 3.279 cm(P<0.001)。
- Easy Pulse 组 EtCO2 为 33.4 mmHg,高于 LUCAS™ 3 组的 19.8 mmHg(P<0.001)。
- 股动脉血流量 LUCAS™ 3 组 32.55 cm³/s,低于 Easy Pulse 组的 61.35 cm³/s(P<0.001)。
- 股动脉收缩期直径 Easy Pulse 组 2.0 cm 大于 LUCAS™ 3 组的 1.5 cm,舒张期 0.8 cm 大于 0.4 cm(均 P<0.001)。
- 股动脉 PSV LUCAS™ 3 组 67.6 cm/s 高于 Easy Pulse 组的 55.0 cm/s(P<0.001)。
适用人群与场景
- 8只雄性杂种猪,经静脉注射氯化钾溶液诱发心脏骤停后接受机械心肺复苏
处置流程与药物剂量
- 心脏骤停后随机分两组,分别用 LUCAS™ 3(Physio-Control, Redmond, USA)与 Easy Pulse(Schiller Medizintechnik GMBH, Feldkirchen, Germany)行机械心肺复苏。
- 用超声测量股动脉收缩期峰值流速(PSV)及收缩期、舒张期股动脉直径,血流量按收缩期 PSV 乘以股动脉横截面积计算。
- 同时测量呼气末二氧化碳(EtCO2)与按压深度,并经动脉导管测量收缩压、平均压和舒张压。
研究设计与关键结果
- 动物实验,8只雄性杂种猪随机分为 LUCAS™ 3 组与 Easy Pulse 组,比较两种机械按压装置的心肺复苏效率。
- 按压深度 LUCAS™ 3 组 6.80 cm 深于 Easy Pulse 组 3.279 cm(P<0.001);EtCO2 LUCAS™ 3 组 19.8 mmHg 低于 Easy Pulse 组 33.4 mmHg(P<0.001)。
- 股动脉 PSV LUCAS™ 3 组 67.6 cm/s 高于 Easy Pulse 组 55.0 cm/s(P<0.001);收缩期直径 LUCAS™ 3 组 1.5 cm 小于 Easy Pulse 组 2.0 cm(P<0.001),舒张期直径 LUCAS™ 3 组 0.4 cm 小于 Easy Pulse 组 0.8 cm(P<0.001)。
- 股动脉血流量 LUCAS™ 3 组 32.55 cm³/s 低于 Easy Pulse 组 61.35 cm³/s(P<0.001)。
证据类型、等级与局限
- 证据类型:动物实验(猪心脏骤停模型)。
- 局限:原文未报告局限性。
对床旁实践的改变
- 作者认为,Easy Pulse 可能产生并维持更有效的胸腔内压。
提炼说明
- 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开
BACKGROUND: During manual chest compression, maintaining accurate compression depth and consistency is a challenge. Therefore, mechanical chest compression devices(mCCDs) have been increasingly incorporated in clinical practice. Evaluation and comparison of the efficacy of these devices is critical for extensive clinical application. Hence, this study compared the cardiopulmonary resuscitation(CPR) efficiency of two chest compression devices, LUCAS™ 3(Physio-Control, Redmond, USA) and Easy Pulse (Schiller Medizintechnik GMBH, Feldkirchen, Germany), in terms of blood flow using ultrasonography(USG) in a swine model.
METHODS: A swine model was used to compare two mCCDs, LUCAS™ 3 and Easy Pulse. Cardiac arrest was induced by injecting potassium chloride(KCl) solution in eight male mongrel pigs and the animals were randomly divided into two groups. Mechanical CPR was provided to two groups using LUCAS™ 3(LUCAS™ 3 group) and Easy Pulse(Easy Pulse group). USG was used to measure hemodynamic parameters including femoral peak systolic velocity(PSV) and femoral artery diameters(diameter during systole and diastole). Blood flow rate was calculated by multiplying the PSV and cross-sectional area of the femoral artery during systole. The end-tidal carbon dioxide(EtCo2), chest compression depth was measured. Systolic blood pressure, mean blood pressure, and diastolic blood pressure were also measured using an arterial catheter.
RESULTS: The chest compression depth was much deeper in LUCAS™ 3 group than Easy Pulse group(LUCAS™ 3: 6.80 cm; Easy Pulse: 3.279 cm, p < 0.001). However, EtCo2 was lower in the LUCAS™ 3 group(LUCAS™ 3: 19.8 mmHg; Easy Pulse: 33.4 mmHg, p < 0.001). The PSV was higher in the LUCAS™ 3 group(LUCAS™ 3: 67.6 cm s-1; Easy Pulse: 55.0 cm s-1, p < 0.001), while the systolic(LUCAS™ 3: 1.5 cm; Easy Pulse: 2.0 cm, p < 0.001) and diastolic diameters were larger in the Easy Pulse group(LUCAS™ 3: 0.4; Easy Pulse: 0.8 cm, p < 0.001). The femoral flood flow rate was also lower in the LUCAS™ 3 group(LUCAS™ 3: 32.55 cm3/s; Easy Pulse: 61.35 cm3/s, p < 0.001).
CONCLUSION: The Easy Pulse had a shallower compression depth and slower PSV but had a wider systolic diameter in the femoral artery as compared to that in LUCAS™ 3. Blood flow and EtCo2 were higher in the easy pulse group probably because of the wider diameter. Therefore, an easy pulse may create and maintain more effective intrathoracic pressure.