氧储备指数(ORI)检测重症患者高氧血症的能力

The ability of Oxygen Reserve Index® to detect hyperoxia in critically ill patients.
PubMed 文献 · Ann Intensive Care · 2022-05-16
一句话结论在神经重症监护病房的脑损伤患者中,ORI 诊断高氧血症的能力较低,SpO2 检测效果优于 ORI。

本研究为单中心研究,在神经重症监护病房的脑损伤患者中评估 ORI 检测动脉氧分压(PaO2)> 100 mmHg 和 > 120 mmHg 的诊断准确性,并与 SpO2 比较。共收集 62 例患者的 696 次同步 ORI 与 PaO2 测量,每 6 小时一次,连续 3 天。ORI 检测 PaO2 > 100 mmHg 的 AUROC 为 0.567(95%CI 0.566-0.569),灵敏度 0.233(95%CI 0.230-0.235),特异度 0.909(95%CI 0.907-0.910);SpO2 的 AUROC 为 0.771(95%CI 0.770-0.773),灵敏度 0.715(95%CI 0.712-0.718),特异度 0.700(95%CI 0.697-0.703)。检测 PaO2 > 120 mmHg 时,ORI 与 SpO2 的 AUROC 分别为 0.584(95%CI 0.582-0.586)和 0.764(95%CI 0.762-0.766),SpO2 的 AUROC 显著高于 ORI(p < 0.01)。作者认为,在该特定人群中 ORI 诊断高氧血症的能力较低,SpO2 提供更好的检测。

要点

  • 在神经重症监护病房的脑损伤患者中,ORI 检测 PaO2 > 100 mmHg 的 AUROC 仅为 0.567(95%CI 0.566-0.569),灵敏度 0.233(95%CI 0.230-0.235),特异度 0.909(95%CI 0.907-0.910)。
  • SpO2 检测 PaO2 > 100 mmHg 的 AUROC 为 0.771(95%CI 0.770-0.773),灵敏度 0.715(95%CI 0.712-0.718),特异度 0.700(95%CI 0.697-0.703),优于 ORI。
  • 检测 PaO2 > 120 mmHg 时,ORI 的 AUROC 为 0.584(95%CI 0.582-0.586),SpO2 为 0.764(95%CI 0.762-0.766),SpO2 显著高于 ORI(p < 0.01)。
  • ORI 与 PaO2 的相关性为 r = 0.13(考虑重复测量)。
  • 镇静、去甲肾上腺素输注、动脉血二氧化碳分压、血红蛋白水平和灌注指数均未影响诊断性能。

适用人群与场景

  • 入住神经重症监护病房的脑损伤患者。

研究设计与关键结果

  • 单中心研究,共收集 62 例患者的 696 次同步 ORI 与 PaO2 测量,每 6 小时一次,连续 3 天。
  • 考虑重复测量,ORI 与 PaO2 的相关性为 r = 0.13。
  • ORI 检测 PaO2 > 100 mmHg 的 AUROC 为 0.567(95%CI 0.566-0.569),灵敏度 0.233(95%CI 0.230-0.235),特异度 0.909(95%CI 0.907-0.910)。
  • SpO2 检测 PaO2 > 100 mmHg 的 AUROC 为 0.771(95%CI 0.770-0.773),灵敏度 0.715(95%CI 0.712-0.718),特异度 0.700(95%CI 0.697-0.703)。
  • ORI 检测 PaO2 > 120 mmHg 的 AUROC 为 0.584(95%CI 0.582-0.586),SpO2 为 0.764(95%CI 0.762-0.766)。
  • SpO2 的 AUROC 显著高于 ORI(p < 0.01)。
  • 诊断性能不受镇静、去甲肾上腺素输注、动脉血二氧化碳分压、血红蛋白水平和灌注指数影响。

证据类型、等级与局限

  • 单中心研究。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 作者认为,在入住神经重症监护病房的脑损伤患者中,ORI 诊断高氧血症的能力较低,SpO2 提供更好的检测。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「诊断标准与关键阈值」「处置流程与药物剂量」「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

BACKGROUND: Hyperoxia is associated with increased morbidity and mortality in the intensive care unit. Classical noninvasive measurements of oxygen saturation with pulse oximeters are unable to detect hyperoxia. The Oxygen Reserve Index (ORI) is a continuous noninvasive parameter provided by a multi-wave pulse oximeter that can detect hyperoxia. Primary objective was to evaluate the diagnostic accuracy of the ORI for detecting arterial oxygen tension (PaO2) > 100 mmHg in neurocritical care patients. Secondary objectives were to test the ability of ORI to detect PaO2 > 120 mmHg and the ability of pulse oximetry (SpO2) to detect PaO2 > 100 mmHg and PaO2 > 120 mmHg.

METHODS: In this single-center study, we collected ORI and arterial blood samples every 6 h for 3 consecutive days. Diagnostic performance was estimated using the area under the receiver operating characteristic curve (AUROC).

RESULTS: There were 696 simultaneous measurements of ORI and PaO2 in 62 patients. Considering the repeated measurements, the correlation between ORI and PaO2 was r = 0.13. The area under the receiver operating characteristic curve (AUROC), obtained to test the ability of ORI to detect PaO2 > 100 mmHg, was 0.567 (95% confidence interval = 0.566-0.569) with a sensitivity of 0.233 (95%CI = 0.230-0.235) and a specificity of 0.909 (95%CI = 0.907-0.910). The AUROC value obtained to test the ability of SpO2 to detect a PaO2 > 100 mmHg was 0.771 (95%CI = 0.770-0.773) with a sensitivity of 0.715 (95%CI = 0.712-0.718) and a specificity of 0.700 (95%CI = 0.697-0.703). The diagnostic performance of ORI and SpO2 for detecting PaO2 > 120 mmHg was AUROC = 0.584 (95%CI = 0.582-0.586) and 0.764 (95%CI = 0.762-0.766), respectively. The AUROC obtained for SpO2 was significantly higher than that for ORI (p < 0.01). Diagnostic performance was not affected by sedation, norepinephrine infusion, arterial partial pressure of carbon dioxide, hemoglobin level and perfusion index.

CONCLUSION: In a specific population of brain-injured patients hospitalized in a neurointensive care unit, our results suggest that the ability of ORI to diagnose hyperoxia is relatively low and that SpO2 provides better detection.

来源:Ann Intensive Care,查看原文。版权归原作者所有。

参考来源

  1. de Courson H, Julien-Laferrière T, Georges D, et al. The ability of Oxygen Reserve Index® to detect hyperoxia in critically ill patients. Ann Intensive Care. 2022. PMID: 35576087. doi:10.1186/s13613-022-01012-w

供执业医师抢救与诊疗参考,不替代临床判断;剂量请结合体重、肝肾功能和本院药品规格复核,以指南原文为准。