文献速递复苏、休克与脓毒症创伤与烧伤

先进部分阻断控制器在猪失血性休克模型中提高靶向区域优化精度的可行性研究

Advanced partial occlusion controller allows for increased precision during targeted regional optimization in a porcine model of hemorrhagic shock.
PubMed 文献 · J Trauma Acute Care Surg · 2022-01-01
一句话结论在猪失血性休克模型中,先进部分阻断控制器(APOC)在目标远端平均动脉压25 mmHg和45 mmHg时较手动调节更精准地维持靶向区域优化(TRO),35 mmHg时无差异。

本研究为猪失血性休克模型研究,48头雌性猪随机分为目标远端平均动脉压25、35、45 mmHg三组,每组分别采用手动(MAN)或先进部分阻断控制器(APOC)调节(每组n=8)。通过肝裂伤制造非控制性出血,行靶向区域优化(TRO)85分钟,随后手术止血并进入6小时重症监护阶段,连续监测近端和远端平均动脉压及流量。结果显示,目标远端平均动脉压25 mmHg时APOC与MAN达到的平均压无差异(26.2±1.05 vs 26.1±1.78 mmHg),但APOC偏差显著更小(10.9% vs 14.9%,p<0.0001);45 mmHg时平均压无差异(44.0±0.900 vs 45.2±1.31 mmHg),APOC偏差也更小(9.34% vs 11.9%,p<0.0001);35 mmHg时平均压(34.6 vs 33.7 mmHg)和偏差(9.95% vs 10.4%)均无差异。APOC平均每实验进行77次球囊体积调整,手动调节为29次。结论认为APOC能持续实现精准调节的TRO,且在25 mmHg和45 mmHg组较手动调节偏差更小。

要点

  • 在猪失血性休克模型中,APOC在目标远端平均动脉压25 mmHg时达到的平均压与手动调节无差异(26.2±1.05 vs 26.1±1.78 mmHg),但偏差显著更低(10.9% vs 14.9%,p<0.0001)。
  • 目标远端平均动脉压45 mmHg时,APOC与手动调节的平均压无差异(44.0±0.900 vs 45.2±1.31 mmHg),APOC偏差更低(9.34% vs 11.9%,p<0.0001)。
  • 目标远端平均动脉压35 mmHg时,APOC与手动调节的平均压(34.6 vs 33.7 mmHg)和偏差(9.95% vs 10.4%)均无差异。
  • APOC平均每实验进行77次球囊体积调整,手动调节为29次。

适用人群与场景

  • 雌性猪(n=48,体重68.1±0.7 kg),肝裂伤致非控制性出血模型。

诊断标准与关键阈值

  • 目标远端平均动脉压设定为25 mmHg、35 mmHg或45 mmHg。

处置流程与药物剂量

  • 行靶向区域优化(TRO)85分钟,随后手术止血并进入6小时重症监护阶段。
  • 采用手动(MAN)或先进部分阻断控制器(APOC)调节球囊体积以维持目标远端平均动脉压。

研究设计与关键结果

  • 研究类型为猪失血性休克模型随机对照实验,样本量n=48(每组n=8),比较APOC与手动调节在三个目标远端平均动脉压水平的控制精度。
  • 目标远端平均动脉压25 mmHg:APOC组达到26.2±1.05 mmHg,MAN组26.1±1.78 mmHg,平均压无差异;APOC偏差10.9%,MAN偏差14.9%,p<0.0001。
  • 目标远端平均动脉压45 mmHg:APOC组达到44.0±0.900 mmHg,MAN组45.2±1.31 mmHg,平均压无差异;APOC偏差9.34%,MAN偏差11.9%,p<0.0001。
  • 目标远端平均动脉压35 mmHg:APOC组平均压34.6 mmHg,MAN组33.7 mmHg,无差异;APOC偏差9.95%,MAN偏差10.4%,无差异。
  • APOC平均每实验进行77次球囊体积调整,手动调节平均29次。

证据类型、等级与局限

  • 证据类型:动物实验(猪失血性休克模型)。
  • 局限:原文未报告局限性。

对床旁实践的改变

  • 在猪失血性休克模型中,APOC在目标远端平均动脉压25 mmHg和45 mmHg时可减少血压偏差,但35 mmHg时与手动调节无差异。

提炼说明

  • 本条依据论文摘要整理;摘要未涉及「红旗征与不能漏的情况」「指南或共识出处」,故未列出这些栏目;全文请点「阅读原文」。
英文原文(摘要)· 点击展开

BACKGROUND: Targeted regional optimization (TRO), a partial resuscitative endovascular balloon occlusion of the aorta strategy, may mitigate distal ischemia and extend the window of effectiveness for this adjunct. An automated device may allow greater control and precise regulation of flow past the balloon, while being less resource-intensive. The objective of this study was to assess the technical feasibility of the novel advanced partial occlusion controller (APOC) in achieving TRO at multiple distal pressures.

METHODS: Female swine (n = 48, 68.1 ± 0.7 kg) were randomized to a target distal mean arterial pressure (MAP) of 25 mm Hg, 35 mm Hg, or 45 mm Hg by either manual (MAN) or APOC regulation (n = 8 per group). Uncontrolled hemorrhage was generated by liver laceration. Targeted regional optimization was performed for 85 minutes, followed by surgical control and a 6-hour critical care phase. Proximal and distal MAP and flow rates were measured continuously.

RESULTS: At a target distal MAP of 25 mm Hg, there was no difference in the MAP attained (APOC: 26.2 ± 1.05 vs. MAN: 26.1 ± 1.78 mm Hg) but the APOC had significantly less deviance (10.9%) than manual titration (14.9%, p < 0.0001). Similarly, at a target distal MAP of 45 mm Hg, there was no difference in mean pressure (44.0 ± 0.900 mm Hg vs. 45.2 ± 1.31 mm Hg) but APOC had less deviance (9.34% vs. 11.9%, p < 0.0001). There was no difference between APOC and MAN in mean (34.6 mm Hg vs. 33.7 mm Hg) or deviance (9.95% vs. 10.4%) at a target distal MAP of 35 mm Hg, respectively. The APOC made on average 77 balloon volume adjustments per experiment compared with 29 by manual titrations.

CONCLUSION: The novel APOC consistently achieved and sustained precisely regulated TRO across all groups and demonstrated reduced deviance at the 25 mm Hg and 45 mm Hg groups compared with manual titration.

来源:J Trauma Acute Care Surg,查看原文。版权归原作者所有。